Evaluation of patient organ doses from kilovoltage cone-beam CT imaging in radiation therapy.

Alper Özseven1, Bahar Dirican2.   

Abstract

BACKGROUND: Currently, CBCT system is an indispensable component of radiation therapy units. Because of that, it is important in treatment planning and diagnosis. CBCT is also an crucial tool for patient positioning and verification in image-guided radiation therapy (IGRT). Therefore, it is critical to investigate the patient organ doses arising from CBCT imaging. The purpose of this study is to evaluate patient organ doses and effective dose to patients from three different protocols of Elekta Synergy XVI system for kV CBCT imaging examinations in image guided radiation therapy.
MATERIALS AND METHODS: Organ dose measurements were done with thermoluminescent dosimeters in Alderson RA NDO male phantom for head & neck (H&N), chest and pelvis protocols of the Elekta Synergy XVI kV CBCT system. From the measured organ dose, effective dose to patients were calculated according to the International Commission on Radiological Protection 103 report recommendations.
RESULTS: For H&N, chest and pelvis scans, the organ doses were in the range of 0.03-3.43 mGy, 6.04-22.94 mGy and 2.5-25.28 mGy, respectively. The calculated effective doses were 0.25 mSv, 5.56 mSv and 4.72 mSv, respectively.
CONCLUSION: The obtained results were consistent with the most published studies in the literature. Although the doses to patient organs from the kV CBCT system were relatively low when compared with the prescribed treatment dose, the amount of delivered dose should be monitored and recorded carefully in order to avoid secondary cancer risk, especially in pediatric examinations.
© 2021 Greater Poland Cancer Centre.

Entities:  

Keywords:  Alderson Rando phantom; effective dose; equivalent organ dose; image guided radiation therapy; kV CBCT

Year:  2021        PMID: 34211776      PMCID: PMC8241296          DOI: 10.5603/RPOR.a2021.0038

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  22 in total

Review 1.  Emergent technologies for 3-dimensional image-guided radiation delivery.

Authors:  David A Jaffray
Journal:  Semin Radiat Oncol       Date:  2005-07       Impact factor: 5.934

2.  Organ and effective doses from verification techniques in image-guided radiotherapy.

Authors:  V Dufek; I Horakova; L Novak
Journal:  Radiat Prot Dosimetry       Date:  2011-08-04       Impact factor: 0.972

3.  TLD measurements and Monte Carlo calculations of head and neck organ and effective doses for cone beam computed tomography using 3D Accuitomo 170.

Authors:  Marina Ernst; Peter Manser; Karl Dula; Werner Volken; Marco Fm Stampanoni; Michael K Fix
Journal:  Dentomaxillofac Radiol       Date:  2017-07-27       Impact factor: 2.419

4.  Determination of the effective dose delivered by image guided radiotherapy in head & neck and breast treatments.

Authors:  Mireille Conrad; Grégory Bolard; Marie Nowak; Berardino De Bari; Wendy Jeanneret-Sozzi; Jean Bourhis; Jean-François Germond; François Bochud; Raphaël Moeckli
Journal:  Z Med Phys       Date:  2018-02-06       Impact factor: 4.820

5.  Age-dependent doses to members of the public from intake of radionuclides: Part 2. Ingestion dose coefficients. A report of a Task Group of Committee 2 of the International Commission on Radiological Protection.

Authors: 
Journal:  Ann ICRP       Date:  1993

6.  Accuracy of radiotherapy dose calculations based on cone-beam CT: comparison of deformable registration and image correction based methods.

Authors:  T E Marchant; K D Joshi; C J Moore
Journal:  Phys Med Biol       Date:  2018-03-12       Impact factor: 3.609

7.  Radiation exposure to patients from image guidance procedures and techniques to reduce the imaging dose.

Authors:  George X Ding; Peter Munro
Journal:  Radiother Oncol       Date:  2013-07-02       Impact factor: 6.280

Review 8.  Exposure Risks Among Children Undergoing Radiation Therapy: Considerations in the Era of Image Guided Radiation Therapy.

Authors:  Clayton B Hess; Holly M Thompson; Stanley H Benedict; J Anthony Seibert; Kenneth Wong; Andrew T Vaughan; Allen M Chen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2016-01-05       Impact factor: 7.038

9.  An organ and effective dose study of XVI and OBI cone-beam CT systems.

Authors:  Daniel E Hyer; Christopher F Serago; Siyong Kim; Jonathan G Li; David E Hintenlang
Journal:  J Appl Clin Med Phys       Date:  2010-04-17       Impact factor: 2.102

10.  Measurement of skin dose from cone-beam computed tomography imaging.

Authors:  Sercan Akyalcin; Jeryl D English; Kenneth M Abramovitch; Xiujiang J Rong
Journal:  Head Face Med       Date:  2013-10-09       Impact factor: 2.151

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.