Evaluating the accuracy of continuous glucose-monitoring sensors: continuous glucose-error grid analysis illustrated by TheraSense Freestyle Navigator data.

Boris P Kovatchev1, Linda A Gonder-Frederick, Daniel J Cox, William L Clarke.   

Abstract

OBJECTIVE: The objective of this study was to introduce continuous glucose-error grid analysis (CG-EGA) as a method of evaluating the accuracy of continuous glucose-monitoring sensors in terms of both accurate blood glucose (BG) values and accurate direction and rate of BG fluctuations and to illustrate the application of CG-EGA with data from the TheraSense Freestyle Navigator. RESEARCH DESIGN AND METHODS: We approach the design of CG-EGA from the understanding that continuous glucose sensors (CGSs) allow the observation of BG fluctuations as a process in time. We account for specifics of process characterization (location, speed, and direction) and for biological limitations of the observed processes (time lags associated with interstitial sensors). CG-EGA includes two interacting components: 1) point-error grid analysis (P-EGA) evaluates the sensor's accuracy in terms of correct presentation of BG values and 2) rate-error grid analysis (R-EGA) assesses the sensor's ability to capture the direction and rate of BG fluctuations.
RESULTS: CG-EGA revealed that the accuracy of the Navigator, measured as a percentage of accurate readings plus benign errors, was significantly different at hypoglycemia (73.5%), euglycemia (99%), and hyperglycemia (95.4%). Failure to detect hypoglycemia was the most common error. The point accuracy of the Navigator was relatively stable over a wide range of BG rates of change, and its rate accuracy decreased significantly at high BG levels.
CONCLUSIONS: Traditional self-monitoring of BG device evaluation methods fail to capture the important temporal characteristics of the continuous glucose-monitoring process. CG-EGA addresses this problem, thus providing a comprehensive assessment of sensor accuracy that appears to be a useful adjunct to other CGS performance measures.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15277418     DOI: 10.2337/diacare.27.8.1922

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  90 in total

1.  Evaluating the clinical accuracy of GlucoMen®Day: a novel microdialysis-based continuous glucose monitor.

Authors:  Francesco Valgimigli; Fausto Lucarelli; Cosimo Scuffi; Sara Morandi; Iolanda Sposato
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Modeling the effects of subcutaneous insulin administration and carbohydrate consumption on blood glucose.

Authors:  Matthew W Percival; Wendy C Bevier; Youqing Wang; Eyal Dassau; Howard C Zisser; Lois Jovanovič; Francis J Doyle
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 3.  Blood glucose measurements in critically ill patients.

Authors:  Tom Van Herpe; Dieter Mesotten
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

4.  The need for clinical accuracy guidelines for blood glucose monitors.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

5.  Real-time glucose estimation algorithm for continuous glucose monitoring using autoregressive models.

Authors:  Yenny Leal; Winston Garcia-Gabin; Jorge Bondia; Eduardo Esteve; Wifredo Ricart; Jose-Manuel Fernández-Real; Josep Vehí
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

6.  Alarm characterization for a continuous glucose monitor that replaces traditional blood glucose monitoring.

Authors:  Geoffrey McGarraugh
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

7.  Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus.

Authors:  Raymond J Davey; Timothy W Jones; Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

Review 8.  Technologies for continuous glucose monitoring: current problems and future promises.

Authors:  Santhisagar Vaddiraju; Diane J Burgess; Ioannis Tomazos; Faquir C Jain; Fotios Papadimitrakopoulos
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

Review 9.  Biocompatible materials for continuous glucose monitoring devices.

Authors:  Scott P Nichols; Ahyeon Koh; Wesley L Storm; Jae Ho Shin; Mark H Schoenfisch
Journal:  Chem Rev       Date:  2013-02-07       Impact factor: 60.622

10.  Effect of BGM Accuracy on the Clinical Performance of CGM: An In-Silico Study.

Authors:  Enrique Campos-Náñez; Marc D Breton
Journal:  J Diabetes Sci Technol       Date:  2017-05-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.