Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study.

Franziska Haupt1, M Hülsmann2.   

Abstract

OBJECTIVES: To assess the consistency of electronic determination of endodontic working length obtained from four identical electronic root canal length measurement devices (ERCLMD) from five different types of ERCLMD each under various in vitro conditions.
MATERIALS AND METHODS: Eight extracted teeth, seven single-rooted teeth, and one molar were accessed. Root length was measured and instruments were inserted and fixed with the tip placed beyond and short of the apical constriction, in roots with an artificial perforation or an open apex. Devices tested were Root ZX (Morita, Kyoto, Japan), Dentaport ZX (Morita), Apex ID (SybronEndo, Glendora, USA), ProPex II (Dentsply Maillefer, Victoria, Australia), and Raypex 6 (VDW-Antaeos, Munich, Germany). Teeth were irrigated with different solutions (NaOCl, EDTA, CHX). ERCLMDs were connected and measurements were recorded. Consistency was classified by the scores 0-4. Comparisons were carried out using the Kruskal-Wallis test (α = 0.05). For multiple testing, the level of significance was adjusted and analysis was performed using the Mann-Whitney U test.
RESULTS: Among the five types of ERCLMD, Apex ID and Raypex 6 showed the highest consistency. There was no statistically significant difference between the settings and conditions. Raypex 6 showed the highest consistency for measurements in case of a perforation.
CONCLUSION: Different devices from one type of ERCLMD show a high consistency. Nevertheless, general statements on the accuracy of one type of ERCLMD taken from studies investigating only one device per type should be drawn with caution. CLINICAL RELEVANCE: The study refers to the reliability and repeatability of determination of endodontic working length by using ERCLMDs.

Entities:  

Keywords:  Electronic root canal length measurement device; Intra-device consistency; Open apex; Perforation; Working length

Mesh:

Year:  2018        PMID: 29516188     DOI: 10.1007/s00784-018-2404-3

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  25 in total

1.  The influence of sodium hypochlorite irrigation on the accuracy of the Root ZX electronic apex locator.

Authors:  W Anthony Meares; H Robert Steiman
Journal:  J Endod       Date:  2002-08       Impact factor: 4.171

2.  Comparison of in vivo and in vitro readings when testing the accuracy of the Root ZX apex locator.

Authors:  Fernando Duran-Sindreu; Eva Stöber; Montse Mercadé; Jorge Vera; Marc Garcia; Rufino Bueno; Miguel Roig
Journal:  J Endod       Date:  2011-12-05       Impact factor: 4.171

3.  Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters.

Authors:  A K Ebrahim; R Wadachi; H Suda
Journal:  Aust Dent J       Date:  2006-09       Impact factor: 2.291

4.  Influence of embedding media on the assessment of electronic apex locators.

Authors:  Járcio V Baldi; Fausto R Victorino; Ricardo A Bernardes; Ivaldo G de Moraes; Clóvis M Bramante; Roberto B Garcia; Norberti Bernardineli
Journal:  J Endod       Date:  2007-02-22       Impact factor: 4.171

5.  Accuracies of seven different apex locators under various conditions.

Authors:  Jung-A Kang; Sung Kyo Kim
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-07-07

6.  Comparison of working length determination with radiographs and four electronic apex locators.

Authors:  J P Vieyra; J Acosta
Journal:  Int Endod J       Date:  2011-02-10       Impact factor: 5.264

7.  Critical diameter of apical foramen and of file size using the Root ZX apex locator: an in vitro study.

Authors:  Manuela Herrera; Camilo Ábalos; Cristina Lucena; Amparo Jiménez-Planas; Rafael Llamas
Journal:  J Endod       Date:  2011-07-16       Impact factor: 4.171

8.  An in vitro test of a simplified model to demonstrate the operation of electronic root canal measuring devices.

Authors:  R J Czerw; M S Fulkerson; J C Donnelly
Journal:  J Endod       Date:  1994-12       Impact factor: 4.171

9.  The Influence of Various Irrigants on the Accuracy of 2 Electronic Apex Locators in Locating Simulated Root Perforations.

Authors:  Demet Altunbaş; Alper Kuştarcı; Mustafa Toyoğlu
Journal:  J Endod       Date:  2017-01-16       Impact factor: 4.171

10.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

View more
  2 in total

1.  The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.

Authors:  Carlos G Adorno; Sandra M Solaeche; Ileana E Ferreira; Andrea Pedrozo; Patricia M Escobar; Vicente R Fretes
Journal:  Clin Oral Investig       Date:  2021-02-13       Impact factor: 3.573

Review 2.  Accuracy of different generations of apex locators in determining working length; a systematic review and meta-analysis.

Authors:  Kaveh Nasiri; Karl-Thomas Wrbas
Journal:  Saudi Dent J       Date:  2021-09-25
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.