Technical Note: Measurement of bow tie profiles in CT scanners using radiochromic film.

Bruce R Whiting1, Andreea C Dohatcu2, Joshua D Evans3, David G Politte4, Jeffrey F Williamson3.   

Abstract

PURPOSE: To provide a noninvasive technique to measure the intensity profile of the fan beam in a computed tomography (CT) scanner that is cost effective and easily implemented without the need to access proprietary scanner information or service modes.
METHODS: The fabrication of an inexpensive aperture is described, which is used to expose radiochromic film in a rotating CT gantry. A series of exposures is made, each of which is digitized on a personal computer document scanner, and the resulting data set is analyzed to produce a self-consistent calibration of relative radiation exposure. The bow tie profiles were analyzed to determine the precision of the process and were compared to two other measurement techniques, direct measurements from CT gantry detectors and a dynamic dosimeter.
RESULTS: The radiochromic film method presented here can measure radiation exposures with a precision of ∼ 6% root-mean-square relative error. The intensity profiles have a maximum 25% root-mean-square relative error compared with existing techniques.
CONCLUSIONS: The proposed radiochromic film method for measuring bow tie profiles is an inexpensive (∼$100 USD + film costs), noninvasive method to measure the fan beam intensity profile in CT scanners.

Mesh:

Year:  2015        PMID: 26127044      PMCID: PMC4441714          DOI: 10.1118/1.4921120

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  On two-parameter models of photon cross sections: application to dual-energy CT imaging.

Authors:  Jeffrey F Williamson; Sicong Li; Slobodan Devic; Bruce R Whiting; Fritz A Lerma
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

2.  Properties of preprocessed sinogram data in x-ray computed tomography.

Authors:  Bruce R Whiting; Parinaz Massoumzadeh; Orville A Earl; Joseph A O'Sullivan; Donald L Snyder; Jeffrey F Williamson
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

3.  Validation of CT dose-reduction simulation.

Authors:  Parinaz Massoumzadeh; Steven Don; Charles F Hildebolt; Kyongtae T Bae; Bruce R Whiting
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

4.  A method to generate equivalent energy spectra and filtration models based on measurement for multidetector CT Monte Carlo dosimetry simulations.

Authors:  Adam C Turner; Di Zhang; Hyun J Kim; John J DeMarco; Chris H Cagnon; Erin Angel; Dianna D Cody; Donna M Stevens; Andrew N Primak; Cynthia H McCollough; Michael F McNitt-Gray
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

5.  Measurement-based model of a wide-bore CT scanner for Monte Carlo dosimetric calculations with GMCTdospp software.

Authors:  Witold Skrzyński
Journal:  Phys Med       Date:  2014-07-11       Impact factor: 2.685

6.  Measurement of bow tie profiles in CT scanners using a real-time dosimeter.

Authors:  Bruce R Whiting; Joshua D Evans; Andreea C Dohatcu; Jeffrey F Williamson; David G Politte
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

7.  Characterization of calibration curves and energy dependence GafChromic™ XR-QA2 model based radiochromic film dosimetry system.

Authors:  Nada Tomic; Chrystian Quintero; Bruce R Whiting; Saad Aldelaijan; Hamed Bekerat; LiHeng Liang; François DeBlois; Jan Seuntjens; Slobodan Devic
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

8.  Nonlinear effects of the recording material on the image quality of a Fourier hologram.

Authors:  H Kasprzak; N Sultanova; H Podbielska
Journal:  J Opt Soc Am A       Date:  1987-05       Impact factor: 2.129

9.  Generalized image combinations in dual KVP digital radiography.

Authors:  L A Lehmann; R E Alvarez; A Macovski; W R Brody; N J Pelc; S J Riederer; A L Hall
Journal:  Med Phys       Date:  1981 Sep-Oct       Impact factor: 4.071

10.  Experimental implementation of a polyenergetic statistical reconstruction algorithm for a commercial fan-beam CT scanner.

Authors:  Joshua D Evans; Bruce R Whiting; David G Politte; Joseph A O'Sullivan; Paul F Klahr; Jeffrey F Williamson
Journal:  Phys Med       Date:  2013-01-21       Impact factor: 2.685

View more
  2 in total

1.  A new technique to characterize CT scanner bow-tie filter attenuation and applications in human cadaver dosimetry simulations.

Authors:  Xinhua Li; Jim Q Shi; Da Zhang; Sarabjeet Singh; Atul Padole; Alexi Otrakji; Mannudeep K Kalra; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

2.  Direct and fast measurement of CT beam filter profiles with simultaneous geometrical calibration.

Authors:  Kai Yang; Xinhua Li; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2017-01       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.