Measuring output factors of small fields formed by collimator jaws and multileaf collimator using plastic scintillation detectors.

David M Klein1, Ramesh C Tailor, Louis Archambault, Lilie Wang, Francois Therriault-Proulx, A Sam Beddar.   

Abstract

PURPOSE: As the practice of using high-energy photon beams to create therapeutic radiation fields of subcentimeter dimensions (as in intensity-modulated radiotherapy or stereotactic radiosurgery) grows, so too does the need for accurate verification of beam output at these small fields in which standard practices of dose verification break down. This study investigates small-field output factors measured using a small plastic scintillation detector (PSD), as well as a 0.01 cm3 ionization chamber. Specifically, output factors were measured with both detectors using small fields that were defined by either the X-Y collimator jaws or the multileaf collimator (MLC).
METHODS: A PSD of 0.5 mm diameter and 2 mm length was irradiated with 6 and 18 MV linac beams. The PSD was positioned vertically at a source-to-axis distance of 100 cm, at 10 cm depth in a water phantom, and irradiated with fields ranging in size from 0.5 x 0.5 to 10 x 10 cm2. The field sizes were defined either by the collimator jaws alone or by a MLC alone. The MLC fields were constructed in two ways: with the closed leaves (i.e., those leaves that were not opened to define the square field) meeting at either the field center line or at a 4 cm offset from the center line. Scintillation light was recorded using a CCD camera and an estimation of error in the median-filtered signals was made using the bootstrapping technique. Measurements were made using a CC01 ionization chamber under conditions identical to those used for the PSD.
RESULTS: Output factors measured by the PSD showed close agreement with those measured using the ionization chamber for field sizes of 2.0 x 2.0 cm2 and above. At smaller field sizes, the PSD obtained output factors as much as 15% higher than those found using the ionization chamber by 0.6 x 0.6 cm2 jaw-defined fields. Output factors measured with no offset of the closed MLC leaves were as much as 20% higher than those measured using a 4 cm leaf offset.
CONCLUSIONS: The authors' results suggest that PSDs provide a useful and possibly superior alternative to existing dosimetry systems for small fields, as they are inherently less susceptible to volume-averaging and perturbation effects than larger, air-filled ionization chambers. Therefore, PSDs may provide more accurate small-field output factor determination, regardless of the collimation mechanism.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21089789      PMCID: PMC2962666          DOI: 10.1118/1.3488981

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  29 in total

1.  Miniature scintillating detector for small field radiation therapy.

Authors:  D Létourneau; J Pouliot; R Roy
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

2.  The value of the PinPoint ion chamber for characterization of small field segments used in intensity-modulated radiotherapy.

Authors:  C Martens; C De Wagter; W De Neve
Journal:  Phys Med Biol       Date:  2000-09       Impact factor: 3.609

3.  The volume effect of detectors in the dosimetry of small fields used in IMRT.

Authors:  Wolfram U Laub; Tony Wong
Journal:  Med Phys       Date:  2003-03       Impact factor: 4.071

4.  Dosimetric considerations for validation of a sequential IMRT process with a commercial treatment planning system.

Authors:  P Cadman; R Bassalow; N P S Sidhu; G Ibbott; A Nelson
Journal:  Phys Med Biol       Date:  2002-08-21       Impact factor: 3.609

5.  Precision dosimetry for narrow photon beams used in radiosurgery-determination of Gamma Knife output factors.

Authors:  Andreas Mack; Stefan G Scheib; Jörg Major; Stefano Gianolini; Gyorgy Pazmandi; Harald Feist; Heinz Czempiel; Hans-Jürg Kreiner
Journal:  Med Phys       Date:  2002-09       Impact factor: 4.071

6.  A quality assurance phantom for IMRT dose verification.

Authors:  C M Ma; S B Jiang; T Pawlicki; Y Chen; J S Li; J Deng; A L Boyer
Journal:  Phys Med Biol       Date:  2003-03-07       Impact factor: 3.609

7.  Intensity-modulated stereotactic radiosurgery using dynamic micro-multileaf collimation.

Authors:  S H Benedict; R M Cardinale; Q Wu; R D Zwicker; W C Broaddus; R Mohan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-07-01       Impact factor: 7.038

8.  Toward a real-time in vivo dosimetry system using plastic scintillation detectors.

Authors:  Louis Archambault; Tina M Briere; Falk Pönisch; Luc Beaulieu; Deborah A Kuban; Andrew Lee; Sam Beddar
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-16       Impact factor: 7.038

9.  Water equivalence of plastic organic scintillators in megavoltage radiotherapy bremsstrahlung beams.

Authors:  M A Cliftt; R A Sutton; D V Webb
Journal:  Phys Med Biol       Date:  2000-07       Impact factor: 3.609

10.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

View more
  14 in total

1.  Technical note: removing the stem effect when performing Ir-192 HDR brachytherapy in vivo dosimetry using plastic scintillation detectors: a relevant and necessary step.

Authors:  Francois Therriault-Proulx; Sam Beddar; Tina M Briere; Louis Archambault; Luc Beaulieu
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

2.  Study of the response of plastic scintillation detectors in small-field 6 MV photon beams by Monte Carlo simulations.

Authors:  Lilie L W Wang; Sam Beddar
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

3.  Investigating the Electronic Portal Imaging Device for Small Radiation Field Measurements.

Authors:  Arpita Agarwal; Nikhil Rastogi; K J Maria Das; S A Yoganathan; D Udayakumar; Shaleen Kumar
Journal:  J Med Phys       Date:  2017 Apr-Jun

Review 4.  Stereotactic radiosurgery alone for multiple brain metastases? A review of clinical and technical issues.

Authors:  Arjun Sahgal; Mark Ruschin; Lijun Ma; Wilko Verbakel; David Larson; Paul D Brown
Journal:  Neuro Oncol       Date:  2017-04-01       Impact factor: 12.300

5.  Determination of the quenching correction factors for plastic scintillation detectors in therapeutic high-energy proton beams.

Authors:  L L W Wang; L A Perles; L Archambault; N Sahoo; D Mirkovic; S Beddar
Journal:  Phys Med Biol       Date:  2012-11-06       Impact factor: 3.609

6.  Variation of kQclin,Qmsr (fclin,fmsr) for the small-field dosimetric parameters percentage depth dose, tissue-maximum ratio, and off-axis ratio.

Authors:  Paolo Francescon; Sam Beddar; Ninfa Satariano; Indra J Das
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

7.  Temperature dependence of BCF plastic scintillation detectors.

Authors:  Landon Wootton; Sam Beddar
Journal:  Phys Med Biol       Date:  2013-04-11       Impact factor: 3.609

8.  Effect of Magnetic Field Strength on Plastic Scintillation Detector Response.

Authors:  F Therriault-Proulx; Z Wen; G Ibbott; S Beddar
Journal:  Radiat Meas       Date:  2018-06-09       Impact factor: 1.898

9.  Integral T-shaped phantom-dosimeter system to measure transverse and longitudinal dose distributions simultaneously for stereotactic radiosurgery dosimetry.

Authors:  Wook Jae Yoo; Jinsoo Moon; Kyoung Won Jang; Ki-Tek Han; Sang Hun Shin; Dayeong Jeon; Jang-Yeon Park; Byung Gi Park; Bongsoo Lee
Journal:  Sensors (Basel)       Date:  2012-05-14       Impact factor: 3.576

10.  Development of a Small-Sized, Flexible, and Insertable Fiber-Optic Radiation Sensor for Gamma-Ray Spectroscopy.

Authors:  Wook Jae Yoo; Sang Hun Shin; Dong Eun Lee; Kyoung Won Jang; Seunghyun Cho; Bongsoo Lee
Journal:  Sensors (Basel)       Date:  2015-08-28       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.