R Ghanbarnezhad Farshi1,2, A Mesbahi2, M Johari3, Ü Kara4, N Gharehaghaji5.
Abstract
BACKGROUND: While the benefits of cone-beam computed tomography (CBCT) are well known in maxillofacial imaging, the use of this modality is not risk-free.Entities:
Keywords: Dentistry; Entrance Surface Dose ; Maxillofacial Imaging ; Radiation Dosimetry ; Thermoluminescent Dosimetry ; Cone-Beam Computed Tomography
Year: 2019 PMID: 30881934 PMCID: PMC6409370
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Demographical and anatomical information of patients
| Center A | Center B | |||||
|---|---|---|---|---|---|---|
| Male | Female | Total | Male | Female | Total | |
| Number of patients | 16 | 16 | 32 | 20 | 12 | 32 |
| Percentage (%) | 50 | 50 | 100 | 62.5 | 37.5 | 100 |
|
|
| |||||
| Male | Female | Total | Male | Female | Total | |
| Age (years) | 45.25 | 52.5 | 48.8 | 48.4 | 51 | 49.14 |
| Weight (kg) | 87.75 | 62.75 | 75.25 | 79.4 | 74.5 | 78 |
| Height (cm) | 179.5 | 157.5 | 168.6 | 169.4 | 164.5 | 168 |
| BMI (kg.m-2) | 27.26 | 25.19 | 26.22 | 27.83 | 27.55 | 27.75 |
| Chin-thyroid distance (cm) | 5.7 | 6.1 | 5.9 | 5.9 | 6.75 | 6.1 |
Figure1Phantom of cone-beam computed tomography (CBCT) dosimetry. (a): A cylindrical head-size poly-methyl methacrylate (PMMA) phantom with five cylindrical slots for the insertion of an ionization chamber. (b): A schematic illustration of slot arrangement.
Entrance surface doses (mGy) for the eyes, parotid and thyroid glands for two maxillofacial imaging centers
| Center A | Center B | |||||||
|---|---|---|---|---|---|---|---|---|
| Min | Max | Mean | SD | Min | Max | Mean | SD | |
| Eyes | 2.22 | 2.88 | 2.57 | 0.25 | 0.18 | 0.50 | 0.35 | 0.12 |
| Parotid | 1.97 | 2.87 | 2.33 | 0.33 | 1.38 | 2.90 | 2.11 | 0.55 |
| Thyroid | 0.22 | 0.38 | 0.28 | 0.06 | 0.18 | 0.62 | 0.37 | 0.16 |
SD: Standard deviation
Exposure parameters and doses for two CBCT devices in phantom dosimetry
| Voltage (kV) | Amperage (mA) | Exposure time(s) | FOV (cm2) | dose (mGy) | |
|---|---|---|---|---|---|
| CBCT unit | 110 | Auto | Auto | 8×8 | 2.63 |
| NewTom VGi | 110 | Auto | Auto | 6×6 | 2.08 |
| 84 | 8 | 12 | 8×8 | 1.37 | |
| Planmeca Promax 3D | 84 | 12 | 12 | 8×8 | 2.36 |
| 84 | 16 | 12 | 8×8 | 3.24 | |
| . | 84 | 8 | 12 | 5×8 | 0.98 |
| 84 | 12 | 12 | 5×8 | 1.56 | |
| 84 | 16 | 12 | 5×8 | 2.33 |
FOV: field of view (height × diameter)