Dosimetric properties of improved GafChromic films for seven different digitizers.

Slobodan Devic1, Jan Seuntjens, Gyorgy Hegyi, Ervin B Podgorsak, Christopher G Soares, Assen S Kirov, Imad Ali, Jeffrey F Williamson, Angel Elizondo.   

Abstract

Two recently introduced GafChromic film models, HS and XR-T, have been developed as more sensitive and uniform alternatives to GafChromic MD-55-2 film. The HS model has been specifically designed for measurement of absorbed dose in high-energy photon beams (above 1 MeV), while the XR-T model has been introduced for dose measurements of low energy (0.1 MeV) photons. The goal of this study is to compare the sensitometric curves and estimated dosimetric uncertainties associated with seven different GafChromic film dosimetry systems for the two new film models. The densitometers tested are: LKB Pharmacia UltroScan XL, Molecular Dynamics Personal Densitometer, Nuclear Associates Radiochromic Densitometer Model 37-443, Photoelectron Corporation CMR-604, Laser Pro 16, Vidar VXR-16, and AGFA Arcus II document scanner. Pieces of film were exposed to different doses in a dose range from 0.5 to 50 Gy using 6 MV photon beam. Functional forms for dose vs net optical density have been determined for each of the GafChromic film-dosimetry systems used in this comparison. Two sources of uncertainties in dose measurements, governed by the experimental measurement and calibration curve fit procedure, have been compared for the densitometers used. Among the densitometers tested, it is found that for the HS film type the uncertainty caused by the experimental measurement varies from 1% to 3% while the calibration fit uncertainty ranges from 2% to 4% for doses above 5 Gy. Corresponding uncertainties for XR-T film model are somewhat higher and range from 1% to 5% for experimental and from 2% to 7% for the fit uncertainty estimates. Notwithstanding the significant variations in sensitivity, the studied densitometers exhibit very similar precision for GafChromic film based dose measurements above 5 Gy.

Entities:  

Mesh:

Year:  2004        PMID: 15487718     DOI: 10.1118/1.1776691

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  40 in total

1.  Technical note: Heterogeneity dose calculation accuracy in IMRT: study of five commercial treatment planning systems using an anthropomorphic thorax phantom.

Authors:  Scott E Davidson; Richard A Popple; Geoffrey S Ibbott; David S Followill
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

2.  Energy dependence of radiochromic dosimetry films for use in radiotherapy verification.

Authors:  Krzysztof Chełmiński; Wojciech Bulski; Dietmar Georg; Dominika Bodzak; Zbigniew Maniakowski; Dominika Oborska; Joanna Rostkowska; Małgorzata Kania
Journal:  Rep Pract Oncol Radiother       Date:  2010-03-30

3.  Scanning irradiation device for mice in vivo with pulsed and continuous proton beams.

Authors:  Christoph Greubel; Walter Assmann; Christian Burgdorf; Günther Dollinger; Guanghua Du; Volker Hable; Alexander Hapfelmeier; Ralf Hertenberger; Peter Kneschaurek; Dörte Michalski; Michael Molls; Sabine Reinhardt; Barbara Röper; Stefan Schell; Thomas E Schmid; Christian Siebenwirth; Tatiana Wenzl; Olga Zlobinskaya; Jan J Wilkens
Journal:  Radiat Environ Biophys       Date:  2011-05-10       Impact factor: 1.925

4.  Investigation into the feasibility of using PRESAGE/optical-CT dosimetry for the verification of gating treatments.

Authors:  Samuel L Brady; William E Brown; Corey G Clift; Sua Yoo; Mark Oldham
Journal:  Phys Med Biol       Date:  2010-03-26       Impact factor: 3.609

5.  Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry.

Authors:  David Lewis; Maria F Chan
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

6.  Investigation of EBT2 and EBT3 films for proton dosimetry in the 4-20 MeV energy range.

Authors:  S Reinhardt; M Würl; C Greubel; N Humble; J J Wilkens; M Hillbrand; A Mairani; W Assmann; K Parodi
Journal:  Radiat Environ Biophys       Date:  2015-01-09       Impact factor: 1.925

7.  Dosimetric characterization and output verification for conical brachytherapy surface applicators. Part I. Electronic brachytherapy source.

Authors:  Regina K Fulkerson; John A Micka; Larry A DeWerd
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

8.  Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation.

Authors:  Somayeh Gholami; Hamid Reza Mirzaei; Ali Jabbary Arfaee; Ramin Jaberi; Hassan Ali Nedaie; Seied Rabi Mahdavi; Eftekhar Rajab Bolookat; Ali S Meigooni
Journal:  Rep Pract Oncol Radiother       Date:  2016-07-18

9.  Modification and validation of an analytical source model for external beam radiotherapy Monte Carlo dose calculations.

Authors:  Scott E Davidson; Jing Cui; Stephen Kry; Joseph O Deasy; Geoffrey S Ibbott; Milos Vicic; R Allen White; David S Followill
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

10.  Phantom scatter corrections of radiochromic films in high-energy brachytherapy dosimetry: a Monte Carlo study.

Authors:  Mishra Subhalaxmi; T Palani Selvam
Journal:  Radiol Phys Technol       Date:  2015-03-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.