Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters.

A K Ebrahim1, R Wadachi, H Suda.   

Abstract

BACKGROUND: The aim of this ex vivo study was to evaluate the accuracy of four electronic apex locators (EALs) to determine the working length in teeth with various foramen diameters. Our previous study revealed that electronically measured canal length was influenced by the root canal diameter. It is not known whether foramen size would interfere with the reading accuracy of an EAL.
METHODS: A total of 36 extracted human lower single rooted premolar teeth were divided into four groups of nine teeth each. In groups A, B and C, the root canals were instrumented using #10-80, #10-100 and #10-120 K-files, and the tip of size #80, #100 and #120 K-files were permitted to pass through the apical foramen to a length of 1mm, respectively. In group D, the teeth were instrumented using #10-140 K-files and the tip of #140 K-file was permitted to pass through the apical foramen to a length of 5 mm. Thus, the average apical foramen diameters in groups A, B, C and D were approximately 0.82 mm, 1.02 mm, 1.22mm and 1.5 mm, respectively. The teeth were then mounted in 1% agar and four EALs were used: Root ZX, Foramatron D10, Apex NRG and Apit 7. For electronic measurement, sizes #10 and #80, #10 and #100, #10 and #120, and #10 and #140 K-files were used for groups A, B, C and D, respectively. During electronic measurement the canals were flushed with 6% sodium hypochlorite solution.
RESULTS: Three-way ANOVA and Bonferroni test showed that EAL, file size and foramen size all had a significant influence on the measurement error (P<0.0001), with all the interactions between these three factors being significant (P<0.0001).
CONCLUSIONS: The four EALs were unreliable to determine the working length of teeth with a wide apical foramen, when using a small size file. The Root ZX and Foramatron D10 showed significantly better scores than the other two EALs and may be more reliable to determine the working length of teeth with a wide apical foramen, if a tight-fit file is used.

Entities:  

Mesh:

Year:  2006        PMID: 17037894     DOI: 10.1111/j.1834-7819.2006.tb00439.x

Source DB:  PubMed          Journal:  Aust Dent J        ISSN: 0045-0421            Impact factor:   2.291


  9 in total

1.  Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study.

Authors:  Franziska Haupt; M Hülsmann
Journal:  Clin Oral Investig       Date:  2018-03-07       Impact factor: 3.573

2.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

3.  Effect of pre-flaring and file size on the accuracy of two electronic apex locators.

Authors:  Manoel Brito-Júnior; Carla Cristina Camilo; Gil Moreira-Júnior; Jesus Djalma Pecora; Manoel Damião Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

4.  In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters.

Authors:  Ugur Aydin; Emrah Karataslioglu; Fatih Aksoy; Cihan Yildirim
Journal:  J Conserv Dent       Date:  2015 Jan-Feb

5.  In vitro comparison of working length determination using three different electronic apex locators.

Authors:  Alper Kuştarci; Dilara Arslan; Demet Altunbaş
Journal:  Dent Res J (Isfahan)       Date:  2014-09

6.  Influence of root canal curvature on the accuracy of an electronic apex locator: An in vitro study.

Authors:  Lekha Santhosh; Pooja Raiththa; Srirekha Aswathanarayana; Srinivas Panchajanya; Jayakumar Thimmaraya Reddy; Shwetha Rajanna Susheela
Journal:  J Conserv Dent       Date:  2014-11

7.  Influence of embedding media on the accuracy of working length determination by means of apex locator: an ex vivo study.

Authors:  Thomas Gerhard Wolf; Anna Krauß-Mironjuk; Richard Johannes Wierichs; Benjamín Briseño-Marroquín
Journal:  Sci Rep       Date:  2021-02-08       Impact factor: 4.379

8.  Accuracy and stability of electronic apex locator length measurements in root canals with wide apical foramen: an ex vivo study.

Authors:  Maayan Shacham; Avi Levin; Avi Shemesh; Alex Lvovsky; Joe Ben Itzhak; Michael Solomonov
Journal:  BDJ Open       Date:  2020-11-17

9.  An In Vitro Comparison of Propex II Apex Locator to Standard Radiographic Method.

Authors:  Kalyan Vinayak Chakravarthy Pishipati
Journal:  Iran Endod J       Date:  2013-08-01
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.