Design considerations for a new, high resolution Micro-Angiographic Fluoroscope based on a CMOS sensor (MAF-CMOS).

Brendan Loughran1, S N Swetadri Vasan1, Vivek Singh1, Ciprian N Ionita1, Amit Jain1, Daniel R Bednarek1, Albert Titus1, Stephen Rudin1.   

Abstract

The detectors that are used for endovascular image-guided interventions (EIGI), particularly for neurovascular interventions, do not provide clinicians with adequate visualization to ensure the best possible treatment outcomes. Developing an improved x-ray imaging detector requires the determination of estimated clinical x-ray entrance exposures to the detector. The range of exposures to the detector in clinical studies was found for the three modes of operation: fluoroscopic mode, high frame-rate digital angiographic mode (HD fluoroscopic mode), and DSA mode. Using these estimated detector exposure ranges and available CMOS detector technical specifications, design requirements were developed to pursue a quantum limited, high resolution, dynamic x-ray detector based on a CMOS sensor with 50 μm pixel size. For the proposed MAF-CMOS, the estimated charge collected within the full exposure range was found to be within the estimated full well capacity of the pixels. Expected instrumentation noise for the proposed detector was estimated to be 50-1,300 electrons. Adding a gain stage such as a light image intensifier would minimize the effect of the estimated instrumentation noise on total image noise but may not be necessary to ensure quantum limited detector operation at low exposure levels. A recursive temporal filter may decrease the effective total noise by 2 to 3 times, allowing for the improved signal to noise ratios at the lowest estimated exposures despite consequent loss in temporal resolution. This work can serve as a guide for further development of dynamic x-ray imaging prototypes or improvements for existing dynamic x-ray imaging systems.

Entities:  

Keywords:  CMOS; MAF; ROI; angiography; detector design; fluoroscopy; neurovascular interventions; x-ray imaging

Year:  2013        PMID: 24353389      PMCID: PMC3864963          DOI: 10.1117/12.2006430

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  9 in total

1.  New head equivalent phantom for task and image performance evaluation representative for neurovascular procedures occurring in the Circle of Willis.

Authors:  Ciprian N Ionita; Brendan Loughran; Amit Jain; S N Swetadri Vasan; Daniel R Bednarek; Elad Levy; Adnan H Siddiqui; Kenneth V Snyder; L N Hopkins; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2012-02-23

2.  Empirical electro-optical and x-ray performance evaluation of CMOS active pixels sensor for low dose, high resolution x-ray medical imaging.

Authors:  C D Arvanitis; S E Bohndiek; G Royle; A Blue; H X Liang; A Clark; M Prydderch; R Turchetta; R Speller
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

3.  Signal and noise transfer properties of CMOS based active pixel flat panel imager coupled to structured CsI:Tl.

Authors:  C D Arvanitis; S E Bohndiek; J Blakesley; A Olivo; R D Speller
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

4.  A practical exposure-equivalent metric for instrumentation noise in x-ray imaging systems.

Authors:  G K Yadava; A T Kuhls-Gilcrist; S Rudin; V K Patel; K R Hoffmann; D R Bednarek
Journal:  Phys Med Biol       Date:  2008-08-22       Impact factor: 3.609

5.  SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes.

Authors:  G Poludniowski; G Landry; F DeBlois; P M Evans; F Verhaegen
Journal:  Phys Med Biol       Date:  2009-09-01       Impact factor: 3.609

6.  A theoretical and experimental evaluation of the microangiographic fluoroscope: A high-resolution region-of-interest x-ray imager.

Authors:  Amit Jain; D R Bednarek; Ciprian Ionita; S Rudin
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

7.  Endovascular coil embolization of a very small ruptured aneurysm using a novel microangiographic technique: technical note.

Authors:  Peter Kan; Parham Yashar; Ciprian N Ionita; Amit Jain; Stephen Rudin; Elad I Levy; Adnan H Siddiqui
Journal:  J Neurointerv Surg       Date:  2012-01-21       Impact factor: 5.836

8.  Region-of-Interest Micro-Angiographic Fluoroscope Detector Used in Aneurysm and Artery Stenosis Diagnoses and Treatment.

Authors:  Weiyuan Wang; Ciprian Ionita; Ying Huang; Bin Qu; Ashish Panse; Amit Jain; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2012-02-23

9.  Use of the microangiographic fluoroscope for coiling of intracranial aneurysms.

Authors:  Mandy J Binning; David Orion; Parham Yashar; Sharon Webb; Ciprian N Ionita; Amit Jain; Stephen Rudin; L Nelson Hopkins; Adnan H Siddiqui; Elad I Levy
Journal:  Neurosurgery       Date:  2011-11       Impact factor: 4.654

  9 in total
  5 in total

1.  Quantitative comparison using Generalized Relative Object Detectability (G-ROD) metrics of an amorphous selenium detector with high resolution Microangiographic Fluoroscopes (MAF) and standard flat panel detectors (FPD).

Authors:  M Russ; A Shankar; A Jain; S V Setlur Nagesh; C N Ionita; C Scott; K S Karim; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-22

2.  Implementation of material decomposition using an EMCCD and CMOS-based micro-CT system.

Authors:  Alexander R Podgorsak; Sv Setlur Nagesh; Daniel R Bednarek; Stephen Rudin; Ciprian N Ionita
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-13

3.  Evaluation of the microangiographic fluoroscope (MAF) using generalized system performance metrics.

Authors:  Amit Jain; Daniel R Bednarek; Stephen Rudin
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

4.  Workflow for the use of a high-resolution image detector in endovascular interventional procedures.

Authors:  R Rana; B Loughran; S N Swetadri Vasan; L Pope; C N Ionita; A Siddiqui; N Lin; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19

5.  Detector system comparison using relative CNR for specific imaging tasks related to neuro-endovascular image-guided interventions (neuro-EIGIs).

Authors:  Brendan Loughran; S N Swetadri Vasan; Vivek Singh; Ciprian N Ionita; Amit Jain; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.