Variation of kQclin,Qmsr (fclin,fmsr) for the small-field dosimetric parameters percentage depth dose, tissue-maximum ratio, and off-axis ratio.

Paolo Francescon1, Sam Beddar2, Ninfa Satariano1, Indra J Das3.   

Abstract

PURPOSE: Evaluate the ability of different dosimeters to correctly measure the dosimetric parameters percentage depth dose (PDD), tissue-maximum ratio (TMR), and off-axis ratio (OAR) in water for small fields.
METHODS: Monte Carlo (MC) simulations were used to estimate the variation of kQclin,Qmsr (fclin,fmsr) for several types of microdetectors as a function of depth and distance from the central axis for PDD, TMR, and OAR measurements. The variation of kQclin,Qmsr (fclin,fmsr) enables one to evaluate the ability of a detector to reproduce the PDD, TMR, and OAR in water and consequently determine whether it is necessary to apply correction factors. The correctness of the simulations was verified by assessing the ratios between the PDDs and OARs of 5- and 25-mm circular collimators used with a linear accelerator measured with two different types of dosimeters (the PTW 60012 diode and PTW PinPoint 31014 microchamber) and the PDDs and the OARs measured with the Exradin W1 plastic scintillator detector (PSD) and comparing those ratios with the corresponding ratios predicted by the MC simulations.
RESULTS: MC simulations reproduced results with acceptable accuracy compared to the experimental results; therefore, MC simulations can be used to successfully predict the behavior of different dosimeters in small fields. The Exradin W1 PSD was the only dosimeter that reproduced the PDDs, TMRs, and OARs in water with high accuracy. With the exception of the EDGE diode, the stereotactic diodes reproduced the PDDs and the TMRs in water with a systematic error of less than 2% at depths of up to 25 cm; however, they produced OAR values that were significantly different from those in water, especially in the tail region (lower than 20% in some cases). The microchambers could be used for PDD measurements for fields greater than those produced using a 10-mm collimator. However, with the detector stem parallel to the beam axis, the microchambers could be used for TMR measurements for all field sizes. The microchambers could not be used for OAR measurements for small fields.
CONCLUSIONS: Compared with MC simulation, the Exradin W1 PSD can reproduce the PDDs, TMRs, and OARs in water with a high degree of accuracy; thus, the correction used for converting dose is very close to unity. The stereotactic diode is a viable alternative because it shows an acceptable systematic error in the measurement of PDDs and TMRs and a significant underestimation in only the tail region of the OAR measurements, where the dose is low and differences in dose may not be therapeutically meaningful.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25281947      PMCID: PMC5175987          DOI: 10.1118/1.4895978

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  37 in total

1.  Measuring output factors of small fields formed by collimator jaws and multileaf collimator using plastic scintillation detectors.

Authors:  David M Klein; Ramesh C Tailor; Louis Archambault; Lilie Wang; Francois Therriault-Proulx; A Sam Beddar
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

2.  A new formalism for reference dosimetry of small and nonstandard fields.

Authors:  R Alfonso; P Andreo; R Capote; M Saiful Huq; W Kilby; P Kjäll; T R Mackie; H Palmans; K Rosser; J Seuntjens; W Ullrich; S Vatnitsky
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

3.  The influence of small field sizes, penumbra, spot size and measurement depth on perturbation factors for microionization chambers.

Authors:  F Crop; N Reynaert; G Pittomvils; L Paelinck; C De Wagter; L Vakaet; H Thierens
Journal:  Phys Med Biol       Date:  2009-04-21       Impact factor: 3.609

4.  On the wall perturbation correction for a parallel-plate NACP-02 chamber in clinical electron beams.

Authors:  K Zink; J Wulff
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

5.  Measurements of dose distributions in small beams of 6 MV x-rays.

Authors:  R K Rice; J L Hansen; G K Svensson; R L Siddon
Journal:  Phys Med Biol       Date:  1987-09       Impact factor: 3.609

6.  X-ray source and the output factor.

Authors:  T C Zhu; B E Bjärngard; H Shackford
Journal:  Med Phys       Date:  1995-06       Impact factor: 4.071

7.  Comments on dose measurements for a narrow beam in radiosurgery.

Authors:  A Wu; R D Zwicker; A M Kalend; Z Zheng
Journal:  Med Phys       Date:  1993 May-Jun       Impact factor: 4.071

8.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

9.  Dosimetric characterization of CyberKnife radiosurgical photon beams using polymer gels.

Authors:  E Pantelis; C Antypas; L Petrokokkinos; P Karaiskos; P Papagiannis; M Kozicki; E Georgiou; L Sakelliou; I Seimenis
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

10.  Monte Carlo calculated correction factors for diodes and ion chambers in small photon fields.

Authors:  D Czarnecki; K Zink
Journal:  Phys Med Biol       Date:  2013-03-21       Impact factor: 3.609

View more
  10 in total

1.  Dosimetric characterization and behaviour in small X-ray fields of a microchamber and a plastic scintillator detector.

Authors:  Massimo Pasquino; Claudia Cutaia; Lorenzo Radici; Serena Valzano; Eva Gino; Carlo Cavedon; Michele Stasi
Journal:  Br J Radiol       Date:  2016-11-09       Impact factor: 3.039

Review 2.  Stereotactic radiosurgery alone for multiple brain metastases? A review of clinical and technical issues.

Authors:  Arjun Sahgal; Mark Ruschin; Lijun Ma; Wilko Verbakel; David Larson; Paul D Brown
Journal:  Neuro Oncol       Date:  2017-04-01       Impact factor: 12.300

3.  Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector.

Authors:  Suresh H Chaudhari; Rishabh Dobhal; Rajesh A Kinhikar; Sudarshan S Kadam; Deepak D Deshpande
Journal:  J Med Phys       Date:  2017 Jan-Mar

4.  Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators.

Authors:  Madelaine K Tyler; Paul Z Y Liu; Christopher Lee; David R McKenzie; Natalka Suchowerska
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

5.  A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams.

Authors:  Božidar Casar; Eduard Gershkevitsh; Ignasi Mendez; Slaven Jurković; M Saiful Huq
Journal:  Med Phys       Date:  2018-12-24       Impact factor: 4.071

6.  Consistency of small-field dosimetry, on and off axis, in beam-matched linacs used for stereotactic radiosurgery.

Authors:  Luis Muñoz; Tomas Kron; Marco Petasecca; Joseph Bucci; Michael Jackson; Peter Metcalfe; Anatoly B Rosenfeld; Giordano Biasi
Journal:  J Appl Clin Med Phys       Date:  2021-01-13       Impact factor: 2.102

7.  Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields.

Authors:  Dustin J Jacqmin; Jessica R Miller; Brendan A Barraclough; Zacariah E Labby
Journal:  J Appl Clin Med Phys       Date:  2022-07-21       Impact factor: 2.243

8.  Commissioning and validation of RayStation treatment planning system for CyberKnife M6.

Authors:  Maude Gondré; Mireille Conrad; Véronique Vallet; Jean Bourhis; François Bochud; Raphaël Moeckli
Journal:  J Appl Clin Med Phys       Date:  2022-07-20       Impact factor: 2.243

9.  Detector-specific correction factors in radiosurgery beams and their impact on dose distribution calculations.

Authors:  Olivia A García-Garduño; Manuel A Rodríguez-Ávila; José M Lárraga-Gutiérrez
Journal:  PLoS One       Date:  2018-05-15       Impact factor: 3.240

10.  Small field output correction factors of the microSilicon detector and a deeper understanding of their origin by quantifying perturbation factors.

Authors:  Carolin Weber; Rafael Kranzer; Jan Weidner; Kevin Kröninger; Björn Poppe; Hui Khee Looe; Daniela Poppinga
Journal:  Med Phys       Date:  2020-04-13       Impact factor: 4.071

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.