OBJECTIVES: The objectives of this study were to determine the accuracy of Periotest to monitor primary implant stability at first-stage surgery, to identify by multivariate analysis the variables associated with early implant failure and to compare Periotest with radiographic study in the diagnosis of implant stability at second-stage surgery (during osseointegration period).
MATERIAL AND METHODS: A 10-year retrospective study was conducted on 1084 Brånemark implants placed in 316 patients. Clinical variables, implant diameter and length, Periotest values (PTVs) and radiological variables were analyzed in bivariate and multivariate studies in order to determine their influence on early implant failure.
RESULTS: After examination of the sensitivity and specificity values obtained for different PTV cutoff points, a cutoff PTV of -2 was selected (84% sensitivity and 39% specificity). In the bivariate analysis, early failure was significantly related to smoking habits, implant location, bone type, implant features and PTVs (-2 and >or=-2). In the final multiple logistic model, only age (odds ratio (OR)=4.53; 95% confidence interval (CI), 1.34-15.27), smoking habits (OR=2.5; 95% CI, 1.3-4.79), bone type (OR=1.93; 95% CI, 1.01-3.7) and PTV at first surgery (OR=3.01; 95% CI, 1.5-6.02) were independently related to early failure.
CONCLUSIONS: The Periotest (with -2 cutoff) at first surgery offers high sensitivity in the prognosis of early implant loss and shows a greater capacity to evaluate stability during the osseointegration period compared with radiographic study.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16907779     DOI: 10.1111/j.1600-0501.2006.01250.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  12 in total

1.  Relation between insertion torque and bone-implant contact percentage: an artificial bone study.

Authors:  Cheng Liu; Ming-Tzu Tsai; Heng-Li Huang; Michael Yuan-Chien Chen; Jui-Ting Hsu; Kuo-Chih Su; Chih-Han Chang; Aaron Yu-Jen Wu
Journal:  Clin Oral Investig       Date:  2012-01-10       Impact factor: 3.573

2.  Effect of Platelet-Rich-Plasma (PRP) and Implant Surface Topography on Implant Stability and Bone.

Authors:  Renu Kundu; Manu Rathee
Journal:  J Clin Diagn Res       Date:  2014-06-20

3.  Orthodontic mini-implant stability at different insertion depths : Sensitivity of three stability measurement methods.

Authors:  Manuel Nienkemper; Natascha Santel; Ralf Hönscheid; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2016-06-07       Impact factor: 1.938

4.  The occurrence of dental implant malpositioning and related factors: A cross-sectional cone-beam computed tomography survey.

Authors:  Yaser Safi; Reza Amid; Fariba Zadbin; Mitra Ghazizadeh Ahsaie; Hamed Mortazavi
Journal:  Imaging Sci Dent       Date:  2021-05-06

5.  Measurement of mini-implant stability using resonance frequency analysis.

Authors:  Manuel Nienkemper; Benedict Wilmes; Agamemnon Panayotidis; Alexander Pauls; Vladimir Golubovic; Frank Schwarz; Dieter Drescher
Journal:  Angle Orthod       Date:  2012-08-29       Impact factor: 2.079

6.  Prospective randomized clinical trial of hydrophilic tapered implant placement at maxillary posterior area: 6 weeks and 12 weeks loading.

Authors:  Seong-Beom Kim; Pil-Young Yun; Sang-Yun Kim; Yang-Jin Yi; Ji-Yun Kim; Young-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2016-10-21       Impact factor: 1.904

Review 7.  The impact of the alveolar bone sites on early implant failure: a systematic review with meta-analysis.

Authors:  Atef Abdel Hameed Fouda
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2020-06-30

8.  The effect of osseodensification and different thread designs on the dental implant primary stability.

Authors:  Abdullah Saleh Almutairi; Maher Abdullatif Walid; Mohamed Ahmed Alkhodary
Journal:  F1000Res       Date:  2018-12-05

9.  Long-Term Influence of Platelet-Rich Plasma (PRP) on Dental Implants after Maxillary Augmentation: Retrospective Clinical and Radiological Outcomes of a Randomized Controlled Clinical Trial.

Authors:  Sameh Attia; Clara Narberhaus; Heidrun Schaaf; Philipp Streckbein; Jörn Pons-Kühnemann; Christian Schmitt; Friedrich Wilhelm Neukam; Hans-Peter Howaldt; Sebastian Böttger
Journal:  J Clin Med       Date:  2020-01-28       Impact factor: 4.241

10.  Osseointegration Effect of Micro-Nano Implants Loaded With Kaempferol in Osteoporotic Rats.

Authors:  Anyue Wang; Wenhong Yuan; Yu Song; Yanjun Zang; Yanling Yu
Journal:  Front Bioeng Biotechnol       Date:  2022-02-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.