An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

David Lewis1, Andre Micke, Xiang Yu, Maria F Chan.   

Abstract

PURPOSE: Radiochromic film provides dose measurement at high spatial resolution, but often is not preferred for routine evaluation of patient-specific intensity modulated radiation therapy (IMRT) plans owing to ease-of-use factors. The authors have established an efficient protocol that combines calibration and measurement in a single scan and enables measurement results to be obtained in less than 30 min. This avoids complications due to postexposure changes in radiochromic film that delay the completion of a measurement, often for up to 24 h, in commonly used methods. In addition, the protocol addresses the accuracy and integrity of the measurement by eliminating environmental and interscan variability issues.
METHODS: The authors collected dose-response data from six production lots of Gafchromic EBT3 film and three production lots of EBT2 film at doses up to 480 cGy. In this work, the authors used seven different scanners of two different models-Epson 10000XL and V700; postexposure times before scanning from 30 min to 9 days; ambient temperatures for scanning spanning 11 °C; and two film orientations. Scanning was in 48-bit RGB format at 72 dpi resolution. Dose evaluation was conducted using a triple-channel dosimetry method. To evaluate the measurement protocol, patient specific IMRT and volumetric modulated arc therapy (VMAT) plans were exposed onto EBT3 films on a Varian Trilogy Linac. Film scanning was done following the protocol under a number of different conditions and the dose maps were analyzed to demonstrate the equivalence of results.
RESULTS: The results indicated that the dose-response data could be fit by a set of related rational functions leading to the description of a generic calibration curve. A simplified dosimetry protocol was established where dose-response data for a specific film lot, scanner, and scanning conditions could be derived from two films exposed to known doses. In most cases only one calibrated exposure was required since the dose for one of the films could be zero. Using the Gamma test criterion of 2%∕2 mm to evaluate the measurements, similar passing rates ranging between about 95% and 99% for the fields studied were obtained from application films digitized under a variety of conditions all of them different than the conditions under which the calibration films were scanned.
CONCLUSIONS: The authors have developed a simplified and efficient protocol to measure doses delivered by an IMRT or VMAT plan using only the patient film, one calibration film, one unexposed film, and applying a single scan to acquire a digital image for calculation and analysis. The simplification and timesaving offer a potential practical solution for using radiochromic film for routine treatment plan quality assurance without sacrificing spatial resolution for convenience.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23039670      PMCID: PMC9381144          DOI: 10.1118/1.4754797

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.506


  32 in total

1.  Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry.

Authors:  Joannie Desroches; Hugo Bouchard; Frédéric Lacroix
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

2.  Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies.

Authors:  Bijan Arjomandy; Ramesh Tailor; Aman Anand; Narayan Sahoo; Michael Gillin; Karl Prado; Milos Vicic
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

3.  Dosimetric properties of improved GafChromic films for seven different digitizers.

Authors:  Slobodan Devic; Jan Seuntjens; Gyorgy Hegyi; Ervin B Podgorsak; Christopher G Soares; Assen S Kirov; Imad Ali; Jeffrey F Williamson; Angel Elizondo
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

4.  Characterization and use of EBT radiochromic film for IMRT dose verification.

Authors:  Omar A Zeidan; Stacy Ann L Stephenson; Sanford L Meeks; Thomas H Wagner; Twyla R Willoughby; Patrick A Kupelian; Katja M Langen
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

5.  Comparison of the epson expression 1680 flatbed and the vidar VXR-16 dosimetry PRO film scanners for use in IMRT dosimetry using gafchromic and radiographic film.

Authors:  Ellen Wilcox; George Daskalov; Lucy Nedialkova
Journal:  Med Phys       Date:  2007-01       Impact factor: 4.071

6.  Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance.

Authors:  Martina Fuss; Eva Sturtewagen; Carlos De Wagter; Dietmar Georg
Journal:  Phys Med Biol       Date:  2007-06-18       Impact factor: 3.609

7.  Radiochromic film dosimetry with flatbed scanners: a fast and accurate method for dose calibration and uniformity correction with single film exposure.

Authors:  L Menegotti; A Delana; A Martignano
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

8.  Investigation of energy dependence of EBT and EBT-2 gafchromic film.

Authors:  Patricia Lindsay; Alexandra Rink; Mark Ruschin; David Jaffray
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

9.  Confirmation of skin doses resulting from bolus effect of intervening alpha-cradle and carbon fiber couch in radiotherapy.

Authors:  Maria F Chan; Sou-Tung Chiu-Tsao; Jingdong Li; Karen Schupak; Preeti Parhar; Chandra Burman
Journal:  Technol Cancer Res Treat       Date:  2012-06-15

10.  Linearization of dose-response curve of the radiochromic film dosimetry system.

Authors:  Slobodan Devic; Nada Tomic; Saad Aldelaijan; Francois Deblois; Jan Seuntjens; Maria F Chan; Dave Lewis
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.506

View more
  73 in total

1.  Scattered Dose Calculations and Measurements in a Life-Like Mouse Phantom.

Authors:  David Welch; Leah Turner; Michael Speiser; Gerhard Randers-Pehrson; David J Brenner
Journal:  Radiat Res       Date:  2017-01-31       Impact factor: 2.841

2.  Dosimetric accuracy of dynamic couch rotation during volumetric modulated arc therapy (DCR-VMAT) for primary brain tumours.

Authors:  Gregory Smyth; Philip M Evans; Jeffrey C Bamber; Henry C Mandeville; A Rollo Moore; Liam C Welsh; Frank H Saran; James L Bedford
Journal:  Phys Med Biol       Date:  2019-04-05       Impact factor: 3.609

3.  Technical Note: Measurement of bow tie profiles in CT scanners using radiochromic film.

Authors:  Bruce R Whiting; Andreea C Dohatcu; Joshua D Evans; David G Politte; Jeffrey F Williamson
Journal:  Med Phys       Date:  2015-06       Impact factor: 4.071

4.  Influence of metallic dental implants and metal artefacts on dose calculation accuracy.

Authors:  Manuel Maerz; Oliver Koelbl; Barbara Dobler
Journal:  Strahlenther Onkol       Date:  2014-10-31       Impact factor: 3.621

5.  Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry.

Authors:  David Lewis; Maria F Chan
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

6.  Investigation of EBT2 and EBT3 films for proton dosimetry in the 4-20 MeV energy range.

Authors:  S Reinhardt; M Würl; C Greubel; N Humble; J J Wilkens; M Hillbrand; A Mairani; W Assmann; K Parodi
Journal:  Radiat Environ Biophys       Date:  2015-01-09       Impact factor: 1.925

7.  Design and characterization of an economical (192)Ir hemi-brain small animal irradiator.

Authors:  Michael P Grams; Zachary C Wilson; Terence T Sio; Chris J Beltran; Erik J Tryggestad; Shiv K Gupta; Charles R Blackwell; Kevin P McCollough; Jann N Sarkaria; Keith M Furutani
Journal:  Int J Radiat Biol       Date:  2014-06-25       Impact factor: 2.694

8.  Evaluation of the incident directional dependence of radiochromic film by use of Monte Carlo simulation and measurement.

Authors:  Yuki Yuasa; Shinji Kawamura; Takehiro Shiinoki; Takuya Uehara; Masahiro Koike; Ryuji Kanzaki; Hideki Hanazawa; Shotaro Takahashi; Keiko Shibuya
Journal:  Radiol Phys Technol       Date:  2016-03-24

9.  Patient-Specific QA of Spot-Scanning Proton Beams using Radiochromic Film.

Authors:  Maria F Chan; Chin-Cheng Chen; Chengyu Shi; Jingdong Li; Xiaoli Tang; Xiang Li; Dennis Mah
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2017-05-16

10.  Correction of lateral response artifacts from flatbed scanners for dual-channel radiochromic film dosimetry.

Authors:  Yuichi Akino; Hiroya Shiomi; Fumiaki Isohashi; Osamu Suzuki; Yuji Seo; Keisuke Tamari; Takero Hirata; Hirokazu Mizuno; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2021-03-10       Impact factor: 2.724

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.