Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Richard L Weinstein1, Sherwyn L Schwartz, Ronald L Brazg, Jolyon R Bugler, Thomas A Peyser, Geoffrey V McGarraugh.   

Abstract

OBJECTIVE: The purpose of this study was to compare the accuracy of measurements of glucose in interstitial fluid made with the FreeStyle Navigator Continuous Glucose Monitoring System with Yellow Springs Instrument laboratory reference measurements of venous blood glucose. RESEARCH DESIGN AND METHODS: Fifty-eight subjects with type 1 diabetes, aged 18-64 years, were enrolled in a multicenter, prospective, single-arm study. Each subject wore two sensors simultaneously, which were calibrated with capillary fingerstick measurements at 10, 12, 24, and 72 h after insertion. Measurements from the FreeStyle Navigator system were collected at 1-min intervals and compared with venous measurements taken once every 15 min for 50 h over the 5-day period of sensor wear in an in-patient clinical research center. Periods of high rates of change of glucose were induced by insulin and glucose challenges.
RESULTS: Comparison of the FreeStyle Navigator measurements with the laboratory reference method (n = 20,362) gave mean and median absolute relative differences (ARDs) of 12.8 and 9.3%, respectively. The percentage in the clinically accurate Clarke error grid A zone was 81.7% and that in the in the benign error B zone was 16.7%. During low rates of change (< +/-1 mg x dl(-1) x min(-1)), the percentage in the A zone was higher (84.9%) and the mean and median ARDs were lower (11.7 and 8.5%, respectively).
CONCLUSIONS: Measurements with the FreeStyle Navigator system were found to be consistent and accurate compared with venous measurements made using a laboratory reference method over 5 days of sensor wear (82.5% in the A zone on day 1 and 80.9% on day 5).

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17337488     DOI: 10.2337/dc06-1602

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  83 in total

1.  Hypoglycemia prevention via pump attenuation and red-yellow-green "traffic" lights using continuous glucose monitoring and insulin pump data.

Authors:  Colleen S Hughes; Stephen D Patek; Marc D Breton; Boris P Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Continuous glucose monitoring considerations for the development of a closed-loop artificial pancreas system.

Authors:  D Barry Keenan; Benyamin Grosman; Harry W Clark; Anirban Roy; Stuart A Weinzimer; Rajiv V Shah; John J Mastrototaro
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

3.  Toward an injectable continuous osmotic glucose sensor.

Authors:  Erik Johannessen; Olga Krushinitskaya; Andrey Sokolov; Häfliger Philipp; Arno Hoogerwerf; Christian Hinderling; Kari Kautio; Jaakko Lenkkeri; Esko Strömmer; Vasily Kondratyev; Tor Inge Tønnessen; Tom Eirik Mollnes; Henrik Jakobsen; Even Zimmer; Bengt Akselsen
Journal:  J Diabetes Sci Technol       Date:  2010-07-01

4.  Real-time glucose estimation algorithm for continuous glucose monitoring using autoregressive models.

Authors:  Yenny Leal; Winston Garcia-Gabin; Jorge Bondia; Eduardo Esteve; Wifredo Ricart; Jose-Manuel Fernández-Real; Josep Vehí
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

5.  Continuous glucose monitoring: real-time algorithms for calibration, filtering, and alarms.

Authors:  B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

6.  Alarm characterization for a continuous glucose monitor that replaces traditional blood glucose monitoring.

Authors:  Geoffrey McGarraugh
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

7.  Alarm characterization for continuous glucose monitors used as adjuncts to self-monitoring of blood glucose.

Authors:  Geoffrey McGarraugh
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

Review 8.  Technologies for continuous glucose monitoring: current problems and future promises.

Authors:  Santhisagar Vaddiraju; Diane J Burgess; Ioannis Tomazos; Faquir C Jain; Fotios Papadimitrakopoulos
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

9.  Use of subcutaneous interstitial fluid glucose to estimate blood glucose: revisiting delay and sensor offset.

Authors:  Kerstin Rebrin; Norman F Sheppard; Garry M Steil
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 10.  Biocompatible materials for continuous glucose monitoring devices.

Authors:  Scott P Nichols; Ahyeon Koh; Wesley L Storm; Jae Ho Shin; Mark H Schoenfisch
Journal:  Chem Rev       Date:  2013-02-07       Impact factor: 60.622

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.