Patient-Specific QA of Spot-Scanning Proton Beams using Radiochromic Film.

Maria F Chan1, Chin-Cheng Chen2, Chengyu Shi1, Jingdong Li1, Xiaoli Tang1, Xiang Li1, Dennis Mah2.   

Abstract

Radiochromic film for spot-scanning QA provides high spatial resolution and efficiency gains from one-shot irradiation for multiple depths. However, calibration can be a tedious procedure which may limit widespread use. Moreover, since there may be an energy dependence, which manifests as a depth dependence, this may require additional measurements for each patient. We present a one-scan protocol to simplify the procedure. A calibration using an EBT3 film, exposed by a 6-level step-wedge plan on a Proteus®PLUS proton system (IBA, Belgium), was performed at depths of 18, 20, 24cm using Plastic Water® (CIRS, Norfolk, VA). The calibration doses ranged from 65-250 cGy(RBE) (relative biological effectiveness) for proton energies of 170-200 MeV. A clinical prostate+nodes plan was used for validation. The planar doses at selected depths were measured with EBT3 films and analyzed using One-scan protocol (one-scan digitization of QA film and at least one film exposed to a known dose). The gamma passing rates, dose-difference maps, and profiles of 2D planar doses measured with EBT3 film and IBA MatriXX-PT, versus the RayStation TPS calculations were analyzed and compared. The EBT3 film measurement results matched well with the TPS calculation data with an average passing rate of ~95% for 2%/2mm and slightly lower passing rates were obtained from an ion chamber array detector. We were able to demonstrate that the use of a proton step-wedge provided clinically acceptable results and minimized variations between film-scanner orientation, inter-scan, and scanning conditions. Furthermore, for relative dosimetry (calibration is not done at the time of experiment) it could be derived from no more than two films exposed to known doses (one could be zero) for rescaling the master calibration curve at each depth. The sensitivity of the calibration to depth variations has been explored. One-scan protocol results appear to be comparable to that of the ion chamber array detector. The use of a proton step-wedge for calibration of EBT3 film potentially increases efficiency in patient-specific QA of proton beams.

Entities:  

Keywords:  GafChromic EBT3; Patient-specific QA; Proton Therapy; film dosimetry

Year:  2017        PMID: 28620561      PMCID: PMC5469417          DOI: 10.4236/ijmpcero.2017.62011

Source DB:  PubMed          Journal:  Int J Med Phys Clin Eng Radiat Oncol        ISSN: 2168-5436


  27 in total

1.  Gafchromic EBT film dosimetry in proton beams.

Authors:  Li Zhao; Indra J Das
Journal:  Phys Med Biol       Date:  2010-04-29       Impact factor: 3.609

2.  Technical Note: Spot characteristic stability for proton pencil beam scanning.

Authors:  Chin-Cheng Chen; Chang Chang; Michael F Moyers; Mingcheng Gao; Dennis Mah
Journal:  Med Phys       Date:  2016-02       Impact factor: 4.071

3.  Fast range measurement of spot scanning proton beams using a volumetric liquid scintillator detector.

Authors:  CheukKai Hui; Daniel Robertson; Fahed Alsanea; Sam Beddar
Journal:  Biomed Phys Eng Express       Date:  2015-07-30

4.  Proposed linear energy transfer areal detector for protons using radiochromic film.

Authors:  Rulon Mayer; Liyong Lin; Marcus Fager; Dan Douglas; James McDonough; Alejandro Carabe
Journal:  Rev Sci Instrum       Date:  2015-04       Impact factor: 1.523

5.  The effect of surgical titanium rods on proton therapy delivered for cervical bone tumors: experimental validation using an anthropomorphic phantom.

Authors:  Isabelle Dietlicher; Margherita Casiraghi; Carmen Ares; Alessandra Bolsi; Damien C Weber; Antony J Lomax; Francesca Albertini
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

6.  Measurements of 2D distributions of absorbed dose in protontherapy with Gafchromic EBT3 films.

Authors:  G Gambarini; V Regazzoni; E Artuso; D Giove; A Mirandola; M Ciocca
Journal:  Appl Radiat Isot       Date:  2015-07-03       Impact factor: 1.513

7.  Evaluation of Gafchromic® EBT3 films characteristics in therapy photon, electron and proton beams.

Authors:  J Sorriaux; A Kacperek; S Rossomme; J A Lee; D Bertrand; S Vynckier; E Sterpin
Journal:  Phys Med       Date:  2012-10-26       Impact factor: 2.685

8.  Dosimetric accuracy assessment of a treatment plan verification system for scanned proton beam radiotherapy: one-year experimental results and Monte Carlo analysis of the involved uncertainties.

Authors:  S Molinelli; A Mairani; A Mirandola; G Vilches Freixas; T Tessonnier; S Giordanengo; K Parodi; M Ciocca; R Orecchia
Journal:  Phys Med Biol       Date:  2013-05-16       Impact factor: 3.609

9.  Calibration of GafChromic EBT3 for absorbed dose measurements in 5 MeV proton beam and (60)Co γ-rays.

Authors:  M Vadrucci; G Esposito; C Ronsivalle; R Cherubini; F Marracino; R M Montereali; L Picardi; M Piccinini; M Pimpinella; M A Vincenti; C De Angelis
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

10.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

Authors:  David Lewis; Andre Micke; Xiang Yu; Maria F Chan
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

View more
  2 in total

1.  The Value of On-Site Proton Audits.

Authors:  Paige A Taylor; Jessica Lowenstein; David Followill; Stephen F Kry
Journal:  Int J Radiat Oncol Biol Phys       Date:  2021-11-13       Impact factor: 7.038

2.  Monte Carlo calculation of the mass stopping power of EBT3 and EBT-XD films for protons for energy ranges of 50-400 MeV.

Authors:  Chengyu Shi; Chin-Cheng Chen; Dennis Mah; Maria F Chan
Journal:  Precis Radiat Oncol       Date:  2018-11-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.