Literature DB >> 17703968

Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30.

Ricardo Affonso Bernardes1, Marco A H Duarte, Bruno C Vasconcelos, Ivaldo G Moraes, Norberti Bernardineli, Roberto B Garcia, Jarcio V Baldi, Fausto R Victorino, Clovis M Bramante.   

Abstract

OBJECTIVE: The aim of this study was to perform a comparative analysis of precision of 3 apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30. STUDY
DESIGN: Forty extracted single-rooted human teeth were selected. After endodontic access, measurement of the anatomical tooth length was visually performed by insertion of a K-file size 10 until its tip could be observed at the apical foramen with aid of a light microscope at 8x magnification. Following, the teeth were placed in a plastic box with alginate and electronically measured with the 3 apex locators at 1-mm short of the apical foramen.
RESULTS: The results revealed a precision rate of 97.5% for Root ZX, 95% for Elements Diagnostic Unit and Apex Locator, and 92.5% for RomiAPEX D-30, with no statistically significant difference between them (chi-square test, P > .05).
CONCLUSION: The results confirm that all these electronic devices can accurately determine the root canal length within 1 mm from the apical constriction.

Entities:  

Mesh:

Year:  2007        PMID: 17703968     DOI: 10.1016/j.tripleo.2007.05.016

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  10 in total

1.  Comparison of Radiovisiography, an Apex Locator and an Integrated Endomotor-inbuilt Apex Locator in Primary Teeth Endometrics.

Authors:  Suleman Abbas Khan; Rinee Khanna; Saumya Navit; Seema Jabeen; Nishi Grover; Saurabh Pramanik
Journal:  Int J Clin Pediatr Dent       Date:  2022

2.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

3.  In vivo accuracy of conventional and digital radiographic methods in confirming root canal working length determination by Root ZX.

Authors:  Fernando Accorsi Orosco; Norberti Bernardineli; Roberto Brandão Garcia; Clovis Monteiro Bramante; Marco Antonio Húngaro Duarte; Ivaldo Gomes de Moraes
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

4.  Effect of pre-flaring and file size on the accuracy of two electronic apex locators.

Authors:  Manoel Brito-Júnior; Carla Cristina Camilo; Gil Moreira-Júnior; Jesus Djalma Pecora; Manoel Damião Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

5.  In vivo assessment of the impedance ratio method used in electronic foramen locators.

Authors:  Marcos V H Rambo; Humberto R Gamba; Gustavo B Borba; Joaquim M Maia; Carlos A S Ramos
Journal:  Biomed Eng Online       Date:  2010-09-06       Impact factor: 2.819

6.  Root ZX Electronic Foramen Locator: An Ex Vivo Study of Its Three Models' Precision and Reproducibility.

Authors:  Bernardo Almeida Aguiar; Rafael Santos Reinaldo; Luciana Maria Arcanjo Frota; Mônica Sampaio do Vale; Bruno Carvalho de Vasconcelos
Journal:  Int J Dent       Date:  2017-03-07

7.  In vitro detection of simulated apical root perforation with two electronic apex locators.

Authors:  Kiumars Nazari Moghaddam; Shahrzad Nazari; Leila Shakeri; Kiamars Honardar; Farshid Mirmotalebi
Journal:  Iran Endod J       Date:  2010-02-20

8.  Working Length Determination of Root Canal of Young Permanent Tooth: An In vitro Study.

Authors:  A Diwanji; As Rathore; R Arora; V Dhar; A Madhusudan; J Doshi
Journal:  Ann Med Health Sci Res       Date:  2014-07

9.  Evaluation of complications of root canal treatment performed by undergraduate dental students.

Authors:  Mothanna K AlRahabi
Journal:  Libyan J Med       Date:  2017-12       Impact factor: 1.657

10.  Comparative evaluation of three methods to measure working length - Manual tactile sensation, digital radiograph, and multidetector computed tomography: An in vitro study.

Authors:  Kaushik Dutta; Priti D Desai; Utapal Kumar Das; Soumayabrata Sarkar
Journal:  J Conserv Dent       Date:  2017 Mar-Apr
  10 in total