Literature DB >> 24844370

Design and characterization of an economical (192)Ir hemi-brain small animal irradiator.

Michael P Grams1, Zachary C Wilson1, Terence T Sio1, Chris J Beltran1, Erik J Tryggestad1, Shiv K Gupta1, Charles R Blackwell1, Kevin P McCollough1, Jann N Sarkaria1, Keith M Furutani1.   

Abstract

PURPOSE: To describe the design and dosimetric characterization of a simple and economical small animal irradiator.
MATERIALS AND METHODS: A high dose rate (HDR) (192)Ir brachytherapy source from a commercially available afterloader was used with a 1.3 cm thick tungsten collimator to provide sharp beam penumbra suitable for hemi-brain irradiation of mice. The unit was equipped with continuous gas anesthesia to allow robust animal immobilization. Dosimetric characterization of the device was performed with Gafchromic film measurements.
RESULTS: The tungsten collimator provided a sharp penumbra suitable for hemi-brain irradiation, and dose rates on the order of 200 cGy/minute were achieved. The sharpness of the penumbra attainable with this device compares favorably to those measured experimentally for 6 MV photons, and 6 and 20 MeV electron beams from a linear accelerator, and was comparable to those measured for a 300 kVp orthovoltage beam and a Monte Carlo simulated 90 MeV proton beam.
CONCLUSIONS: Due to its simplicity and low cost, the apparatus described is an attractive alternative for small animal irradiation experiments requiring steep dose gradients.

Entities:  

Keywords:  Iridium-192 HDR; Small animal irradiator; hemi-brain

Mesh:

Substances:

Year:  2014        PMID: 24844370      PMCID: PMC4564891          DOI: 10.3109/09553002.2014.922719

Source DB:  PubMed          Journal:  Int J Radiat Biol        ISSN: 0955-3002            Impact factor:   2.694


  12 in total

1.  Surface dose in intracavitary orthovoltage radiotherapy.

Authors:  M B Podgorsak; L J Schreiner; E B Podgorsak
Journal:  Med Phys       Date:  1990 Jul-Aug       Impact factor: 4.071

2.  Enhanced epidermal dose caused by localized electron contamination from lead cutouts used in kilovoltage radiotherapy.

Authors:  J E Lye; D J Butler; D V Webb
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

3.  The water equivalence of solid phantoms for low energy photon beams.

Authors:  Robin Hill; Zdenka Kuncic; Clive Baldock
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

4.  MicroRT-small animal conformal irradiator.

Authors:  S Stojadinovic; D A Low; A J Hope; M Vicic; J O Deasy; J Cui; D Khullar; P J Parikh; K T Malinowski; E W Izaguirre; S Mutic; P W Grigsby
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

5.  Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine.

Authors:  A Niroomand-Rad; C R Blackwell; B M Coursey; K P Gall; J M Galvin; W L McLaughlin; A S Meigooni; R Nath; J E Rodgers; C G Soares
Journal:  Med Phys       Date:  1998-11       Impact factor: 4.071

Review 6.  Small animal radiotherapy research platforms.

Authors:  Frank Verhaegen; Patrick Granton; Erik Tryggestad
Journal:  Phys Med Biol       Date:  2011-05-26       Impact factor: 3.609

7.  Dosimetric calibration and characterization for experimental mouse thoracic irradiation using orthovoltage x rays.

Authors:  C D Arndt; I Z Wang; N G Saito; M B Podgorsak
Journal:  Radiat Res       Date:  2011-03-30       Impact factor: 2.841

8.  Feasibility of small animal cranial irradiation with the microRT system.

Authors:  Erich L Kiehl; Strahinja Stojadinovic; Kathleen T Malinowski; David Limbrick; Sarah C Jost; Joel R Garbow; Joshua B Rubin; Joseph O Deasy; Divya Khullar; Enrique W Izaguirre; Parag J Parikh; Daniel A Low; Andrew J Hope
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

9.  High-resolution, small animal radiation research platform with x-ray tomographic guidance capabilities.

Authors:  John Wong; Elwood Armour; Peter Kazanzides; Iulian Iordachita; Erik Tryggestad; Hua Deng; Mohammad Matinfar; Christopher Kennedy; Zejian Liu; Timothy Chan; Owen Gray; Frank Verhaegen; Todd McNutt; Eric Ford; Theodore L DeWeese
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-08-01       Impact factor: 7.038

10.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

Authors:  David Lewis; Andre Micke; Xiang Yu; Maria F Chan
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

View more
  2 in total

1.  Delineation of MGMT Hypermethylation as a Biomarker for Veliparib-Mediated Temozolomide-Sensitizing Therapy of Glioblastoma.

Authors:  Shiv K Gupta; Sani H Kizilbash; Brett L Carlson; Ann C Mladek; Felix Boakye-Agyeman; Katrina K Bakken; Jenny L Pokorny; Mark A Schroeder; Paul A Decker; Ling Cen; Jeanette E Eckel-Passow; Gobinda Sarkar; Karla V Ballman; Joel M Reid; Robert B Jenkins; Roeland G Verhaak; Erik P Sulman; Gaspar J Kitange; Jann N Sarkaria
Journal:  J Natl Cancer Inst       Date:  2015-11-27       Impact factor: 13.506

2.  Half brain irradiation in a murine model of breast cancer brain metastasis: magnetic resonance imaging and histological assessments of dose-response.

Authors:  Niloufar Zarghami; Donna H Murrell; Michael D Jensen; Frederick A Dick; Ann F Chambers; Paula J Foster; Eugene Wong
Journal:  Radiat Oncol       Date:  2018-06-01       Impact factor: 3.481

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.