Alarm characterization for continuous glucose monitors used as adjuncts to self-monitoring of blood glucose.

Geoffrey McGarraugh1.   

Abstract

BACKGROUND: Continuous glucose monitoring (CGM) devices available in the United States are approved for use as adjuncts to self-monitoring of blood glucose (SMBG). Alarm evaluation in the Clinical and Laboratory Standards Institute (CLSI) guideline for CGM does not specifically address devices that employ both CGM and SMBG. In this report, an alarm evaluation method is proposed for these devices.
METHOD: The proposed method builds on the CLSI method using data from an in-clinic study of subjects with type 1 diabetes. CGM was used to detect glycemic events, and SMBG was used to determine treatment. To optimize detection of a single glucose level, such as 70 mg/dl, a range of alarm threshold settings was evaluated. The alarm characterization provides a choice of alarm settings that trade off detection and false alarms. Detection of a range of high glucose levels was similarly evaluated.
RESULTS: Using low glucose alarms, detection of 70 mg/dl within 30 minutes increased from 64 to 97% as alarm settings increased from 70 to 100 mg/dl, and alarms that did not require treatment (SMBG >85 mg/dl) increased from 18 to 52%. Using high glucose alarms, detection of 180 mg/dl within 30 minutes increased from 87 to 96% as alarm settings decreased from 180 to 165 mg/dl, and alarms that did not require treatment (SMBG <180 mg/dl) increased from 24 to 42%.
CONCLUSION: The proposed alarm evaluation method provides information for choosing appropriate alarm thresholds and reflects the clinical utility of CGM alarms. 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20167166      PMCID: PMC2825623          DOI: 10.1177/193229681000400106

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  4 in total

1.  Response to nocturnal alarms using a real-time glucose sensor.

Authors:  Bruce Buckingham; Jen Block; Jonathan Burdick; Andrea Kalajian; Craig Kollman; Michael Choy; Darrell M Wilson; Peter Chase
Journal:  Diabetes Technol Ther       Date:  2005-06       Impact factor: 6.118

2.  Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Authors:  Richard L Weinstein; Sherwyn L Schwartz; Ronald L Brazg; Jolyon R Bugler; Thomas A Peyser; Geoffrey V McGarraugh
Journal:  Diabetes Care       Date:  2007-03-02       Impact factor: 19.112

3.  Detection of hypoglycemia with the GlucoWatch biographer.

Authors:  K R Pitzer; S Desai; T Dunn; S Edelman; Y Jayalakshmi; J Kennedy; J A Tamada; R O Potts
Journal:  Diabetes Care       Date:  2001-05       Impact factor: 19.112

4.  Detection of hypoglycemia with continuous interstitial and traditional blood glucose monitoring using the FreeStyle Navigator Continuous Glucose Monitoring System.

Authors:  Geoffrey McGarraugh; Richard Bergenstal
Journal:  Diabetes Technol Ther       Date:  2009-03       Impact factor: 6.118

  4 in total
  6 in total

1.  Hypoglycemia prevention via pump attenuation and red-yellow-green "traffic" lights using continuous glucose monitoring and insulin pump data.

Authors:  Colleen S Hughes; Stephen D Patek; Marc D Breton; Boris P Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Impaired driving from medical conditions: a 70-year-old man trying to decide if he should continue driving.

Authors:  Matthew Rizzo
Journal:  JAMA       Date:  2011-03-01       Impact factor: 56.272

3.  New Opportunities for Digital Health to Thrive.

Authors:  David C Klonoff; Fraya King; David Kerr
Journal:  J Diabetes Sci Technol       Date:  2019-01-24

Review 4.  Translating cognitive neuroscience to the driver's operational environment: a neuroergonomic approach.

Authors:  Monica N Lees; Joshua D Cosman; John D Lee; Nicola Fricke; Matthew Rizzo
Journal:  Am J Psychol       Date:  2010

5.  FreeStyle Navigator Continuous Glucose Monitoring System with TRUstart algorithm, a 1-hour warm-up time.

Authors:  McGarraugh Geoffrey; Ronald Brazg; Weinstein Richard
Journal:  J Diabetes Sci Technol       Date:  2011-01-01

6.  "Turn it off!": diabetes device alarm fatigue considerations for the present and the future.

Authors:  Joseph P Shivers; Linda Mackowiak; Henry Anhalt; Howard Zisser
Journal:  J Diabetes Sci Technol       Date:  2013-05-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.