Dose reduction of cone beam CT scanning for the entire oral and maxillofacial regions with thyroid collars.

X M Qu1, G Li, G C H Sanderink, Z Y Zhang, X C Ma.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the influence of thyroid collars on radiation dose during cone beam CT (CBCT) scanning.
METHODS: Average tissue-absorbed dose for a NewTom 9000 CBCT scanner (Quantitative Radiology, Verona, Italy) was measured using thermoluminescent dosemeter chips in a phantom. The scans were carried out with and without thyroid collars. Effective organ dose and total effective dose were derived using International Commission on Radiological Protection 2007 recommendations.
RESULTS: The effective organ doses for the thyroid gland and oesophagus were 31.0 µSv and 2.4 µSv, respectively, during CBCT scanning without a collar around the neck. When the thyroid collars were used loosely around the neck, no effective organ dose reduction was observed. When one thyroid collar was used tightly on the front of the neck, the effective organ dose for the thyroid gland and oesophagus were reduced to 15.9 µSv (48.7% reduction) and 1.4 µSv (41.7% reduction), respectively. Similar organ dose reduction (46.5% and 41.7%) was achieved when CBCT scanning was performed with two collars tightly on the front and back of the neck. However, the differences to the total effective dose were not significant among the scans with and without collars around the neck (p = 0.775).
CONCLUSIONS: Thyroid collars can effectively reduce the radiation dose to the thyroid and oesophagus if used appropriately.

Mesh:

Year:  2012        PMID: 22707330      PMCID: PMC3520363          DOI: 10.1259/dmfr/30200901

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  20 in total

1.  Dose reduction in maxillofacial imaging using low dose Cone Beam CT.

Authors:  Kostas Tsiklakis; Catherine Donta; Sophia Gavala; Kety Karayianni; Vasiliki Kamenopoulou; Costas J Hourdakis
Journal:  Eur J Radiol       Date:  2005-06-22       Impact factor: 3.528

2.  Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT.

Authors:  J B Ludlow; L E Davies-Ludlow; S L Brooks; W B Howerton
Journal:  Dentomaxillofac Radiol       Date:  2006-07       Impact factor: 2.419

3.  Comparison of clinical, periapical radiograph, and cone-beam volume tomography measurement techniques for assessing bone level changes following regenerative periodontal therapy.

Authors:  Brently A Grimard; Matt J Hoidal; Michael P Mills; James T Mellonig; Pirkka V Nummikoski; Brian L Mealey
Journal:  J Periodontol       Date:  2009-01       Impact factor: 6.993

4.  American Academy of Oral and Maxillofacial Radiology executive opinion statement on performing and interpreting diagnostic cone beam computed tomography.

Authors:  Laurie Carter; Allan G Farman; James Geist; William C Scarfe; Christos Angelopoulos; Madhu K Nair; Charles F Hildebolt; Donald Tyndall; Michael Shrout
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-10

5.  Effective dose from cone beam CT examinations in dentistry.

Authors:  J A Roberts; N A Drage; J Davies; D W Thomas
Journal:  Br J Radiol       Date:  2008-10-13       Impact factor: 3.039

6.  Thyroid shields versus z-axis automatic tube current modulation for dose reduction at neck CT.

Authors:  David A Leswick; Megan M Hunt; Steven T Webster; Derek A Fladeland
Journal:  Radiology       Date:  2008-09-09       Impact factor: 11.105

7.  Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology.

Authors:  John B Ludlow; Marija Ivanovic
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-05-27

8.  Cone beam computed tomography: a useful tool in orthodontic diagnosis and treatment planning.

Authors:  Sarah J Merrett; Nicholas A Drage; Peter Durning
Journal:  J Orthod       Date:  2009-09

9.  Influence of CBCT exposure conditions on radiation dose.

Authors:  J Martin Palomo; Pejavar S Rao; Mark G Hans
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-04-18

10.  Detection of vertical root fractures in endodontically treated teeth by a cone beam computed tomography scan.

Authors:  Bassam Hassan; Maria Elissavet Metska; Ahmet Rifat Ozok; Paul van der Stelt; Paul Rudolf Wesselink
Journal:  J Endod       Date:  2009-05       Impact factor: 4.171

View more
  9 in total

Review 1.  Effective dose of cone beam CT (CBCT) of the facial skeleton: a systematic review.

Authors:  A Al-Okshi; C Lindh; H Salé; M Gunnarsson; M Rohlin
Journal:  Br J Radiol       Date:  2015-01       Impact factor: 3.039

2.  Thyroid shielding in cone beam computed tomography: recommendations towards appropriate use.

Authors:  Ruben Pauwels; Keith Horner; Jenia Vassileva; Madan M Rehani
Journal:  Dentomaxillofac Radiol       Date:  2019-07-09       Impact factor: 2.419

3.  Effect of leaded glasses and thyroid shielding on cone beam CT radiation dose in an adult female phantom.

Authors:  A D Goren; R D Prins; L T Dauer; B Quinn; A Al-Najjar; R D Faber; G Patchell; I Branets; D C Colosi
Journal:  Dentomaxillofac Radiol       Date:  2013-02-14       Impact factor: 2.419

4.  Patient radiation dose and protection from cone-beam computed tomography.

Authors:  Gang Li
Journal:  Imaging Sci Dent       Date:  2013-06-14

5.  Effectiveness of thyroid gland shielding in dental CBCT using a paediatric anthropomorphic phantom.

Authors:  A Hidalgo; J Davies; K Horner; C Theodorakou
Journal:  Dentomaxillofac Radiol       Date:  2014-11-20       Impact factor: 2.419

Review 6.  Cone beam computed tomography in paediatric dentistry: overview of recent literature.

Authors:  J K M Aps
Journal:  Eur Arch Paediatr Dent       Date:  2013-06

Review 7.  Effective dose of dental CBCT-a meta analysis of published data and additional data for nine CBCT units.

Authors:  J B Ludlow; R Timothy; C Walker; R Hunter; E Benavides; D B Samuelson; M J Scheske
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

8.  Measurement of skin dose from cone-beam computed tomography imaging.

Authors:  Sercan Akyalcin; Jeryl D English; Kenneth M Abramovitch; Xiujiang J Rong
Journal:  Head Face Med       Date:  2013-10-09       Impact factor: 2.151

9.  Digital orthodontic radiographic set versus cone-beam computed tomography: an evaluation of the effective dose.

Authors:  Lillian Atsumi Simabuguro Chinem; Beatriz de Souza Vilella; Cláudia Lúcia de Pinho Maurício; Lucia Viviana Canevaro; Luiz Fernando Deluiz; Oswaldo de Vasconcellos Vilella
Journal:  Dental Press J Orthod       Date:  2016 Jul-Aug
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.