Assessment of damping capacity as an index of root proximity in self-drilling orthodontic mini-implants.

Mitsuru Motoyoshi1, Yasuki Uchida, Miho Matsuoka, Mizuki Inaba, Hiroaki Iwai, Yoko Karasawa, Noriyoshi Shimizu.   

Abstract

OBJECTIVES: This study aims to investigate orthodontic mini-implant root proximity, placement torque, and damping capacity and to determine whether placement torque and damping capacity (Periotest value (PTV)) are useful indices for the estimation of mini-implant root proximity.
MATERIALS AND METHODS: The root proximity of 143 orthodontic mini-implants (1.6 mm diameter, 8 mm screw thread length) was evaluated in 79 patients (24 males, 55 females; mean age, 22.5 ± 8 years) using cone-beam computed tomography. The placement torque and PTV of each implant were determined using a torque tester and the Periotest, respectively. Variability in these values according to root proximity was evaluated.
RESULTS: PTVs of mini-implants with multiple (two or more) points of contact between the root and implant were significantly larger than those of mini-implants with no root contact in the self-drilling group. Placement torque did not differ significantly according to root proximity. In the self-drilling group, the odds ratio for root contact was 20.82 (P = 0.000) for a PTV >6.
CONCLUSIONS: Placement torque could not be used to estimate root proximity. The PTV was significantly affected by root proximity in the self-drilling group. CLINICAL RELEVANCE: A threshold of PTV >6 could be applied clinically for the estimation of self-drilling mini-implant root proximity.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23519743     DOI: 10.1007/s00784-013-0965-8

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  24 in total

1.  Safe placement techniques for self-drilling orthodontic mini-implants.

Authors:  R Tachibana; M Motoyoshi; A Shinohara; T Shigeeda; N Shimizu
Journal:  Int J Oral Maxillofac Surg       Date:  2012-07-02       Impact factor: 2.789

2.  Resonance frequency analysis and damping capacity assessment. Part I: an in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability.

Authors:  Stefan Lachmann; Berthold Jäger; Detlef Axmann; German Gomez-Roman; Martin Groten; Heiner Weber
Journal:  Clin Oral Implants Res       Date:  2006-02       Impact factor: 5.977

3.  A three-dimensional radiographic-surgical guide for mini-implant placement.

Authors:  Sérgio Estelita Cavalcante Barros; Guilherme Janson; Kelly Chiqueto; Marcos Roberto de Freitas; José Fernando Castanha Henriques; Arnaldo Pinzan
Journal:  J Clin Orthod       Date:  2006-09

4.  Effect of cortical bone thickness and implant placement torque on stability of orthodontic mini-implants.

Authors:  Mitsuru Motoyoshi; Tohru Yoshida; Akiko Ono; Noriyoshi Shimizu
Journal:  Int J Oral Maxillofac Implants       Date:  2007 Sep-Oct       Impact factor: 2.804

5.  Root proximity is a major factor for screw failure in orthodontic anchorage.

Authors:  Shingo Kuroda; Kazuyo Yamada; Toru Deguchi; Takashi Hashimoto; Hee-Moon Kyung; Teruko Takano-Yamamoto
Journal:  Am J Orthod Dentofacial Orthop       Date:  2007-04       Impact factor: 2.650

6.  A precise wire guide for positioning interradicular miniscrews.

Authors:  Hyung Joo Choi; Tae Woo Kim; Hye Won Kim
Journal:  J Clin Orthod       Date:  2007-05

7.  A comparative study of two noninvasive techniques to evaluate implant stability: Periotest and Osstell Mentor.

Authors:  Ji-Su Oh; Su-Gwan Kim; Sung-Chul Lim; Joo L Ong
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-12-17

8.  Evaluation of primary stability of inclined orthodontic mini-implants.

Authors:  Mizuki Inaba
Journal:  J Oral Sci       Date:  2009-09       Impact factor: 1.556

9.  Distal movement of mandibular molars in adult patients with the skeletal anchorage system.

Authors:  Junji Sugawara; Takayoshi Daimaruya; Mikako Umemori; Hiroshi Nagasaka; Ichiro Takahashi; Hiroshi Kawamura; Hideo Mitani
Journal:  Am J Orthod Dentofacial Orthop       Date:  2004-02       Impact factor: 2.650

10.  Measurement of mini-implant stability using resonance frequency analysis.

Authors:  Manuel Nienkemper; Benedict Wilmes; Agamemnon Panayotidis; Alexander Pauls; Vladimir Golubovic; Frank Schwarz; Dieter Drescher
Journal:  Angle Orthod       Date:  2012-08-29       Impact factor: 2.079

View more
  4 in total

Review 1.  A Scoping Review about the Characteristics and Success-Failure Rates of Temporary Anchorage Devices in Orthodontics.

Authors:  Daniel Jaramillo-Bedoya; Gustavo Villegas-Giraldo; Andrés A Agudelo-Suárez; Diana Milena Ramírez-Ossa
Journal:  Dent J (Basel)       Date:  2022-05-06

2.  Insertion torque recordings for the diagnosis of contact between orthodontic mini-implants and dental roots: protocol for a systematic review.

Authors:  Reint Meursinge Reynders; Luisa Ladu; Laura Ronchi; Nicola Di Girolamo; Jan de Lange; Nia Roberts; Annette Plüddemann
Journal:  Syst Rev       Date:  2015-04-02

Review 3.  Insertion torque recordings for the diagnosis of contact between orthodontic mini-implants and dental roots: a systematic review.

Authors:  Reint Meursinge Reynders; Luisa Ladu; Laura Ronchi; Nicola Di Girolamo; Jan de Lange; Nia Roberts; Annette Plüddemann
Journal:  Syst Rev       Date:  2016-03-31

4.  Role of anatomical sites and correlated risk factors on the survival of orthodontic miniscrew implants: a systematic review and meta-analysis.

Authors:  Hisham Mohammed; Khaled Wafaie; Mumen Z Rizk; Mohammed Almuzian; Rami Sosly; David R Bearn
Journal:  Prog Orthod       Date:  2018-09-24       Impact factor: 2.750

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.