Limitations of anti-scatter grids when used with high resolution image detectors.

V Singh1, A Jain1, D R Bednarek1, S Rudin1.   

Abstract

Anti-scatter grids are used in fluoroscopic systems to improve image quality by absorbing scattered radiation. A stationary Smit Rontgen X-ray grid (line density: 70 lines/cm, grid ratio: 13:1) was used with a flat panel detector (FPD) of pixel size 194 micron and a high-resolution CMOS detector, the Dexela 1207 with pixel size of 75 microns. To investigate the effectiveness of the grid, a simulated artery block was placed in a modified uniform frontal head phantom and imaged with both the FPD and the Dexela for an approximately 15 × 15 cm field of view (FOV). The contrast improved for both detectors with the grid. The contrast-to-noise ratio (CNR) does not increase as much in the case of the Dexela as it improves in the case of the FPD. Since the total noise in a single frame increases substantially for the Dexela compared to the FPD when the grid is used, the CNR is degraded. The increase in the quantum noise per frame would be similar for both detectors when the grid is used due to the attenuation of radiation, but the fixed pattern noise caused by the grid was substantially higher for the Dexela compared to the FPD and hence caused a severe reduction of CNR. Without further corrective methods this grid should not be used with high-resolution fluoroscopic detectors because the CNR does not improve significantly and the visibility of low contrast details may be reduced. Either an anti-scatter grid of different design or an additional image processing step when using a similar grid would be required to deal with the problem of scatter for high resolution detectors and the structured noise of the grid pattern.

Entities:  

Keywords:  CNR; anti-scatter grid; high-resolution detector; image quality; scatter

Year:  2014        PMID: 25309101      PMCID: PMC4189125          DOI: 10.1117/12.2043063

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  8 in total

Review 1.  Contrast and scatter in x-ray imaging.

Authors:  G T Barnes
Journal:  Radiographics       Date:  1991-03       Impact factor: 5.333

2.  New head equivalent phantom for task and image performance evaluation representative for neurovascular procedures occurring in the Circle of Willis.

Authors:  Ciprian N Ionita; Brendan Loughran; Amit Jain; S N Swetadri Vasan; Daniel R Bednarek; Elad Levy; Adnan H Siddiqui; Kenneth V Snyder; L N Hopkins; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2012-02-23

3.  Grids or air gaps for scatter reduction in digital radiography: a model calculation.

Authors:  U Neitzel
Journal:  Med Phys       Date:  1992 Mar-Apr       Impact factor: 4.071

4.  A study of grid artifacts formation and elimination in computed radiographic images.

Authors:  Chih-Yang Lin; Wen-Jeng Lee; Shyh-Jye Chen; Ching-Hwa Tsai; Jei-Han Lee; Chia-Hung Chang; Yu-Tai Ching
Journal:  J Digit Imaging       Date:  2006-12       Impact factor: 4.056

5.  Grid line artifact formation: A comprehensive theory.

Authors:  David M Gauntt; Gary T Barnes
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

6.  Grid artifact reduction for direct digital radiography detectors based on rotated stationary grids with homomorphic filtering.

Authors:  Dong Sik Kim; Sanggyun Lee
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

7.  Artifacts produced by moving grids.

Authors:  D R Bednarek; S Rudin; R Wong
Journal:  Radiology       Date:  1983-04       Impact factor: 11.105

8.  Modification of the NEMA XR21-2000 cardiac phantom for testing of imaging systems used in endovascular image guided interventions.

Authors:  C N Ionita; A Dohatcu; A Jain; C Keleshis; K R Hoffmann; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2009
  8 in total
  4 in total

1.  Scatter reduction for high resolution image detectors with a region of interest attenuator.

Authors:  Amit Jain; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19

2.  Anti-scatter grid artifact elimination for high resolution x-ray imaging CMOS detectors.

Authors:  R Rana; V Singh; A Jain; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-03-18

3.  Real time implementation of anti-scatter grid artifact elimination method for high resolution x-ray imaging CMOS detectors using Graphics Processing Units (GPUs).

Authors:  R Rana; S V Setlur Nagesh; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-09

4.  Scatter estimation and removal of anti-scatter grid-line artifacts from anthropomorphic head phantom images taken with a high resolution image detector.

Authors:  R Rana; A Jain; A Shankar; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-31
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.