Characteristics of an OSLD in the diagnostic energy range.

Rani M Al-Senan1, Mustapha R Hatab.   

Abstract

PURPOSE: Optically stimulated luminescence (OSL) dosimetry has been recently introduced in radiation therapy as a potential alternative to the thermoluminescent dosimeter (TLD) system. The aim of this study was to investigate the feasibility of using OSL point dosimeters in the energy range used in diagnostic imaging.
METHODS: NanoDot OSL dosimeters (OSLDs) were used in this study, which started with testing the homogeneity of a new packet of nanoDots. Reproducibility and the effect of optical treatment (bleaching) were then examined, followed by an investigation of the effect of accumulated dose on the OSLD indicated doses. OSLD linearity, angular dependence, and energy dependence were also studied. Furthermore, comparison with LiF:Mg,Ti TLD chips using standard CT dose phantoms at 80 and 120 kVp settings was performed.
RESULTS: Batch homogeneity showed a coefficient of variation of <5%. Single-irradiation measurements with bleaching after each OSL readout was found to be associated with a 3.3% reproducibility (one standard deviation measured with a 8 mGy test dose), and no systematic change in OSLDs sensitivity could be noted from measurement to measurement. In contrast, the multiple-irradiation readout without bleaching in between measurements was found to be associated with an uncertainty (using a 6 mGy test dose) that systematically increased with accumulated dose, reaching 42% at 82 mGy. Good linearity was shown by nanoDots under general x-ray, CT, and mammography units with an R2 > 0.99. The angular dependence test showed a drop of approximately 70% in the OSLD response at 90 degrees in mammography (25 kVp). With the general radiography unit, the maximum drop was 40% at 80 kVp and 20% at 120 kVp, and it was only 10% with CT at both 80 and 120 kVp. The energy dependence study showed a range of ion chamber-to-OSLDs ratios between 0.81 and 1.56, at the energies investigated (29-62 keV). A paired t-test for comparing the OSLDs and TLDs showed no significant variation (p > 0.1).
CONCLUSIONS: OSLDs exhibited good batch homogeneity (<5%) and reproducibility (3.3%), as well as a linear response. In addition, they showed no statistically significant difference with TLDs in CT measurements (p > 0.1). However, high uncertainty (42%) in the dose estimate was found as a result of relatively high accumulated dose. Furthermore, nanoDots showed high angular dependence (up to 70%) in low kVp techniques. Energy dependence of about 60% was found, and correction factors were suggested for the range of energies investigated. Therefore, if angular and energy dependences are taken into consideration and the uncertainty associated with accumulated dose is avoided, OSLDs (nanoDots) can be suitable for use as point dosimeters in diagnostic settings.

Entities:  

Mesh:

Year:  2011        PMID: 21859040     DOI: 10.1118/1.3602456

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  Radiation dose uncertainty and correction for a mouse orthotopic and xenograft irradiation model.

Authors:  Gregory N Gan; Cem Altunbas; John J Morton; Justin Eagles; Jennifer Backus; Wayne Dzingle; David Raben; Antonio Jimeno
Journal:  Int J Radiat Biol       Date:  2015-12-21       Impact factor: 2.694

2.  A method to acquire CT organ dose map using OSL dosimeters and ATOM anthropomorphic phantoms.

Authors:  Da Zhang; Xinhua Li; Yiming Gao; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

3.  Practical calibration curve of small-type optically stimulated luminescence (OSL) dosimeter for evaluation of entrance skin dose in the diagnostic X-ray region.

Authors:  Kazuki Takegami; Hiroaki Hayashi; Hiroki Okino; Natsumi Kimoto; Itsumi Maehata; Yuki Kanazawa; Tohru Okazaki; Ikuo Kobayashi
Journal:  Radiol Phys Technol       Date:  2015-05-15

4.  The effective dose of different scanning protocols using the Sirona GALILEOS(®) comfort CBCT scanner.

Authors:  D Chambers; R Bohay; L Kaci; R Barnett; J Battista
Journal:  Dentomaxillofac Radiol       Date:  2014-10-31       Impact factor: 2.419

5.  Characterization of the nanoDot OSLD dosimeter in CT.

Authors:  Sarah B Scarboro; Dianna Cody; Paola Alvarez; David Followill; Laurence Court; Francesco C Stingo; Di Zhang; Michael McNitt-Gray; Stephen F Kry
Journal:  Med Phys       Date:  2015-04       Impact factor: 4.071

6.  Potential of dosage reduction in cone-beam-computed tomography (CBCT) for radiological diagnostics of the paranasal sinuses.

Authors:  C Güldner; A Ningo; J Voigt; I Diogo; J Heinrichs; R Weber; T Wilhelm; M Fiebich
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-09-18       Impact factor: 2.503

7.  In vitro dose measurements in a human cadaver with abdomen/pelvis CT scans.

Authors:  Da Zhang; Atul Padole; Xinhua Li; Sarabjeet Singh; Ranish Deedar Ali Khawaja; Diego Lira; Tianyu Liu; Jim Q Shi; Alexi Otrakji; Mannudeep K Kalra; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

8.  A more accurate and safer method for the measurement of scattered radiation in X-ray examination rooms.

Authors:  Tokiko Nakamura; Suzuki Shoichi; Yasutaka Takei; Masanao Kobayashi; Vergil Cruz; Ikuo Kobayashi; Satoshi Asegawa; Kyoichi Kato
Journal:  Radiol Phys Technol       Date:  2019-12-23

9.  Physical validation of a Monte Carlo-based, phantom-derived approach to computed tomography organ dosimetry under tube current modulation.

Authors:  Elliott J Stepusin; Daniel J Long; Kayla R Ficarrotta; David E Hintenlang; Wesley E Bolch
Journal:  Med Phys       Date:  2017-09-22       Impact factor: 4.071

10.  Radiation Doses in Patient Eye Lenses during Interventional Neuroradiology Procedures.

Authors:  R M Sánchez; E Vañó; J M Fernández; S Rosati; L López-Ibor
Journal:  AJNR Am J Neuroradiol       Date:  2015-11-05       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.