Significance of bone morphology and quality on the primary stability of orthodontic mini-implants: in vitro comparison between human bone substitute and artificial bone.

Sachin Chhatwani1, Ouafaa Kouji-Diehl2, Kristian Kniha3, Ali Modabber3, Frank Hölzle3, Jozsef Szalma4, Gholamreza Danesh2, Stephan Christian Möhlhenrich2.   

Abstract

AIM: This study evaluated artificial bone models against a human bone substitute to assess the primary stability of orthodontic mini-implants (OMIs) at varying implant sites with different morphologies and qualities.
MATERIALS AND METHODS: A total of 1200 OMI placements of four types were inserted into four artificial bone models of different density (D1, D2, D3, D4) and into a human bone substitute (HB). The implants varied in diameter (2.0 and 2.3 mm) and length (9 and 11 mm). Each specimen had four implant sites: no defect, one-wall defect, three-wall defect, and circular defect. The implant stability quotient (ISQ) values were measured using resonance frequency analysis (RFA) and insertion placement torque values (IPT) were assessed for primary stability. Correlation analysis was performed to evaluate the different models.
RESULTS: The highest IPT value was registered for the 2.0 mm × 11 mm implant inserted into D1 with no defect (37.53 ± 3.02 Ncm). The lowest ISQ value was measured for the 2.3 mm × 9 mm OMI inserted into D3 with a circular defect (12.33 ± 5.88) and the highest for the 2.3 mm × 9 mm implant inserted into HB with no defect (63.23 ± 2.57). A strong correlation (r = 0.64) for IPT values and a very strong correlation (r = 0.8) for ISQ values was found between D2 and HB.
CONCLUSION: Bone defects and bone quality affected the primary stability of implants in terms of ISQ and IPT values. Results for bone model D2 correlated very well with the HB substitution material.
© 2022. The Author(s).

Entities:  

Keywords:  Bone defects; Bone quality; Insertion placement torque; Resonance frequency analysis; Skeletal anchorage

Year:  2022        PMID: 35304617     DOI: 10.1007/s00056-022-00385-8

Source DB:  PubMed          Journal:  J Orofac Orthop        ISSN: 1434-5293            Impact factor:   1.938


  27 in total

1.  Effects of the taper shape, dual-thread, and length on the mechanical properties of mini-implants.

Authors:  Young-Kyun Kim; Yoon-Ji Kim; Pil-Young Yun; Jong-Wan Kim
Journal:  Angle Orthod       Date:  2009-09       Impact factor: 2.079

2.  A comparison of maxillary and mandibular bone mineral densities.

Authors:  H Devlin; K Horner; D Ledgerton
Journal:  J Prosthet Dent       Date:  1998-03       Impact factor: 3.426

3.  Side effects and complications associated with third molar surgery.

Authors:  M Chiapasco; L De Cicco; G Marrone
Journal:  Oral Surg Oral Med Oral Pathol       Date:  1993-10

4.  Primary stability determination by means of insertion torque and RFA in a sample of 4,135 implants.

Authors:  Marco Degidi; Giuseppe Daprile; Adriano Piattelli
Journal:  Clin Implant Dent Relat Res       Date:  2010-09-17       Impact factor: 3.932

5.  Implant stability measurements for implants placed at the time of extraction: a cohort, prospective clinical trial.

Authors:  William Becker; Lars Sennerby; Edwin Bedrossian; Burton E Becker; Jean Pierre Lucchini
Journal:  J Periodontol       Date:  2005-03       Impact factor: 6.993

6.  A comparison of the implant stability among various implant systems: clinical study.

Authors:  Jae-Min Kim; Sun-Jong Kim; Inho Han; Sang-Wan Shin; Jae-Jun Ryu
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

7.  Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test.

Authors:  Dorjpalam Bayarchimeg; Hee Namgoong; Byung Kook Kim; Myung Duk Kim; Sungtae Kim; Tae-Il Kim; Yang Jo Seol; Yong Moo Lee; Young Ku; In-Chul Rhyu; Eun Hee Lee; Ki-Tae Koo
Journal:  J Periodontal Implant Sci       Date:  2013-02-28       Impact factor: 2.614

8.  Mechanical properties of self-drilling orthodontic micro-implants with different diameters.

Authors:  Yan Chen; Hee Moon Kyung; Limin Gao; Won-Jae Yu; Eun-Ju Bae; Sang-Min Kim
Journal:  Angle Orthod       Date:  2010-09       Impact factor: 2.079

9.  Three-dimensional quantitative assessment of palatal bone height for insertion of orthodontic implants - a retrospective CBCT study.

Authors:  Sachin Chhatwani; Viola Rose-Zierau; Bassel Haddad; Mohammed Almuzian; Christian Kirschneck; Gholamreza Danesh
Journal:  Head Face Med       Date:  2019-04-01       Impact factor: 2.151

10.  An in vitro study of factors affecting the primary stability of orthodontic mini-implants.

Authors:  Lindsy Holm; Susan J Cunningham; Aviva Petrie; Richard R J Cousley
Journal:  Angle Orthod       Date:  2012-05-07       Impact factor: 2.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.