Graphical and numerical evaluation of continuous glucose sensing time lag.

Boris P Kovatchev1, Devin Shields, Marc Breton.   

Abstract

AIM: This study introduces a new method for graphical and numerical evaluation of time lags typically associated with subcutaneous glucose sensing, based on Poincaré-type plot and a maximum statistical agreement criterion.
METHODS: The proposed method is illustrated by retrospective analysis of 56 continuous glucose monitor (CGM) time series collected by the FreeStyle Navigator (Abbott Diabetes Care, Alameda, CA) from 28 patients with type 1 diabetes mellitus, each wearing simultaneously two sensors (on arm and abdomen) and parallel reference blood glucose (BG) collected with a reference YSI (Yellow Springs, OH) analyzer every 15 min. The average duration of a time series was 111 h; there were approximately 10,000 sensor-reference data pairs.
RESULTS: When sliding in time CGM readings versus BG, the point of minimal spread of a Poincaré-type plot marks visually the time of CGM delay. The same point is numerically estimated by minimizing the distance between BG and CGM readings. The average observed time lag between reference BG and CGM was 12.5 min. Stratified by BG rate of change, the time lag was longer (16.8 min) when BG was falling, compared to steady or rising BG (11.7 min and 9.9 min, respectively) (P < 0.005). The time lags at the two sensor locations were not significantly different: 12.4 min on the arm, 12.6 min on the abdomen.
CONCLUSIONS: In this data set, substantial blood-to-sensor time delays were observed, possibly because of both blood-to-interstitial glucose transport and instrumental delay. Analysis of BG-CGM co-dynamics that is free from mathematical approximation of glucose fluctuations resulted in convenient visualization and numerical estimation of these delays.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19216687      PMCID: PMC2939841          DOI: 10.1089/dia.2008.0044

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  19 in total

1.  Determination of time delay between blood and interstitial adipose tissue glucose concentration change by microdialysis in healthy volunteers.

Authors:  K J Wientjes; A J Schoonen
Journal:  Int J Artif Organs       Date:  2001-12       Impact factor: 1.595

2.  Poincaré plot interpretation using a physiological model of HRV based on a network of oscillators.

Authors:  Michael Brennan; Marimuthu Palaniswami; Peter Kamen
Journal:  Am J Physiol Heart Circ Physiol       Date:  2002-11       Impact factor: 4.733

Review 3.  Fascinating rhythm: a primer on chaos theory and its application to cardiology.

Authors:  T A Denton; G A Diamond; R H Helfant; S Khan; H Karagueuzian
Journal:  Am Heart J       Date:  1990-12       Impact factor: 4.749

4.  A novel approach to mitigating the physiological lag between blood and interstitial fluid glucose measurements.

Authors:  Philip J Stout; Joel R Racchini; Michael E Hilgers
Journal:  Diabetes Technol Ther       Date:  2004-10       Impact factor: 6.118

5.  Peculiarities of the continuous glucose monitoring data stream and their impact on developing closed-loop control technology.

Authors:  Boris Kovatchev; William Clarke
Journal:  J Diabetes Sci Technol       Date:  2008-01

6.  Symmetrization of the blood glucose measurement scale and its applications.

Authors:  B P Kovatchev; D J Cox; L A Gonder-Frederick; W Clarke
Journal:  Diabetes Care       Date:  1997-11       Impact factor: 19.112

7.  Interstitial fluid glucose dynamics during insulin-induced hypoglycaemia.

Authors:  G M Steil; K Rebrin; F Hariri; S Jinagonda; S Tadros; C Darwin; M F Saad
Journal:  Diabetologia       Date:  2005-07-07       Impact factor: 10.122

Review 8.  Can interstitial glucose assessment replace blood glucose measurements?

Authors:  K Rebrin; G M Steil
Journal:  Diabetes Technol Ther       Date:  2000       Impact factor: 6.118

9.  Quantifying temporal glucose variability in diabetes via continuous glucose monitoring: mathematical methods and clinical application.

Authors:  Boris P Kovatchev; William L Clarke; Marc Breton; Kenneth Brayman; Anthony McCall
Journal:  Diabetes Technol Ther       Date:  2005-12       Impact factor: 6.118

10.  Modeling of Calibration Effectiveness and Blood-to-Interstitial Glucose Dynamics as Potential Confounders of the Accuracy of Continuous Glucose Sensors during Hyperinsulinemic Clamp.

Authors:  Christopher King; Stacey M Anderson; Marc Breton; William L Clarke; Boris P Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2007-05
View more
  32 in total

1.  Evaluating the clinical accuracy of GlucoMen®Day: a novel microdialysis-based continuous glucose monitor.

Authors:  Francesco Valgimigli; Fausto Lucarelli; Cosimo Scuffi; Sara Morandi; Iolanda Sposato
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Continuous glucose monitoring in subjects with type 1 diabetes: improvement in accuracy by correcting for background current.

Authors:  Joseph El Youssef; Jessica R Castle; Julia M Engle; Ryan G Massoud; W Kenneth Ward
Journal:  Diabetes Technol Ther       Date:  2010-09-30       Impact factor: 6.118

3.  Preclinical Performance Evaluation of Percutaneous Glucose Biosensors: Experimental Considerations and Recommendations.

Authors:  Robert J Soto; Mark H Schoenfisch
Journal:  J Diabetes Sci Technol       Date:  2015-06-17

4.  Discrepancies Between Blood Glucose and Interstitial Glucose-Technological Artifacts or Physiology: Implications for Selection of the Appropriate Therapeutic Target.

Authors:  Thorsten Siegmund; Lutz Heinemann; Ralf Kolassa; Andreas Thomas
Journal:  J Diabetes Sci Technol       Date:  2017-03-21

5.  A miniaturized transcutaneous system for continuous glucose monitoring.

Authors:  Robert A Croce; SanthiSagar Vaddiraju; Jun Kondo; Yan Wang; Liang Zuo; Kai Zhu; Syed K Islam; Diane J Burgess; Fotios Papadimitrakopoulos; Faquir C Jain
Journal:  Biomed Microdevices       Date:  2013-02       Impact factor: 2.838

6.  Minimizing the impact of time lag variability on accuracy evaluation of continuous glucose monitoring systems.

Authors:  Cosimo Scuffi; Fausto Lucarelli; Francesco Valgimigli
Journal:  J Diabetes Sci Technol       Date:  2012-11-01

7.  Optical drug monitoring: photoacoustic imaging of nanosensors to monitor therapeutic lithium in vivo.

Authors:  Kevin J Cash; Chiye Li; Jun Xia; Lihong V Wang; Heather A Clark
Journal:  ACS Nano       Date:  2015-01-21       Impact factor: 15.881

8.  A continuous glucose monitoring device by graphene modified electrochemical sensor in microfluidic system.

Authors:  Zhihua Pu; Chongwei Zou; Ridong Wang; Xiaochen Lai; Haixia Yu; Kexin Xu; Dachao Li
Journal:  Biomicrofluidics       Date:  2016-02-23       Impact factor: 2.800

9.  Glucose Variability: Timing, Risk Analysis, and Relationship to Hypoglycemia in Diabetes.

Authors:  Boris Kovatchev; Claudio Cobelli
Journal:  Diabetes Care       Date:  2016-04       Impact factor: 19.112

10.  Effects of a high-carbohydrate versus high-protein meal on acute responses to hypoxia at rest and exercise.

Authors:  Keyne Charlot; Aurélien Pichon; Jean-Paul Richalet; Didier Chapelot
Journal:  Eur J Appl Physiol       Date:  2012-08-24       Impact factor: 3.078

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.