Measurement of mini-implant stability using resonance frequency analysis.

Manuel Nienkemper1, Benedict Wilmes, Agamemnon Panayotidis, Alexander Pauls, Vladimir Golubovic, Frank Schwarz, Dieter Drescher.   

Abstract

OBJECTIVE: To investigate whether resonance frequency analysis (RFA) is suitable to measure orthodontic mini-implant stability. Implant size significantly affects the level of resonance frequency. Regarding the operating mode of RFA, it has to be proven whether the resonance frequency of mini-implants in bone fits the range of frequency emitted by the Osstell ISQ device.
MATERIAL AND METHODS: For this purpose the SmartPegs in the Osstell ISQ device were modified to fit with the inner screw thread of orthodontic mini-implants, and 110 mini-implants were inserted into porcine pelvic bone. RFA was performed parallel and perpendicular to the run of superficial bone fibers. A suitability test, Periotest, was also performed in the same directions. Compacta thickness was measured using cone-beam computed tomography. Correlation tests and linear regression analysis were carried out between the three methods.
RESULTS: The RFA showed a mean Implant Stability Quotient value of 36.36 ± 2.67, and the Periotest mean value was -2.10 ± 1.17. The differences between the two directions of measurement were statistically significant (P > .001) for RFA and the Periotest. There was a high correlation between RFA and the Periotest (r  =  -0.90) and between RFA and compacta thickness (r  =  0.71). The comparison between the Periotest and compacta thickness showed a correlation coefficient of r  =  -0.64.
CONCLUSION: The present results suggest that RFA is feasible as a measurement method for orthodontic mini-implant stability. As a consequence, it could be used for clinical evaluation of current stability and allow stability-related loading of mini-implants to reduce the failure rate.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22931201      PMCID: PMC8793650          DOI: 10.2319/043012-354.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  36 in total

1.  Influence of titanium implant surface characteristics on bone regeneration in dehiscence-type defects: an experimental study in dogs.

Authors:  Frank Schwarz; Martin Sager; Ines Kadelka; Daniel Ferrari; Jürgen Becker
Journal:  J Clin Periodontol       Date:  2010-05       Impact factor: 8.728

2.  Early implant failure. Prognostic capacity of Periotest: retrospective study of a large sample.

Authors:  Blas Noguerol; Ricardo Muñoz; Francisco Mesa; Juan de Dios Luna; Francisco O'Valle
Journal:  Clin Oral Implants Res       Date:  2006-08       Impact factor: 5.977

3.  Soft-tissue and cortical-bone thickness at orthodontic implant sites.

Authors:  Hee-Jin Kim; Hee-Sun Yun; Hyun-Do Park; Doo-Hyung Kim; Young-Chel Park
Journal:  Am J Orthod Dentofacial Orthop       Date:  2006-08       Impact factor: 2.650

4.  The use of microimplants in orthodontic anchorage.

Authors:  Chung-Ho Chen; Chao-San Chang; Chi-Hsin Hsieh; Yu-Chuan Tseng; Yee-Shyong Shen; I-Yueh Huang; Chia-Fu Yang; Chun-Ming Chen
Journal:  J Oral Maxillofac Surg       Date:  2006-08       Impact factor: 1.895

5.  Resonance frequency analysis and damping capacity assessment. Part I: an in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability.

Authors:  Stefan Lachmann; Berthold Jäger; Detlef Axmann; German Gomez-Roman; Martin Groten; Heiner Weber
Journal:  Clin Oral Implants Res       Date:  2006-02       Impact factor: 5.977

6.  Impact of insertion depth and predrilling diameter on primary stability of orthodontic mini-implants.

Authors:  Benedict Wilmes; Dieter Drescher
Journal:  Angle Orthod       Date:  2009-07       Impact factor: 2.079

7.  Application and effectiveness of a mini-implant- and tooth-borne rapid palatal expansion device: the hybrid hyrax.

Authors:  Benedict Wilmes; Manuel Nienkemper; Dieter Drescher
Journal:  World J Orthod       Date:  2010

8.  Impact of the method of surface debridement and decontamination on the clinical outcome following combined surgical therapy of peri-implantitis: a randomized controlled clinical study.

Authors:  Frank Schwarz; Narja Sahm; Gerhard Iglhaut; Jürgen Becker
Journal:  J Clin Periodontol       Date:  2011-01-11       Impact factor: 8.728

9.  Correlation between implant stability quotient and bone-implant contact: a retrospective histological and histomorphometrical study of seven titanium implants retrieved from humans.

Authors:  Antonio Scarano; Marco Degidi; Giovanna Iezzi; Giovanna Petrone; Adriano Piattelli
Journal:  Clin Implant Dent Relat Res       Date:  2006       Impact factor: 3.932

10.  Bone thickness of the palate for orthodontic mini-implant anchorage in adults.

Authors:  Sungmin Kang; Shin-Jae Lee; Sug-Joon Ahn; Min-Suk Heo; Tae-Woo Kim
Journal:  Am J Orthod Dentofacial Orthop       Date:  2007-04       Impact factor: 2.650

View more
  14 in total

1.  Comparison of the primary stability of orthodontic miniscrews after repeated insertion cycles.

Authors:  Felipe Nenen; Nicolás Garnica; Víctor Rojas; Rodrigo Oyonarte
Journal:  Angle Orthod       Date:  2021-05-01       Impact factor: 2.079

2.  A novel technique for measurement of orthodontic mini-implant stability using the Osstell ISQ device.

Authors:  Yara K Hosein; S Jeffrey Dixon; Amin S Rizkalla; Ali Tassi
Journal:  Angle Orthod       Date:  2018-11-01       Impact factor: 2.079

3.  Assessment of damping capacity as an index of root proximity in self-drilling orthodontic mini-implants.

Authors:  Mitsuru Motoyoshi; Yasuki Uchida; Miho Matsuoka; Mizuki Inaba; Hiroaki Iwai; Yoko Karasawa; Noriyoshi Shimizu
Journal:  Clin Oral Investig       Date:  2013-03-22       Impact factor: 3.573

4.  Orthodontic mini-implant stability at different insertion depths : Sensitivity of three stability measurement methods.

Authors:  Manuel Nienkemper; Natascha Santel; Ralf Hönscheid; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2016-06-07       Impact factor: 1.938

5.  Significance of bone morphology and quality on the primary stability of orthodontic mini-implants: in vitro comparison between human bone substitute and artificial bone.

Authors:  Sachin Chhatwani; Ouafaa Kouji-Diehl; Kristian Kniha; Ali Modabber; Frank Hölzle; Jozsef Szalma; Gholamreza Danesh; Stephan Christian Möhlhenrich
Journal:  J Orofac Orthop       Date:  2022-03-18       Impact factor: 1.938

6.  Impact of mini-implant length on stability at the initial healing period: a controlled clinical study.

Authors:  Manuel Nienkemper; Benedict Wilmes; Alexander Pauls; Dieter Drescher
Journal:  Head Face Med       Date:  2013-10-20       Impact factor: 2.151

7.  Evaluation of primary stability of innovated orthodontic miniscrew system (STS): An ex-vivo study.

Authors:  Massoud Seifi; Negin-Sadat Matini
Journal:  J Clin Exp Dent       Date:  2016-07-01

Review 8.  Clinical evaluation of osseointegration using resonance frequency analysis.

Authors:  Parth Satwalekar; Sandeep Nalla; Ramaswamy Reddy; Sheeba Glory Chowdary
Journal:  J Indian Prosthodont Soc       Date:  2015 Jul-Sep

9.  Torque Loss After Miniscrew Placement: An In-Vitro Study Followed by a Clinical Trial.

Authors:  Marco Migliorati; Sara Drago; Fabrizio Barberis; Irene Schiavetti; Domenico Dalessandri; Stefano Benedicenti; Armando Silvestrini Biavati
Journal:  Open Dent J       Date:  2016-05-31

Review 10.  Does cortical thickness influence the primary stability of miniscrews?: A systematic review and meta-analysis.

Authors:  Mariana Marquezan; Cláudia Trindade Mattos; Eduardo Franzotti Sant'Anna; Margareth Maria Gomes de Souza; Lucianne Cople Maia
Journal:  Angle Orthod       Date:  2014-04-02       Impact factor: 2.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.