FreeStyle Navigator Continuous Glucose Monitoring System with TRUstart algorithm, a 1-hour warm-up time.

McGarraugh Geoffrey1, Ronald Brazg, Weinstein Richard.   

Abstract

BACKGROUND: The first-generation FreeStyle Navigator® Continuous Glucose Monitoring System (FreeStyle Navigator CGM) requires a 10 h warm-up period to avoid inaccurate glucose readings caused by sensor insertion trauma and wound-healing processes. The performance of a second-generation FreeStyle Navigator CGM that begins reporting glucose 1 h after sensor insertion is described.
METHODS: Second-generation FreeStyle Navigator CGM performance was evaluated in an in-clinic study using the YSI Model 2300 STATPlus Glucose Analyzer as reference with 47 subjects with type 1 diabetes. The reference readings were taken at 15 min intervals, and the study was designed to emphasize the first 10 h of use.
RESULTS: The second-generation FreeStyle Navigator CGM exhibited continuous glucose error grid analysis ratings of 93.7% "clinically accurate," 3.6% "benign errors," and 2.8% "clinical errors" and a mean and median absolute relative difference of 14.5% and 10.7%, respectively. The second-generation algorithm detected signal instability in the first 10 h of use and suspended the reporting of 14.1% of first day continuous glucose readings. The clinical accuracy of the second-generation FreeStyle Navigator CGM was similar for the first 10 h versus subsequent hours, with 92.6% and 94.2% "clinically accurate" readings, respectively.
CONCLUSION: The warm-up period for the second-generation FreeStyle Navigator CGM was reduced from 10 to 1 h, with minimal interruption of glucose reporting and without sacrificing clinical performance.
© 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21303631      PMCID: PMC3045239          DOI: 10.1177/193229681100500114

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  10 in total

1.  Evaluating the accuracy of continuous glucose-monitoring sensors: continuous glucose-error grid analysis illustrated by TheraSense Freestyle Navigator data.

Authors:  Boris P Kovatchev; Linda A Gonder-Frederick; Daniel J Cox; William L Clarke
Journal:  Diabetes Care       Date:  2004-08       Impact factor: 19.112

2.  Alarm characterization for a continuous glucose monitor that replaces traditional blood glucose monitoring.

Authors:  Geoffrey McGarraugh
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

3.  Alarm characterization for continuous glucose monitors used as adjuncts to self-monitoring of blood glucose.

Authors:  Geoffrey McGarraugh
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

4.  Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Authors:  Richard L Weinstein; Sherwyn L Schwartz; Ronald L Brazg; Jolyon R Bugler; Thomas A Peyser; Geoffrey V McGarraugh
Journal:  Diabetes Care       Date:  2007-03-02       Impact factor: 19.112

5.  Graphical and numerical evaluation of continuous glucose sensing time lag.

Authors:  Boris P Kovatchev; Devin Shields; Marc Breton
Journal:  Diabetes Technol Ther       Date:  2009-03       Impact factor: 6.118

6.  Peculiarities of the continuous glucose monitoring data stream and their impact on developing closed-loop control technology.

Authors:  Boris Kovatchev; William Clarke
Journal:  J Diabetes Sci Technol       Date:  2008-01

Review 7.  Delays in minimally invasive continuous glucose monitoring devices: a review of current technology.

Authors:  D Barry Keenan; John J Mastrototaro; Gayane Voskanyan; Garry M Steil
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

8.  Evaluating clinical accuracy of systems for self-monitoring of blood glucose.

Authors:  W L Clarke; D Cox; L A Gonder-Frederick; W Carter; S L Pohl
Journal:  Diabetes Care       Date:  1987 Sep-Oct       Impact factor: 19.112

9.  Subcutaneous glucose predicts plasma glucose independent of insulin: implications for continuous monitoring.

Authors:  K Rebrin; G M Steil; W P van Antwerp; J J Mastrototaro
Journal:  Am J Physiol       Date:  1999-09

10.  Detection of hypoglycemia with continuous interstitial and traditional blood glucose monitoring using the FreeStyle Navigator Continuous Glucose Monitoring System.

Authors:  Geoffrey McGarraugh; Richard Bergenstal
Journal:  Diabetes Technol Ther       Date:  2009-03       Impact factor: 6.118

  10 in total
  28 in total

1.  Evaluation of stochastic adjustment for glucose sensor bias during closed-loop insulin delivery.

Authors:  Craig Kollman; Peter Calhoun; John Lum; Werner Sauer; Roy W Beck
Journal:  Diabetes Technol Ther       Date:  2013-11-15       Impact factor: 6.118

2.  Accuracy of subcutaneous continuous glucose monitoring in critically ill adults: improved sensor performance with enhanced calibrations.

Authors:  Lalantha Leelarathna; Shane W English; Hood Thabit; Karen Caldwell; Janet M Allen; Kavita Kumareswaran; Malgorzata E Wilinska; Marianna Nodale; Ahmad Haidar; Mark L Evans; Rowan Burnstein; Roman Hovorka
Journal:  Diabetes Technol Ther       Date:  2013-11-04       Impact factor: 6.118

3.  Continuous glucose monitoring in the subcutaneous tissue over a 14-day sensor wear period.

Authors:  Udo Hoss; Erwin S Budiman; Hanqing Liu; Mark P Christiansen
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

4.  Autoregressive Modeling of Drift and Random Error to Characterize a Continuous Intravascular Glucose Monitoring Sensor.

Authors:  Tony Zhou; Jennifer L Dickson; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2017-07-14

5.  Benefits and Limitations of MARD as a Performance Parameter for Continuous Glucose Monitoring in the Interstitial Space.

Authors:  Lutz Heinemann; Michael Schoemaker; Günther Schmelzeisen-Redecker; Rolf Hinzmann; Adham Kassab; Guido Freckmann; Florian Reiterer; Luigi Del Re
Journal:  J Diabetes Sci Technol       Date:  2019-06-19

6.  Factory-Calibrated Continuous Glucose Sensors: The Science Behind the Technology.

Authors:  Udo Hoss; Erwin Satrya Budiman
Journal:  Diabetes Technol Ther       Date:  2017-05       Impact factor: 6.118

7.  Minimizing the impact of time lag variability on accuracy evaluation of continuous glucose monitoring systems.

Authors:  Cosimo Scuffi; Fausto Lucarelli; Francesco Valgimigli
Journal:  J Diabetes Sci Technol       Date:  2012-11-01

8.  Evaluating the accuracy and large inaccuracy of two continuous glucose monitoring systems.

Authors:  Lalantha Leelarathna; Marianna Nodale; Janet M Allen; Daniela Elleri; Kavita Kumareswaran; Ahmad Haidar; Karen Caldwell; Malgorzata E Wilinska; Carlo L Acerini; Mark L Evans; Helen R Murphy; David B Dunger; Roman Hovorka
Journal:  Diabetes Technol Ther       Date:  2012-12-20       Impact factor: 6.118

9.  Multisite Study of an Implanted Continuous Glucose Sensor Over 90 Days in Patients With Diabetes Mellitus.

Authors:  Andrew Dehennis; Mark A Mortellaro; Sorin Ioacara
Journal:  J Diabetes Sci Technol       Date:  2015-07-29

10.  Effects of a high-carbohydrate versus high-protein meal on acute responses to hypoxia at rest and exercise.

Authors:  Keyne Charlot; Aurélien Pichon; Jean-Paul Richalet; Didier Chapelot
Journal:  Eur J Appl Physiol       Date:  2012-08-24       Impact factor: 3.078

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.