New head equivalent phantom for task and image performance evaluation representative for neurovascular procedures occurring in the Circle of Willis.

Ciprian N Ionita1, Brendan Loughran, Amit Jain, S N Swetadri Vasan, Daniel R Bednarek, Elad Levy, Adnan H Siddiqui, Kenneth V Snyder, L N Hopkins, Stephen Rudin.   

Abstract

Phantom equivalents of different human anatomical parts are routinely used for imaging system evaluation or dose calculations. The various recommendations on the generic phantom structure given by organizations such as the AAPM, are not always accurate when evaluating a very specific task. When we compared the AAPM head phantom containing 3 mm of aluminum to actual neuro-endovascular image guided interventions (neuro-EIGI) occurring in the Circle of Willis, we found that the system automatic exposure rate control (AERC) significantly underestimated the x-ray parameter selection. To build a more accurate phantom for neuro-EIGI, we reevaluated the amount of aluminum which must be included in the phantom. Human skulls were imaged at different angles, using various angiographic exposures, at kV's relevant to neuro-angiography. An aluminum step wedge was also imaged under identical conditions, and a correlation between the gray values of the imaged skulls and those of the aluminum step thicknesses was established. The average equivalent aluminum thickness for the skull samples for frontal projections in the Circle of Willis region was found to be about 13 mm. The results showed no significant changes in the average equivalent aluminum thickness with kV or mAs variation. When a uniform phantom using 13 mm aluminum and 15 cm acrylic was compared with an anthropomorphic head phantom the x-ray parameters selected by the AERC system were practically identical. These new findings indicate that for this specific task, the amount of aluminum included in the head equivalent must be increased substantially from 3 mm to a value of 13 mm.

Entities:  

Keywords:  Circle of Willis; Head equivalent phantom; digital subtracted angiography; fluoroscopy; uniform phantom

Year:  2012        PMID: 24027618      PMCID: PMC3767006          DOI: 10.1117/12.911351

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  11 in total

1.  A new tool for benchmarking cardiovascular fluoroscopes.

Authors:  S Balter; F A Heupler; P J Lin; M H Wondrow
Journal:  Catheter Cardiovasc Interv       Date:  2001-01       Impact factor: 2.692

2.  Development of a phantom for morphometric X-ray absorptiometry.

Authors:  J A Rea; G M Blake; I Fogelman
Journal:  Br J Radiol       Date:  2001-04       Impact factor: 3.039

3.  A patient-equivalent attenuation phantom for estimating patient exposures from automatic exposure controlled x-ray examinations of the abdomen and lumbo-sacral spine.

Authors:  B J Conway; J E Duff; T R Fewell; R J Jennings; L N Rothenberg; R C Fleischman
Journal:  Med Phys       Date:  1990 May-Jun       Impact factor: 4.071

4.  Region-of-Interest Micro-Angiographic Fluoroscope Detector Used in Aneurysm and Artery Stenosis Diagnoses and Treatment.

Authors:  Weiyuan Wang; Ciprian Ionita; Ying Huang; Bin Qu; Ashish Panse; Amit Jain; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2012-02-23

5.  Development and evaluation of a phantom for morphometric X-ray absorptiometry.

Authors:  S A Steel; J A Thorpe; R Walker; S Howey; C M Langton
Journal:  Osteoporos Int       Date:  1999       Impact factor: 4.507

6.  Use of the microangiographic fluoroscope for coiling of intracranial aneurysms.

Authors:  Mandy J Binning; David Orion; Parham Yashar; Sharon Webb; Ciprian N Ionita; Amit Jain; Stephen Rudin; L Nelson Hopkins; Adnan H Siddiqui; Elad I Levy
Journal:  Neurosurgery       Date:  2011-11       Impact factor: 4.654

7.  Patient dose and image quality in five standard x-ray examinations.

Authors:  R Havukainen; M Pirinen
Journal:  Med Phys       Date:  1993 May-Jun       Impact factor: 4.071

8.  Tissue-equivalent materials for construction of tomographic dosimetry phantoms in pediatric radiology.

Authors:  A K Jones; D E Hintenlang; W E Bolch
Journal:  Med Phys       Date:  2003-08       Impact factor: 4.071

9.  Modification of the NEMA XR21-2000 cardiac phantom for testing of imaging systems used in endovascular image guided interventions.

Authors:  C N Ionita; A Dohatcu; A Jain; C Keleshis; K R Hoffmann; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2009

10.  Comparison of six phantoms for entrance skin dose evaluation in 11 standard X-ray examinations.

Authors:  Gaetano Compagnone; Laura Pagan; Carlo Bergamini
Journal:  J Appl Clin Med Phys       Date:  2005-01-12       Impact factor: 2.102

View more
  7 in total

1.  Design considerations for a new, high resolution Micro-Angiographic Fluoroscope based on a CMOS sensor (MAF-CMOS).

Authors:  Brendan Loughran; S N Swetadri Vasan; Vivek Singh; Ciprian N Ionita; Amit Jain; Daniel R Bednarek; Albert Titus; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2013-03-06

2.  Anti-scatter grid artifact elimination for high resolution x-ray imaging CMOS detectors.

Authors:  R Rana; V Singh; A Jain; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-03-18

3.  A novel Region of Interest (ROI) imaging technique for biplane imaging in interventional suites: high-resolution small field-of-view imaging in the frontal plane and dose-reduced, large field-of-view standard-resolution imaging in the lateral plane.

Authors:  S N Swetadri Vasan; C Ionita; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19

4.  Limitations of anti-scatter grids when used with high resolution image detectors.

Authors:  V Singh; A Jain; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19

5.  Detector system comparison using relative CNR for specific imaging tasks related to neuro-endovascular image-guided interventions (neuro-EIGIs).

Authors:  Brendan Loughran; S N Swetadri Vasan; Vivek Singh; Ciprian N Ionita; Amit Jain; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19

6.  Chemical Modifications of Porous Shape Memory Polymers for Enhanced X-ray and MRI Visibility.

Authors:  Grace K Fletcher; Landon D Nash; Lance M Graul; Lindy K Jang; Scott M Herting; Matthew D Wilcox; Tyler J Touchet; Ana Katarina Sweatt; Mary P McDougall; Steven M Wright; Duncan J Maitland
Journal:  Molecules       Date:  2020-10-13       Impact factor: 4.411

7.  Increased X-ray Visualization of Shape Memory Polymer Foams by Chemical Incorporation of Iodine Motifs.

Authors:  Landon D Nash; Mary Beth Browning Monroe; Yong-Hong Ding; Kendal P Ezell; Anthony J Boyle; Ramanathan Kadirvel; David F Kallmes; Duncan J Maitland
Journal:  Polymers (Basel)       Date:  2017-08-20       Impact factor: 4.329

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.