Literature DB >> 17229119

In vivo comparison of working length determination with two electronic apex locators.

K T Wrbas1, A A Ziegler, M J Altenburger, J F Schirrmeister.   

Abstract

AIM: To compare the accuracy of two electronic apex locators (EALs) in the same teeth in vivo.
METHODOLOGY: The working lengths in 20 teeth with a single canal were determined with two different EALs (Root ZX; J. Morita Corporation, Tokyo, Japan and Raypex 5 VDW, Munich, Germany) before extraction. When the first EAL was used the files were advanced until the display indicated the 'apical constriction'. The files were then fixed in removable and replaceable light curing composite patterns. The procedure was repeated in the same tooth with the second EAL and a different file. The teeth were then extracted and the apical 4 mm of the root canals were exposed. After that the apical parts with the repositioned files in the canals were digitally photographed under a light microscope. On the images the minor diameter and the major foramen of each sample were marked and the respective distances of the file tips from these positions were measured with a computer program. Subsequently the values of the two groups of EALs were compared using a paired sample t-test.
RESULTS: The minor foramen was located within the limits of +/-0.5 mm in 75% of the cases with the Root ZX and in 80% of the cases with Raypex 5. The paired sample t-test showed no significant difference between the EALs regarding determination of the minor foramen.
CONCLUSION: The use of EALs is a reliable method for determining working length. The differences between the two EALs were not statistically significant.

Mesh:

Year:  2007        PMID: 17229119     DOI: 10.1111/j.1365-2591.2006.01199.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  15 in total

1.  Clinical evaluation of Root ZX II electronic apex locator in primary teeth.

Authors:  S Saritha; K S Uloopi; C Vinay; R Chandra Sekhar; V V Rao
Journal:  Eur Arch Paediatr Dent       Date:  2012-02

2.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

3.  Comparison of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study.

Authors:  J Paras Mull; Vinutha Manjunath; Mk Manjunath
Journal:  J Conserv Dent       Date:  2012-04

4.  Comparison of third generation versus fourth generation electronic apex locators in detecting apical constriction: An in vivo study.

Authors:  D V Swapna; Akash Krishna; Anand C Patil; K Rashmi; Veena S Pai; M A Ranjini
Journal:  J Conserv Dent       Date:  2015 Jul-Aug

5.  Evaluation of Dentaport ZX and Raypex 6 electronic apex locators: an in vivo study.

Authors:  S Moscoso; K Pineda; J Basilio; C Alvarado; M Roig; F Duran-Sindreu
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2014-03-01

6.  In vitro comparison of working length determination using three different electronic apex locators.

Authors:  Alper Kuştarci; Dilara Arslan; Demet Altunbaş
Journal:  Dent Res J (Isfahan)       Date:  2014-09

7.  Evaluation of Conventional Radiography and an Electronic Apex Locator in Determining the Working Length in C-shaped Canals.

Authors:  Hamid Jafarzadeh; Masoud Beyrami; Maryam Forghani
Journal:  Iran Endod J       Date:  2017

8.  A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.

Authors:  Mugem Asli Gurel; Bagdagul Helvacioglu Kivanc; Adil Ekici
Journal:  J Istanb Univ Fac Dent       Date:  2017-01-02

9.  A comparative evaluation of accuracy of three electronic apex locators using histological section as gold standard: An ex vivo study.

Authors:  Divya Saxena; Suparna Ganguly Saha; Anuj Bharadwaj; Neelam Vijaywargiya; Sandeep Dubey; Shubham Kala
Journal:  J Conserv Dent       Date:  2017 Jul-Aug

10.  A comparative clinical evaluation of accuracy of six apex locators with intraoral periapical radiograph in multirooted teeth: An in vivo study.

Authors:  Prahlad A Saraf; P Ratnakar; Thimmanagowda N Patil; Raghavendra Penukonda; Laxmikant Kamatagi; Sneha S Vanaki
Journal:  J Conserv Dent       Date:  2017 Jul-Aug
View more