Effect of pre-flaring and file size on the accuracy of two electronic apex locators.

Manoel Brito-Júnior1, Carla Cristina Camilo, Gil Moreira-Júnior, Jesus Djalma Pecora, Manoel Damião Sousa-Neto.   

Abstract

OBJECTIVE: This ex vivo study evaluated the effect of pre-flaring and file size on the accuracy of the Root ZX and Novapex electronic apex locators (EALs).
MATERIAL AND METHODS: The actual working length (WL) was set 1 mm short of the apical foramen in the palatal root canals of 24 extracted maxillary molars. The teeth were embedded in an alginate mold, and two examiners performed the electronic measurements using #10, #15, and #20 K-files. The files were inserted into the root canals until the "0.0" or "APEX" signals were observed on the LED or display screens for the Novapex and Root ZX, respectively, retracting to the 1.0 mark. The measurements were repeated after the preflaring using the S1 and SX Pro-Taper instruments. Two measurements were performed for each condition and the means were used. Intra-class correlation coefficients (ICCs) were calculated to verify the intra- and inter-examiner agreement. The mean differences between the WL and electronic length values were analyzed by the three-way ANOVA test (p<0.05).
RESULTS: ICCs were high (>0.8) and the results demonstrated a similar accuracy for both EALs (p>0.05). Statistically significant accurate measurements were verified in the pre-flared canals, except for the Novapex using a #20 K-file.
CONCLUSIONS: The tested EALs showed acceptable accuracy, whereas the pre-flaring procedure revealed a more significant effect than the used file size.

Entities:  

Mesh:

Year:  2012        PMID: 23138740      PMCID: PMC3881801          DOI: 10.1590/s1678-77572012000500008

Source DB:  PubMed          Journal:  J Appl Oral Sci        ISSN: 1678-7757            Impact factor:   2.698


  33 in total

1.  Effect of using electronic apex locators on selected endodontic treatment parameters.

Authors:  A F Fouad; L C Reid
Journal:  J Endod       Date:  2000-06       Impact factor: 4.171

2.  The accuracy of three different electronic root canal measuring devices: an in vitro evaluation.

Authors:  Ali Cemal Tinaz; Murat Maden; Cumhur Aydin; Emin Türköz
Journal:  J Oral Sci       Date:  2002-06       Impact factor: 1.556

Review 3.  Electronic apex locators.

Authors:  M P J Gordon; N P Chandler
Journal:  Int Endod J       Date:  2004-07       Impact factor: 5.264

4.  Influence of pulp condition on the accuracy of an electronic foramen locator in posterior teeth: an in vivo study.

Authors:  Daniel Renner; Renata Grazziotin-Soares; Giulio Gavini; Fernando Branco Barletta
Journal:  Braz Oral Res       Date:  2012 Mar-Apr

5.  An in vivo radiographic evaluation of the accuracy of Apex and iPex electronic Apex locators.

Authors:  Laura Paludo; Sophia Lopes de Souza; Marcus Vinícius Reis Só; Ricardo Abreu da Rosa; Fabiana Vieira Vier-Pelisser; Marco Antônio Húngaro Duarte
Journal:  Braz Dent J       Date:  2012

6.  Apical limit of root canal instrumentation and obturation, part 2. A histological study.

Authors:  D Ricucci; K Langeland
Journal:  Int Endod J       Date:  1998-11       Impact factor: 5.264

7.  In vitro evaluation of the ability of three apex locators to determine the working length during retreatment.

Authors:  Fernando Goldberg; Benjamín Briseño Marroquín; Santiago Frajlich; Cristian Dreyer
Journal:  J Endod       Date:  2005-09       Impact factor: 4.171

8.  Influence of cervical preflaring on apical file size determination.

Authors:  J D Pecora; A Capelli; D M Z Guerisoli; J C E Spanó; C Estrela
Journal:  Int Endod J       Date:  2005-07       Impact factor: 5.264

9.  Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex.

Authors:  G Plotino; N M Grande; L Brigante; B Lesti; F Somma
Journal:  Int Endod J       Date:  2006-05       Impact factor: 5.264

10.  Electronic length measurement using small and large files in enlarged canals.

Authors:  H Q Nguyen; A Y Kaufman; R C Komorowski; S Friedman
Journal:  Int Endod J       Date:  1996-11       Impact factor: 5.264

View more
  3 in total

1.  The accuracy of two electronic apex locators on effect of preflaring and file size: An in vitro study.

Authors:  V Saritha; H Raghu; Twino H Kumar; Shrishail Totad; Laxmikant Kamatagi; Prahlad A Saraf
Journal:  J Conserv Dent       Date:  2021-07-05

2.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

3.  The Precision of Propex Pixi with Different Instruments and Coronal Preflaring Procedures.

Authors:  Inês Ferreira; Ana Cristina Braga; Irene Pina-Vaz
Journal:  Eur Endod J       Date:  2019-07-10
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.