Investigation of energy dependence of EBT and EBT-2 gafchromic film.

Patricia Lindsay1, Alexandra Rink, Mark Ruschin, David Jaffray.   

Abstract

PURPOSE: The purpose of this study was to quantify the extent of energy dependence of Gafchromic film to x-ray energies ranging in quality from 105 kVp to 6 MV, and relate this dependency to the film's chemical composition and date of production.
METHODS: Lots of Gafchromic EBT film manufactured in 2004 and 2005 together with more recent batches produced in 2007 were evaluated for energy dependence. Multiple batches of EBT-2 film were also evaluated. Energy dependence was quantified as Rx-the ratio of net optical density (netOD) measured at a given energy x relative to the netOD measured at 6 MV, as measured on a linear accelerator. Rx was evaluated for beam qualities of 105 and 220 kVp on a clinical orthovoltage unit using two separate techniques-a flatbed scanner (Epson) and a real-time fiber-optic readout system. Neutron activation analysis for chlorine and bromine content was performed on all the films to determine whether the composition of the film had changed between batches of film exhibiting different energy dependence responses.
RESULTS: For batches of EBT manufactured in 2007, R105 kVp was 0.75 and R220 kVp was 0.85, indicating an under-response at orthovoltage energies. These results were confirmed using both the Epson flatbed scanner as well as the real-time readout system. For batches of EBT film manufactured before 2006, Rl05 kVp ranged from 0.9 to 1.0. The results from the neutron activation analysis confirmed a direct relationship between the concentration of chlorine and the magnitude of under-response at orthovoltage energies. EBT-2 film exhibited R105 kVp values ranging from 0.79 (under-response) to 1.20 (over-response) among batches containing varying concentrations of bromine, chlorine, and potassium.
CONCLUSIONS: The results of this study indicated that differences in energy response of EBT and EBT-2 films were due to differences in the chemical composition and therefore the effective atomic number of the film, which have changed over time. To achieve an energy independent dosimeter over a range of kilovoltage energies, the effective atomic number of the dosimeter must be closely matched to that of water. Small deviations in chemical composition can lead to large deviations in response as a function of energy.

Entities:  

Mesh:

Year:  2010        PMID: 20229865     DOI: 10.1118/1.3291622

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Radiation dose uncertainty and correction for a mouse orthotopic and xenograft irradiation model.

Authors:  Gregory N Gan; Cem Altunbas; John J Morton; Justin Eagles; Jennifer Backus; Wayne Dzingle; David Raben; Antonio Jimeno
Journal:  Int J Radiat Biol       Date:  2015-12-21       Impact factor: 2.694

2.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

Authors:  Larry A DeWerd; Geoffrey S Ibbott; Ali S Meigooni; Michael G Mitch; Mark J Rivard; Kurt E Stump; Bruce R Thomadsen; Jack L M Venselaar
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

3.  Investigation of EBT2 and EBT3 films for proton dosimetry in the 4-20 MeV energy range.

Authors:  S Reinhardt; M Würl; C Greubel; N Humble; J J Wilkens; M Hillbrand; A Mairani; W Assmann; K Parodi
Journal:  Radiat Environ Biophys       Date:  2015-01-09       Impact factor: 1.925

4.  High throughput film dosimetry in homogeneous and heterogeneous media for a small animal irradiator.

Authors:  L Wack; W Ngwa; E Tryggestad; P Tsiamas; R Berbeco; S K Ng; J Hesser; P Zygmanski
Journal:  Phys Med       Date:  2013-03-17       Impact factor: 2.685

5.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

Authors:  David Lewis; Andre Micke; Xiang Yu; Maria F Chan
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

6.  Dosimetry in brain tumor phantom at 15 MV 3D conformal radiation therapy.

Authors:  Larissa Thompson; Humberto Galvão Dias; Tarcísio Passos Ribeiro Campos
Journal:  Radiat Oncol       Date:  2013-07-06       Impact factor: 3.481

7.  A novel technique for image-guided local heart irradiation in the rat.

Authors:  Sunil Sharma; Eduardo G Moros; Marjan Boerma; Vijayalakshmi Sridharan; Eun Young Han; Richard Clarkson; Martin Hauer-Jensen; Peter M Corry
Journal:  Technol Cancer Res Treat       Date:  2013-08-31

8.  Is It Possible to Publish a Calibration Function for Radiochromic Film?

Authors:  Maria F Chan; David Lewis; Xiang Yu
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2014-01-29

9.  Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy.

Authors:  Hideharu Miura; Shuichi Ozawa; Fumika Hosono; Naoki Sumida; Toshiya Okazue; Kiyoshi Yamada; Yasushi Nagata
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

10.  Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range.

Authors:  J Eduardo Villarreal-Barajas; Rao F H Khan
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.