Literature DB >> 19216686

Detection of hypoglycemia with continuous interstitial and traditional blood glucose monitoring using the FreeStyle Navigator Continuous Glucose Monitoring System.

Geoffrey McGarraugh1, Richard Bergenstal.   

Abstract

BACKGROUND: The objective of the analysis was to compare detection of hypoglycemic episodes (glucose <70 mg/dL lasting >15 min) with the FreeStyle Navigator Continuous Glucose Monitoring System (FSN-CGM) (Abbott Diabetes Care, Alameda, CA) alarms to detection with traditional finger stick testing at an average frequency of eight tests per day.
METHODS: The performance of FSN-CGM alarms was evaluated in a clinic setting using 58 subjects with type 1 diabetes mellitus (T1DM) monitoring interstitial glucose concentration over a 5-day period compared to reference YSI measurements (instrument manufactured by YSI, Yellow Springs, OH) at 15-min intervals. Finger stick glucose testing was evaluated in the home environment with 91 subjects with TIDM monitoring with the blood glucose meter integrated into the FreeStyle Navigator (FSN-BG) over a 20-day period. The reference was FSN-CGM with results masked from the subjects. Blood glucose values <=85 mg/dL were considered the optimal treatment level to avoid or reverse hypoglycemia.
RESULTS: With a threshold alarm setting of 85 mg/dL, 90.6% of hypoglycemic episodes were detected within +/- 30 min by FSN-CGM in the clinic study. When the alarm was activated, YSI glucose was <= 85 mg/dL 77.2% of the time. In the home environment, the average FSN-BG testing frequency was 7.9 tests per day. Hypoglycemia was verified within +/- 30 min by FSN-BG measurements <= 85 mg/dL at a rate of 27.5%.
CONCLUSIONS: Even with a high rate of FSN-BG testing, hypoglycemia detected by FSN-CGM was verified by patients with T1DM very infrequently. A high rate of hypoglycemia detection with a moderate rate of unnecessary alarms can be attained using FSN-CGM.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19216686     DOI: 10.1089/dia.2008.0047

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  23 in total

1.  Hypoglycemia prevention via pump attenuation and red-yellow-green "traffic" lights using continuous glucose monitoring and insulin pump data.

Authors:  Colleen S Hughes; Stephen D Patek; Marc D Breton; Boris P Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Continuous glucose monitoring: real-time algorithms for calibration, filtering, and alarms.

Authors:  B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

3.  Alarm characterization for a continuous glucose monitor that replaces traditional blood glucose monitoring.

Authors:  Geoffrey McGarraugh
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

4.  Alarm characterization for continuous glucose monitors used as adjuncts to self-monitoring of blood glucose.

Authors:  Geoffrey McGarraugh
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

5.  Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus.

Authors:  Raymond J Davey; Timothy W Jones; Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

6.  Signal processing algorithms implementing the "smart sensor" concept to improve continuous glucose monitoring in diabetes.

Authors:  Andrea Facchinetti; Giovanni Sparacino; Claudio Cobelli
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

7.  Hypoglycemia is more common among type 2 diabetes patients with limited health literacy: the Diabetes Study of Northern California (DISTANCE).

Authors:  Urmimala Sarkar; Andrew J Karter; Jennifer Y Liu; Howard H Moffet; Nancy E Adler; Dean Schillinger
Journal:  J Gen Intern Med       Date:  2010-05-18       Impact factor: 5.128

8.  Overnight closed-loop insulin delivery with model predictive control: assessment of hypoglycemia and hyperglycemia risk using simulation studies.

Authors:  Malgorzata E Wilinska; Erwin S Budiman; Marc B Taub; Daniela Elleri; Janet M Allen; Carlo L Acerini; David B Dunger; Roman Hovorka
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

Review 9.  Clinical need for continuous glucose monitoring in the hospital.

Authors:  Jeffrey I Joseph; Brian Hipszer; Boris Mraovic; Inna Chervoneva; Mark Joseph; Zvi Grunwald
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

10.  Hypoglycemia in Type 2 Diabetes--More Common Than You Think: A Continuous Glucose Monitoring Study.

Authors:  Richa Redhu Gehlaut; Godwin Y Dogbey; Frank L Schwartz; Cynthia R Marling; Jay H Shubrook
Journal:  J Diabetes Sci Technol       Date:  2015-04-27