Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance.

Martina Fuss1, Eva Sturtewagen, Carlos De Wagter, Dietmar Georg.   

Abstract

The suitability of radiochromic EBT film was studied for high-precision clinical quality assurance (QA) by identifying the dose response for a wide range of irradiation parameters typically modified in highly-conformal treatment techniques. In addition, uncertainties associated with varying irradiation conditions were determined. EBT can be used for dose assessment of absorbed dose levels as well as relative dosimetry when compared to absolute absorbed dose calibrated using ionization chamber results. For comparison, a silver halide film (Kodak EDR-2) representing the current standard in film dosimetry was included. As an initial step a measurement protocol yielding accurate and precise results was established for a flatbed transparency scanner (Epson Expression 1680 Pro) that was utilized as a film reading instrument. The light transmission measured by the scanner was found to depend on the position of the film on the scanner plate. For three film pieces irradiated with doses of 0 Gy, approximately 1 Gy and approximately 7 Gy, the pixel values measured in portrait or landscape mode differed by 4.7%, 6.2% and 10.0%, respectively. A study of 200 film pieces revealed an excellent sheet-to-sheet uniformity. On a long time scale, the optical development of irradiated EBT film consisted of a slow but steady increase of absorbance which was not observed to cease during 4 months. Sensitometric curves of EBT films obtained under reference conditions (SSD = 95 cm, FS = 5 x 5 cm(2), d = 5 cm) for 6, 10 and 25 MV photon beams did not show any energy dependence. The average separation between all curves was only 0.7%. The variation of the depth d (range 2-25 cm) in the phantom did not affect the dose response of EBT film. Also the influence of the radiation field size (range 3 x 3-40 x 40 cm(2)) on the sensitometric curve was not significant. For EDR-2 films maximum differences between the calibration curves reached 7-8% for X6MV and X25MV. Radiochromic EBT film, in combination with a flatbed scanner, presents a versatile system for high-precision dosimetry in two dimensions, provided that the intrinsic behaviour of the film reading device is taken into account. EBT film itself presents substantial improvements on formerly available models of radiographic and a radiochromic film and its dosimetric characteristics allow us to measure absorbed dose levels in a large variety of situations with a single calibration curve.

Entities:  

Mesh:

Year:  2007        PMID: 17664604     DOI: 10.1088/0031-9155/52/14/013

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  30 in total

1.  Temperature and hydration effects on absorbance spectra and radiation sensitivity of a radiochromic medium.

Authors:  Alexandra Rink; David F Lewis; Sangya Varma; I Alex Vitkin; David A Jaffray
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

2.  Energy dependence of radiochromic dosimetry films for use in radiotherapy verification.

Authors:  Krzysztof Chełmiński; Wojciech Bulski; Dietmar Georg; Dominika Bodzak; Zbigniew Maniakowski; Dominika Oborska; Joanna Rostkowska; Małgorzata Kania
Journal:  Rep Pract Oncol Radiother       Date:  2010-03-30

3.  An investigation of the accuracy of an IMRT dose distribution using two- and three-dimensional dosimetry techniques.

Authors:  Mark Oldham; Harshad Sakhalkar; Pengyi Guo; John Adamovics
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

4.  Simulation for improvement of system sensitivity of radiochromic film dosimetry with different band-pass filters and scanner light intensities.

Authors:  Takeshi Kamomae; Yuki Miyabe; Akira Sawada; Osamu Matoba; Manabu Nakata; Shinsuke Yano; Toru Takakura; Takashi Mizowaki; Akio Itoh; Masahiro Hiraoka
Journal:  Radiol Phys Technol       Date:  2011-03-17

5.  Tumour bed irradiation of human tumour xenografts in a nude rat model using a common X-ray tube.

Authors:  S V Tokalov; W Enghardt; N Abolmaali
Journal:  J Biosci       Date:  2010-06       Impact factor: 1.826

6.  Characterization of a 137Cs irradiator from a new perspective with modern dosimetric tools.

Authors:  Samuel L Brady; Greta Toncheva; Mark W Dewhirst; Terry T Yoshizumi
Journal:  Health Phys       Date:  2009-09       Impact factor: 1.316

7.  A feasibility study using radiochromic films for fast neutron 2D passive dosimetry.

Authors:  Samuel L Brady; Rathnayaka Gunasingha; Terry T Yoshizumi; Calvin R Howell; Alexander S Crowell; Brent Fallin; Anton P Tonchev; Mark W Dewhirst
Journal:  Phys Med Biol       Date:  2010-08-06       Impact factor: 3.609

8.  Commissioning of a novel microCT/RT system for small animal conformal radiotherapy.

Authors:  Manuel Rodriguez; Hu Zhou; Paul Keall; Edward Graves
Journal:  Phys Med Biol       Date:  2009-05-28       Impact factor: 3.609

9.  Alkali Metal Salts of 10,12-Pentacosadiynoic Acid and Their Dosimetry Applications.

Authors:  Amy V Hall; Osama M Musa; David K Hood; David C Apperley; Dmitry S Yufit; Jonathan W Steed
Journal:  Cryst Growth Des       Date:  2021-03-25       Impact factor: 4.076

10.  Clinical utility of Gafchromic film in an MRI-guided linear accelerator.

Authors:  Ilma Xhaferllari; Joshua P Kim; Ruchira Liyanage; Chang Liu; Dongsu Du; Anthony Doemer; Indrin J Chetty; Ning Wen
Journal:  Radiat Oncol       Date:  2021-06-26       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.