Orthodontic mini-implant stability at different insertion depths : Sensitivity of three stability measurement methods.

Manuel Nienkemper1, Natascha Santel2, Ralf Hönscheid2, Dieter Drescher2.   

Abstract

OBJECTIVES: The purpose of this work was to evaluate the influence of insertion depth on the stability of orthodontic mini-implants. Sensitivity of three different methods to measure implant stability based on differences in insertion depth were determined.
METHODS: A total of 82 mini-implants (2 × 9 mm) were inserted into pelvic bone of Swabian Hall pigs. Each implant was inserted stepwise to depths of 4, 5, 6, 7, and 8 mm. At each of these depths, three different methods were used to measure implant stability, including maximum insertion torque (MIT), resonance frequency analysis (RFA), and Periotest(®). Differences between the recorded values were statistically analyzed and the methods tested for correlations.
RESULTS: Almost linear changes from each insertion depth were measured with the values of RFA [implant stability quotient (ISQ) values range from 1-100], which increased from 6.95 ± 2.85 ISQ at 4 mm to 34.63 ± 5.51 ISQ at 8 mm, and with those of Periotest(®) [periotest values (PTV) range from -8 to 50], which decreased from 13.24 ± 4.03 PTV to -2.89 ± 1.87 PTV. Both methods were found to record highly significant (p < 0.0001) changes for each additional millimeter of insertion depth. The MIT increased significantly (p < 0.0001) from 153.67 ± 69.32 Nmm to 261 ± 103.73 Nmm between 4 and 5 mm of insertion depth but no further significant changes were observed as the implants were driven deeper. The RFA and Periotest(®) values were highly correlated (r = -0.907).
CONCLUSIONS: Mini-implant stability varies significantly with insertion depth. The RFA and the Periotest(®) yielded a linear relationship between stability and insertion depth. MIT does not appear to be an adequate method to determine implant stability based on insertion depth.

Entities:  

Keywords:  Insertion depth; Mini-implants; Models animal; Orthodontics; Skeletal anchorage

Mesh:

Year:  2016        PMID: 27272055     DOI: 10.1007/s00056-016-0036-2

Source DB:  PubMed          Journal:  J Orofac Orthop        ISSN: 1434-5293            Impact factor:   1.938


  24 in total

1.  [Peri-implant bone formation around cylindrical and conical implant systems].

Authors:  A Büchter; J Kleinheinz; H-P Wiesmann; L Seper; U Joos; U Meyer
Journal:  Mund Kiefer Gesichtschir       Date:  2004-07-23

2.  Load-related implant reaction of mini-implants used for orthodontic anchorage.

Authors:  André Büchter; Dirk Wiechmann; Stefan Koerdt; Hans Peter Wiesmann; Josef Piffko; Ulrich Meyer
Journal:  Clin Oral Implants Res       Date:  2005-08       Impact factor: 5.977

3.  Biological and biomechanical evaluation of bone remodelling and implant stability after using an osteotome technique.

Authors:  André Büchter; Johannes Kleinheinz; Hans Peter Wiesmann; Jan Kersken; Manuel Nienkemper; Hendrik von Weyhrother; Ulrich Joos; Ulrich Meyer
Journal:  Clin Oral Implants Res       Date:  2005-02       Impact factor: 5.977

4.  Early implant failure. Prognostic capacity of Periotest: retrospective study of a large sample.

Authors:  Blas Noguerol; Ricardo Muñoz; Francisco Mesa; Juan de Dios Luna; Francisco O'Valle
Journal:  Clin Oral Implants Res       Date:  2006-08       Impact factor: 5.977

5.  Resonance frequency analysis and damping capacity assessment. Part I: an in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability.

Authors:  Stefan Lachmann; Berthold Jäger; Detlef Axmann; German Gomez-Roman; Martin Groten; Heiner Weber
Journal:  Clin Oral Implants Res       Date:  2006-02       Impact factor: 5.977

6.  Parameters affecting primary stability of orthodontic mini-implants.

Authors:  Benedict Wilmes; Carsten Rademacher; Gudrun Olthoff; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2006-05       Impact factor: 1.938

7.  Load-related bone modelling at the interface of orthodontic micro-implants.

Authors:  André Büchter; Dirk Wiechmann; Christoph Gaertner; Marc Hendrik; Martin Vogeler; Hans Peter Wiesmann; Josef Piffko; Ulrich Meyer
Journal:  Clin Oral Implants Res       Date:  2006-12       Impact factor: 5.977

8.  Insertion angle impact on primary stability of orthodontic mini-implants.

Authors:  Benedict Wilmes; Yu-Yu Su; Dieter Drescher
Journal:  Angle Orthod       Date:  2008-11       Impact factor: 2.079

9.  Clinical outcome and bone healing of implants placed with high insertion torque: 12-month results from a multicenter controlled cohort study.

Authors:  T Grandi; P Guazzi; R Samarani; G Grandi
Journal:  Int J Oral Maxillofac Surg       Date:  2012-11-16       Impact factor: 2.789

10.  Measurement of mini-implant stability using resonance frequency analysis.

Authors:  Manuel Nienkemper; Benedict Wilmes; Agamemnon Panayotidis; Alexander Pauls; Vladimir Golubovic; Frank Schwarz; Dieter Drescher
Journal:  Angle Orthod       Date:  2012-08-29       Impact factor: 2.079

View more
  6 in total

1.  Comparison of the primary stability of orthodontic miniscrews after repeated insertion cycles.

Authors:  Felipe Nenen; Nicolás Garnica; Víctor Rojas; Rodrigo Oyonarte
Journal:  Angle Orthod       Date:  2021-05-01       Impact factor: 2.079

2.  Assessment of available sites for palatal orthodontic mini-implants through cone-beam computed tomography.

Authors:  Xinwei Lyu; Jiusi Guo; Liangrui Chen; Yi Gao; Lu Liu; Lingling Pu; Wenli Lai; Hu Long
Journal:  Angle Orthod       Date:  2020-07-01       Impact factor: 2.079

3.  Investigation of the optimal design of orthodontic mini-implants based on the primary stability: A finite element analysis.

Authors:  Amir Hooman Sadr Haghighi; Vahid Pouyafar; Ali Navid; Mahsa Eskandarinezhad; Tannaz Abdollahzadeh Baghaei
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2019-08-14

Review 4.  Effects of low-level laser therapy on the orthodontic mini-implants stability: a systematic review and meta-analysis.

Authors:  Ana Carolina de Figueiredo Costa; Thays Allane Cordeiro Maia; Paulo Goberlânio de Barros Silva; Lucas Guimarães Abreu; Delane Viana Gondim; Pedro César Fernandes Santos
Journal:  Prog Orthod       Date:  2021-02-15       Impact factor: 2.750

Review 5.  Revisiting the Complications of Orthodontic Miniscrew.

Authors:  Van Mai Truong; Soyeon Kim; Jaeheon Kim; Joo Won Lee; Young-Seok Park
Journal:  Biomed Res Int       Date:  2022-08-01       Impact factor: 3.246

6.  Effects of cortical bone thickness and trabecular bone density on primary stability of orthodontic mini-implants.

Authors:  Chin-Yun Pan; Pao-Hsin Liu; Yu-Chuan Tseng; Szu-Ting Chou; Chao-Yi Wu; Hong-Po Chang
Journal:  J Dent Sci       Date:  2019-07-20       Impact factor: 2.080

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.