| Literature DB >> 28808636 |
Fernanda Garcia Tampelini1, Marcelo Santos Coelho2, Marcos de Azevêdo Rios1, Carlos Eduardo Fontana3, Daniel Guimarães Pedro Rocha3, Sergio Luiz Pinheiro3, Carlos Eduardo da Silveira Bueno1.
Abstract
OBJECTIVES: The aim of this in vivo study was to assess the accuracy of 2 third-generation electronic apex locators (EALs), Propex II (Dentsply Maillefer) and Root ZX II (J. Morita), and radiographic technique for locating the major foramen (MF).Entities:
Keywords: Dental pulp cavity; Electronic apex locators; Endodontic; Radiographs; Therapeutics; Working length
Year: 2017 PMID: 28808636 PMCID: PMC5553019 DOI: 10.5395/rde.2017.42.3.200
Source DB: PubMed Journal: Restor Dent Endod ISSN: 2234-7658
Results of Pearson's correlation analyses to determine the correlation between the measurements obtained using direct visualization and those obtained using Ingle's radiographic technique, Root ZX II, and Propex II for MF location
| Measurement correlated | Radiographic | Root ZX II | Propex II |
|---|---|---|---|
| No. (pairs) | 32 | 32 | 32 |
| 0.818 | 0.941 | 0.905 | |
| 95% CI | 0.66–0.91 | 0.88–0.97 | 0.81–0.95 |
| 99% CI | 0.59–0.93 | 0.85–0.98 | 0.77–0.96 |
| < 0.001 | < 0.001 | < 0.001 |
Manufacturer information is follow as; Root ZX II, J. Morita, Tokyo, Japan; Propex II, Dentsply Maillefer, Ballaigues, Switerzland.
CI, confidence interval; MF, major foramen.
Results of descriptive statistics using ANOVA (Tukey's post hoc test) to determine discrepancies among measurements obtained using Ingle's radiographic technique, Root ZX II, and Propex II for MF location
| Radiographic technique | Root ZX II | Propex II |
|---|---|---|
| 0.93 ± 0.84a | 0.31 ± 0.59b | 0.50 ± 0.67b |
Manufacturer information is follow as; Root ZX II, J. Morita, Tokyo, Japan; Propex II, Dentsply Maillefer, Ballaigues, Switerzland.
ANOVA, analysis of variance; MF, major foramen.
a,bThe different superscript letters indicate statistically significant differences (p < 0.05).