A novel technique for measurement of orthodontic mini-implant stability using the Osstell ISQ device.

Yara K Hosein, S Jeffrey Dixon, Amin S Rizkalla, Ali Tassi.   

Abstract

OBJECTIVES: To develop and validate a method for application of the Osstell ISQ device in the assessment of mini-implant stability.
MATERIALS AND METHODS: An adaptor was developed for attachment of Osstell's SmartPeg onto a variety of orthodontic mini-implants. For validation of the adaptor, Benefit mini-implants were inserted into bone blocks that mimicked different stability conditions. The Osstell device was used to assess mini-implant stability with the adaptor (test measurement) and conventional SmartPeg attachment (gold-standard measurement). Implant stability quotient (ISQ) values were assessed for agreement, repeatability, and reproducibility.
RESULTS: Strong positive correlations were found between ISQ values obtained using the novel adaptor and the conventional attachment. Repeatability and reproducibility of ISQ values with the adaptor were similar to those obtained with the conventional attachment.
CONCLUSIONS: A method was developed and validated to assess the stability of orthodontic mini-implants using the Osstell system. The novel mini-implant adaptor provided repeatable and reproducible measurements of mini-implant stability, which agreed with those obtained using a conventional SmartPeg attachment. This adaptor permits noninvasive stability assessment of various designs of mini-implants, most of which are incompatible with the conventional SmartPeg attachment.

Keywords:  Mini-implants; Osstell ISQ; Primary stability

Mesh:

Substances:

Year:  2018        PMID: 30382751      PMCID: PMC8120874          DOI: 10.2319/011518-46.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  16 in total

1.  Assessment of miniscrew implant stability by resonance frequency analysis: a study in human cadavers.

Authors:  Eduardo Yugo Suzuki; Boonsiva Suzuki; Atchara Aramrattana; Kanchana Harnsiriwattanakit; Narumanas Kowanich
Journal:  J Oral Maxillofac Surg       Date:  2010-11       Impact factor: 1.895

2.  Resonance frequency analysis and damping capacity assessment. Part I: an in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability.

Authors:  Stefan Lachmann; Berthold Jäger; Detlef Axmann; German Gomez-Roman; Martin Groten; Heiner Weber
Journal:  Clin Oral Implants Res       Date:  2006-02       Impact factor: 5.977

3.  Impact of implant design on primary stability of orthodontic mini-implants.

Authors:  Benedict Wilmes; Stephanie Ottenstreuer; Yu-Yu Su; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2008-01       Impact factor: 1.938

Review 4.  Mini-implants in orthodontics: a systematic review of the literature.

Authors:  Reint Reynders; Laura Ronchi; Shandra Bipat
Journal:  Am J Orthod Dentofacial Orthop       Date:  2009-05       Impact factor: 2.650

5.  Soft bone primary stability of 3 different miniscrews for orthodontic anchorage: a resonance frequency investigation.

Authors:  Mario Veltri; Beatrice Balleri; Cecilia Goracci; Roberto Giorgetti; Piero Balleri; Marco Ferrari
Journal:  Am J Orthod Dentofacial Orthop       Date:  2009-05       Impact factor: 2.650

6.  Assessment of Osstell™ and Periotest® systems in measuring dental implant stability (in vitro study).

Authors:  Samer Al-Jetaily; Abdullah Alfarraj Al-Dosari
Journal:  Saudi Dent J       Date:  2010-11-13

Review 7.  Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications.

Authors:  Lars Sennerby; Neil Meredith
Journal:  Periodontol 2000       Date:  2008       Impact factor: 7.589

Review 8.  Assessment of implant stability as a prognostic determinant.

Authors:  N Meredith
Journal:  Int J Prosthodont       Date:  1998 Sep-Oct       Impact factor: 1.681

9.  Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis.

Authors:  N Meredith; D Alleyne; P Cawley
Journal:  Clin Oral Implants Res       Date:  1996-09       Impact factor: 5.977

10.  Comparison of stability between cylindrical and conical type mini-implants. Mechanical and histological properties.

Authors:  Jong-Wan Kim; Seung-Hak Baek; Tae-Woo Kim; Young-Il Chang
Journal:  Angle Orthod       Date:  2008-07       Impact factor: 2.079

View more
  2 in total

1.  Comparison of the primary stability of orthodontic miniscrews after repeated insertion cycles.

Authors:  Felipe Nenen; Nicolás Garnica; Víctor Rojas; Rodrigo Oyonarte
Journal:  Angle Orthod       Date:  2021-05-01       Impact factor: 2.079

2.  Optimization Analysis of Two-Factor Continuous Variable between Thread Depth and Pitch of Microimplant under Toque Force.

Authors:  Yushan Ye; Jiuyang Jiao; Song Fan; Jieying He; Yamei Wang; Qinghe Yao; Wei Wang; Jinsong Li; Shaohai Chang
Journal:  Comput Math Methods Med       Date:  2022-06-20       Impact factor: 2.809

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.