Continuous glucose monitoring in the subcutaneous tissue over a 14-day sensor wear period.

Udo Hoss1, Erwin S Budiman, Hanqing Liu, Mark P Christiansen.   

Abstract

BACKGROUND: Glucose monitoring systems using subcutaneously inserted sensors are currently labeled for up to 7 days of wear. In this study, we evaluated the feasibility of a 14-day wear duration using a modified version of the sensor found in the Freestyle Navigator™ continuous glucose monitoring system.
METHODS: Sixty-two subjects with diabetes were enrolled in the study. One sensor per subject was inserted on the arm for a wear time of 14 days. Two different calibration algorithms were applied retrospectively, one that uses periodic sensor recalibrations and one without recalibrations. Sensor in vivo stability was determined by least square regression analysis using capillary blood glucose. Mean absolute relative difference (MARD) and mean relative difference were calculated. Consensus error grid analysis was performed by day and over the 14-day wear period to evaluate accuracy of both systems. The sensor insertion sites were inspected after sensor removal for skin reactions.
RESULTS: Sensor data from 55 subjects were used for the analysis. The accuracy metrics for the system with recalibration were calculated to MARD = 13.9% and 84.0% in zone A (error grid analysis). The system without recalibration performed significantly better, resulting in MARD of 12.2% and 88.0% in zone A (p < .0001). The maximum change of in vivo sensor sensitivity over the 14-day wear period was 2% per day. Two subjects reported pain during the first 5 days of sensor wear, and 1 subject reported itching at the sensor site. No further skin reactions were noticed.
CONCLUSIONS: The study shows that a 14-day sensor wear period is achievable. Moreover, sensors using "wired enzyme" technology showed excellent in vivo stability, with no significant sensitivity loss over the 14-day wear period.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24124948      PMCID: PMC3876365          DOI: 10.1177/193229681300700511

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  15 in total

Review 1.  Characterization of implantable biosensor membrane biofouling.

Authors:  N Wisniewski; F Moussy; W M Reichert
Journal:  Fresenius J Anal Chem       Date:  2000 Mar-Apr

2.  Use of subcutaneous interstitial fluid glucose to estimate blood glucose: revisiting delay and sensor offset.

Authors:  Kerstin Rebrin; Norman F Sheppard; Garry M Steil
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

3.  Evaluation of factors affecting CGMS calibration.

Authors:  Bruce A Buckingham; Craig Kollman; Roy Beck; Andrea Kalajian; Rosanna Fiallo-Scharer; Michael J Tansey; Larry A Fox; Darrell M Wilson; Stuart A Weinzimer; Katrina J Ruedy; William V Tamborlane
Journal:  Diabetes Technol Ther       Date:  2006-06       Impact factor: 6.118

4.  Comparison of the numerical and clinical accuracy of four continuous glucose monitors.

Authors:  Boris Kovatchev; Stacey Anderson; Lutz Heinemann; William Clarke
Journal:  Diabetes Care       Date:  2008-03-13       Impact factor: 19.112

5.  The chemistry of commercial continuous glucose monitors.

Authors:  Geoffrey McGarraugh
Journal:  Diabetes Technol Ther       Date:  2009-06       Impact factor: 6.118

6.  A promising solution to enhance the sensocompatibility of biosensors in continuous glucose monitoring systems.

Authors:  Edith E M van den Bosch; Nik H M de Bont; Jun Qiu; Onko-Jan Gelling
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

7.  Continuous glucose monitoring in subcutaneous tissue using factory-calibrated sensors: a pilot study.

Authors:  Udo Hoss; Iman Jeddi; Mark Schulz; Erwin Budiman; Claire Bhogal; Geoffrey McGarraugh
Journal:  Diabetes Technol Ther       Date:  2010-08       Impact factor: 6.118

8.  New features and performance of a next-generation SEVEN-day continuous glucose monitoring system with short lag time.

Authors:  Timothy Bailey; Howard Zisser; Anna Chang
Journal:  Diabetes Technol Ther       Date:  2009-12       Impact factor: 6.118

9.  The effect of local subcutaneous delivery of vascular endothelial growth factor on the function of a chronically implanted amperometric glucose sensor.

Authors:  W Kenneth Ward; Michael D Wood; Heather M Casey; Matthew J Quinn; Isaac F Federiuk
Journal:  Diabetes Technol Ther       Date:  2004-04       Impact factor: 6.118

10.  Feasibility of 10-day use of a continuous glucose-monitoring system in adults with type 1 diabetes.

Authors:  Satish K Garg; Mary K Voelmle; Peter Gottlieb
Journal:  Diabetes Care       Date:  2008-11-25       Impact factor: 19.112

View more
  17 in total

1.  DTT: 19 Years and Counting ….

Authors:  Satish K Garg
Journal:  Diabetes Technol Ther       Date:  2017-02       Impact factor: 6.118

2.  Factory-Calibrated Continuous Glucose Sensors: The Science Behind the Technology.

Authors:  Udo Hoss; Erwin Satrya Budiman
Journal:  Diabetes Technol Ther       Date:  2017-05       Impact factor: 6.118

3.  Flash Glucose Monitoring: The Future Is Here.

Authors:  Satish K Garg; Halis Kaan Akturk
Journal:  Diabetes Technol Ther       Date:  2017-05       Impact factor: 6.118

4.  First Clinical Experience with Retrospective Flash Glucose Monitoring (FGM) Analysis in South Africa: Characterizing Glycemic Control with Ambulatory Glucose Profile.

Authors:  Larry A Distiller; Iain Cranston; Roger Mazze
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

5.  Cutaneous Complications With Continuous or Flash Glucose Monitoring Use: Systematic Review of Trials and Observational Studies.

Authors:  Nurul A Mohd Asarani; Andrew N Reynolds; Sara E Boucher; Martin de Bock; Benjamin J Wheeler
Journal:  J Diabetes Sci Technol       Date:  2019-08-27

6.  A Multicenter Evaluation of the Performance and Usability of a Novel Glucose Monitoring System in Chinese Adults With Diabetes.

Authors:  Linong Ji; Xiaohui Guo; Lixin Guo; Qian Ren; Nan Yu; Jie Zhang
Journal:  J Diabetes Sci Technol       Date:  2016-09-25

7.  CGM Versus FGM; or, Continuous Glucose Monitoring Is Not Flash Glucose Monitoring.

Authors:  Lutz Heinemann; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2015-09-01

8.  The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System.

Authors:  Timothy Bailey; Bruce W Bode; Mark P Christiansen; Leslie J Klaff; Shridhara Alva
Journal:  Diabetes Technol Ther       Date:  2015-07-14       Impact factor: 6.118

9.  Continuous Subcutaneous Insulin Infusion in Children: A Pilot Study Validating a Protocol to Avoid Hypoglycemia at Initiation.

Authors:  Despoina Manousaki; Johnny Deladoëy; Louis Geoffroy; Patricia Olivier
Journal:  Front Endocrinol (Lausanne)       Date:  2017-04-24       Impact factor: 5.555

Review 10.  A review on B-type natriuretic peptide monitoring: assays and biosensors.

Authors:  Rita Maalouf; Steven Bailey
Journal:  Heart Fail Rev       Date:  2016-09       Impact factor: 4.214

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.