Significant reduction in dental cone beam computed tomography (CBCT) eye dose through the use of leaded glasses.

R Prins1, L T Dauer, D C Colosi, B Quinn, N J Kleiman, G C Bohle, B Holohan, A Al-Najjar, T Fernandez, M Bonvento, R D Faber, H Ching, A D Goren.   

Abstract

OBJECTIVE: In light of the increased recognition of the potential for lens opacification after low-dose radiation exposures, we investigated the effect of leaded eyeglasses worn during dental cone-beam computerized tomography (CBCT) procedures on the radiation absorbed dose to the eye and suggest simple methods to reduce risk of radiation cataract development. STUDY
DESIGN: Dose measurements were conducted with the use of 3 anthropomorphic phantoms: male (Alderson radiation therapy phantom), female (CIRS), and juvenile male (CIRS). All exposures were performed on the same dental CBCT machine (Imtec, Ardmore, OK) using 2 different scanning techniques but with identical machine parameters (120 kVp, 3.8 mA, 7.8 s). Scans were performed with and without leaded glasses and repeated 3 times. All measurements were recorded using calibrated thermoluminescent dosimeters and optical luminescent dosimetry.
RESULTS: Leaded glasses worn by adult and pediatric patients during CBCT scans may reduce radiation dose to the lens of the eye by as much as 67% (from 0.135 ± 0.004 mGy to 0.044 ± 0.002 mGy in pediatric patients).
CONCLUSIONS: Leaded glasses do not appear to have a deleterious effect on the image quality in the area of clinical significance for dental imaging.
Copyright © 2011 Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21802322     DOI: 10.1016/j.tripleo.2011.04.041

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  9 in total

1.  Potential of dosage reduction in cone-beam-computed tomography (CBCT) for radiological diagnostics of the paranasal sinuses.

Authors:  C Güldner; A Ningo; J Voigt; I Diogo; J Heinrichs; R Weber; T Wilhelm; M Fiebich
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-09-18       Impact factor: 2.503

2.  Dosimetry of a cone beam CT device for oral and maxillofacial radiology using Monte Carlo techniques and ICRP adult reference computational phantoms.

Authors:  J J Morant; M Salvadó; I Hernández-Girón; R Casanovas; R Ortega; A Calzado
Journal:  Dentomaxillofac Radiol       Date:  2012-08-29       Impact factor: 2.419

3.  Effect of leaded glasses and thyroid shielding on cone beam CT radiation dose in an adult female phantom.

Authors:  A D Goren; R D Prins; L T Dauer; B Quinn; A Al-Najjar; R D Faber; G Patchell; I Branets; D C Colosi
Journal:  Dentomaxillofac Radiol       Date:  2013-02-14       Impact factor: 2.419

4.  Patient radiation dose and protection from cone-beam computed tomography.

Authors:  Gang Li
Journal:  Imaging Sci Dent       Date:  2013-06-14

5.  Measuring absorbed dose for i-CAT CBCT examinations in child, adolescent and adult phantoms.

Authors:  E Choi; N L Ford
Journal:  Dentomaxillofac Radiol       Date:  2015-03-18       Impact factor: 2.419

Review 6.  Cone beam computed tomography in paediatric dentistry: overview of recent literature.

Authors:  J K M Aps
Journal:  Eur Arch Paediatr Dent       Date:  2013-06

7.  Comparison of Tissue Density in Hounsfield Units in Computed Tomography and Cone Beam Computed Tomography.

Authors:  Masoud Varshowsaz; Sepideh Goorang; Sara Ehsani; Zeynab Azizi; Sepideh Rahimian
Journal:  J Dent (Tehran)       Date:  2016-03

8.  Can leaded glasses protect the eye lens in patients undergoing neck computed tomography?

Authors:  Reza Abedi; Naser Ghaemian; Ali Shabestani Monfared; Mohammad Kiapour; Razzagh Abedi-Firouzjah; Fatemeh Niksirat; Alaba Tolulope Agbele; Kourosh Ebrahimnejad Gorji
Journal:  J Clin Transl Res       Date:  2021-07-16

9.  Measurement of skin dose from cone-beam computed tomography imaging.

Authors:  Sercan Akyalcin; Jeryl D English; Kenneth M Abramovitch; Xiujiang J Rong
Journal:  Head Face Med       Date:  2013-10-09       Impact factor: 2.151

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.