Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Bruno Carvalho de Vasconcelos1, Michelli de Medeiros Bueno, Suyane Maria Luna-Cruz, Marco Antonio Hungaro Duarte, Carlos Augusto de Oliveira Fernandes.   

Abstract

OBJECTIVE: The aim of this study was to evaluate, ex vivo, the precision of five electronic root canal length measurement devices (ERCLMDs) with different operating systems: the Root ZX, Mini Apex Locator, Propex II, iPex, and RomiApex A-15, and the possible influence of the positioning of the instrument tips short of the apical foramen.
MATERIAL AND METHODS: Forty-two mandibular bicuspids had their real canal lengths (RL) previously determined. Electronic measurements were performed 1.0 mm short of the apical foramen (-1.0), followed by measurements at the apical foramen (0.0). The data resulting from the comparison of the ERCLMD measurements and the RL were evaluated by the Wilcoxon and Friedman tests at a significance level of 5%.
RESULTS: Considering the measurements performed at 0.0 and -1.0, the precision rates for the ERCLMDs were: 73.5% and 47.1% (Root ZX), 73.5% and 55.9% (Mini Apex Locator), 67.6% and 41.1% (Propex II), 61.7% and 44.1% (iPex), and 79.4% and 44.1% (RomiApex A-15), respectively, considering ±0.5 mm of tolerance. Regarding the mean discrepancies, no differences were observed at 0.0; however, in the measurements at -1.0, the iPex, a multi-frequency ERCLMD, had significantly more discrepant readings short of the apical foramen than the other devices, except for the Propex II, which had intermediate results. When the ERCLMDs measurements at -1.0 were compared with those at 0.0, the Propex II, iPex and RomiApex A-15 presented significantly higher discrepancies in their readings.
CONCLUSIONS: Under the conditions of the present study, all the ERCLMDs provided acceptable measurements at the 0.0 position. However, at the -1.0 position, the ERCLMDs had a lower precision, with statistically significant differences for the Propex II, iPex, and RomiApex A-15.

Entities:  

Mesh:

Year:  2013        PMID: 23739852      PMCID: PMC3881880          DOI: 10.1590/1678-7757201302188

Source DB:  PubMed          Journal:  J Appl Oral Sci        ISSN: 1678-7757            Impact factor:   2.698


  30 in total

1.  An in vivo comparison of two frequency-based electronic apex locators.

Authors:  Aaron R Welk; J Craig Baumgartner; J Gordon Marshall
Journal:  J Endod       Date:  2003-08       Impact factor: 4.171

Review 2.  The fundamental operating principles of electronic root canal length measurement devices.

Authors:  M H Nekoofar; M M Ghandi; S J Hayes; P M H Dummer
Journal:  Int Endod J       Date:  2006-08       Impact factor: 5.264

3.  In vivo comparison of working length determination with two electronic apex locators.

Authors:  K T Wrbas; A A Ziegler; M J Altenburger; J F Schirrmeister
Journal:  Int Endod J       Date:  2007-02       Impact factor: 5.264

4.  Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters.

Authors:  A K Ebrahim; R Wadachi; H Suda
Journal:  Aust Dent J       Date:  2006-09       Impact factor: 2.291

5.  Influence of embedding media on the assessment of electronic apex locators.

Authors:  Járcio V Baldi; Fausto R Victorino; Ricardo A Bernardes; Ivaldo G de Moraes; Clóvis M Bramante; Roberto B Garcia; Norberti Bernardineli
Journal:  J Endod       Date:  2007-02-22       Impact factor: 4.171

6.  Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30.

Authors:  Ricardo Affonso Bernardes; Marco A H Duarte; Bruno C Vasconcelos; Ivaldo G Moraes; Norberti Bernardineli; Roberto B Garcia; Jarcio V Baldi; Fausto R Victorino; Clovis M Bramante
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2007-08-20

7.  Ex vivo comparison of the accuracy of Root ZX II in detecting apical constriction using different meter's reading.

Authors:  Marco Aurélio Versiani; Bianca Palma Santana; Cristiane Melo Caram; Elizeu Alvaro Pascon; Cássio José Alves de Souza; João Carlos Gabrielli Biffi
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2009-05-22

8.  Influence of cervical preflaring on determination of apical file size in mandibular molars: SEM analysis.

Authors:  Marcia da Silva Schmitz; Roberto Santos; Alexandre Capelli; Marcos Jacobovitz; Júlio César Emboava Spanó; Jesus Djalma Pécora
Journal:  Braz Dent J       Date:  2008

9.  Effect of pre-flaring and file size on the accuracy of two electronic apex locators.

Authors:  Manoel Brito-Júnior; Carla Cristina Camilo; Gil Moreira-Júnior; Jesus Djalma Pecora; Manoel Damião Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

10.  Evaluation of the apical infiltration after root canal disruption and obturation.

Authors:  João Eduardo Gomes-Filho; Renato Nicolás Hopp; Pedro Felício Estrada Bernabé; Mauro Juvenal Nery; José Arlindo Otoboni Filho; Elói Dezan Júnior
Journal:  J Appl Oral Sci       Date:  2008 Sep-Oct       Impact factor: 2.698

View more
  7 in total

1.  In vitro comparison between apex locators, direct and radiographic techniques for determining the root canal length in primary teeth.

Authors:  L Z Adriano; J C Barasuol; M Cardoso; M Bolan
Journal:  Eur Arch Paediatr Dent       Date:  2019-01-02

2.  Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study.

Authors:  Franziska Haupt; M Hülsmann
Journal:  Clin Oral Investig       Date:  2018-03-07       Impact factor: 3.573

3.  In vivo assessment of accuracy of Propex II, Root ZX II, and radiographic measurements for location of the major foramen.

Authors:  Fernanda Garcia Tampelini; Marcelo Santos Coelho; Marcos de Azevêdo Rios; Carlos Eduardo Fontana; Daniel Guimarães Pedro Rocha; Sergio Luiz Pinheiro; Carlos Eduardo da Silveira Bueno
Journal:  Restor Dent Endod       Date:  2017-05-16

4.  A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.

Authors:  Mugem Asli Gurel; Bagdagul Helvacioglu Kivanc; Adil Ekici
Journal:  J Istanb Univ Fac Dent       Date:  2017-01-02

5.  Root ZX Electronic Foramen Locator: An Ex Vivo Study of Its Three Models' Precision and Reproducibility.

Authors:  Bernardo Almeida Aguiar; Rafael Santos Reinaldo; Luciana Maria Arcanjo Frota; Mônica Sampaio do Vale; Bruno Carvalho de Vasconcelos
Journal:  Int J Dent       Date:  2017-03-07

6.  A comparative clinical evaluation of accuracy of six apex locators with intraoral periapical radiograph in multirooted teeth: An in vivo study.

Authors:  Prahlad A Saraf; P Ratnakar; Thimmanagowda N Patil; Raghavendra Penukonda; Laxmikant Kamatagi; Sneha S Vanaki
Journal:  J Conserv Dent       Date:  2017 Jul-Aug

7.  An in vivo comparison of accuracy of two electronic apex locators in determining working length using stainless steel and nickel titanium files.

Authors:  Sanjeeb Chaudhary; Archana Gharti; Bhawana Adhikari
Journal:  Clin Cosmet Investig Dent       Date:  2018-05-22
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.