Impact of insertion depth and predrilling diameter on primary stability of orthodontic mini-implants.

Benedict Wilmes1, Dieter Drescher.   

Abstract

OBJECTIVE: To test the hypothesis that the impact of the insertion depth and predrilling diameter have no effect on the primary stability of mini-implants.
MATERIALS AND METHODS: Twelve ilium bone segments of pigs were embedded in resin. After implant site preparation with different predrilling diameters (1.0, 1.1, 1.2, and 1.3 mm), Dual Top Screws 1.6 x 10 mm (Jeil, Korea) were inserted with three different insertion depths (7.5, 8.5, and 9.5 mm). The insertion torque was recorded to assess primary stability. In each bone, five Dual Top Screws were used as a reference to compensate for the differences of local bone quality.
RESULTS: Both insertion depth and predrilling diameter influenced the measured insertion torques distinctively: the mean insertion torque for the insertion depth of 7.5 mm was 51.62 Nmm (+/-25.22); for insertion depth of 8.5 mm, 65.53 Nmm (+/-29.99); and for the insertion depth of 9.5 mm, 94.38 Nmm (+/-27.61). The mean insertion torque employing the predrill 1.0 mm was 83.50 Nmm (+/-33.56); for predrill 1.1 mm, 77.50 Nmm (+/-27.54); for the predrill 1.2 mm, 61.70 Nmm (+/-28.46); and for the predrill 1.3 mm, 53.10 (+/-32.18). All differences were highly statistically significant (P < .001).
CONCLUSIONS: The hypothesis is rejected. Higher insertion depths result in higher insertion torques and thus primary stability. Larger predrilling diameters result in lower insertion torques.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19537867     DOI: 10.2319/071708-373.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  13 in total

1.  Orthodontic mini-implant stability at different insertion depths : Sensitivity of three stability measurement methods.

Authors:  Manuel Nienkemper; Natascha Santel; Ralf Hönscheid; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2016-06-07       Impact factor: 1.938

2.  Treatment efficiency of mini-implant-borne distalization depending on age and second-molar eruption.

Authors:  M Nienkemper; B Wilmes; A Pauls; S Yamaguchi; B Ludwig; D Drescher
Journal:  J Orofac Orthop       Date:  2014-03-02       Impact factor: 1.938

3.  Comparison of skeletal and conventional anchorage methods in conjunction with pre-operative decompensation of a skeletal class III malocclusion.

Authors:  Benedict Wilmes; Gudrun Olthoff; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2009-08-02       Impact factor: 1.938

4.  Multipurpose use of orthodontic mini-implants to achieve different treatment goals.

Authors:  M Nienkemper; B Wilmes; A Pauls; D Drescher
Journal:  J Orofac Orthop       Date:  2012-10-28       Impact factor: 1.938

5.  Measurement of mini-implant stability using resonance frequency analysis.

Authors:  Manuel Nienkemper; Benedict Wilmes; Agamemnon Panayotidis; Alexander Pauls; Vladimir Golubovic; Frank Schwarz; Dieter Drescher
Journal:  Angle Orthod       Date:  2012-08-29       Impact factor: 2.079

6.  Effects of repeated sterilization cycles on primary stability of orthodontic mini-screws.

Authors:  Sercan Akyalcin; Holly P McIver; Jeryl D English; Joe C Ontiveros; Ron L Gallerano
Journal:  Angle Orthod       Date:  2012-12-17       Impact factor: 2.079

7.  The effect of multiple processing and re-use on orthodontic mini-screw torque values.

Authors:  Saeed Noorollahian; Shiva Alavi; Elahe Rafiei
Journal:  Dent Res J (Isfahan)       Date:  2015 May-Jun

Review 8.  Systematic review of mini-implant displacement under orthodontic loading.

Authors:  Manuel Nienkemper; Jörg Handschel; Dieter Drescher
Journal:  Int J Oral Sci       Date:  2013-12-20       Impact factor: 6.344

9.  An in vitro study of factors affecting the primary stability of orthodontic mini-implants.

Authors:  Lindsy Holm; Susan J Cunningham; Aviva Petrie; Richard R J Cousley
Journal:  Angle Orthod       Date:  2012-05-07       Impact factor: 2.079

Review 10.  Robotic Applications in Orthodontics: Changing the Face of Contemporary Clinical Care.

Authors:  Samar Adel; Abbas Zaher; Nadia El Harouni; Adith Venugopal; Pratik Premjani; Nikhilesh Vaid
Journal:  Biomed Res Int       Date:  2021-06-16       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.