Investigating the Electronic Portal Imaging Device for Small Radiation Field Measurements.

Arpita Agarwal1, Nikhil Rastogi1, K J Maria Das2, S A Yoganathan2, D Udayakumar2, Shaleen Kumar2.   

Abstract

PURPOSE: With the advent of state-of-the-art treatment technologies, the use of small fields has increased, and dosimetry in small fields is highly challenging. In this study, the potential use of Varian electronic portal imaging device (EPID) for small field measurements was explored for 6 and 15 MV photon beams.
MATERIALS AND METHODS: The output factors and profiles were measured for a range of jaw-collimated square field sizes starting from 0.8 cm × 0.8 cm to 10 cm × 10 cm using EPID. For evaluation purpose, reference data were acquired using Exradin A16 microionization chamber (0.007 cc) for output factors and stereotactic field diode for profile measurements in a radiation field analyzer.
RESULTS: The output factors of EPID were in agreement with the reference data for field sizes down to 2 cm × 2 cm and for 2 cm × 2 cm; the difference in output factors was +2.06% for 6 MV and +1.56% for 15 MV. For the lowest field size studied (0.8 cm × 0.8 cm), the differences were maximum; +16% for 6 MV and +23% for 15 MV photon beam. EPID profiles of both energies were closely matching with reference profiles for field sizes down to 2 cm × 2 cm; however, penumbra and measured field size of EPID profiles were slightly lower compared to its counterpart.
CONCLUSIONS: EPID is a viable option for profile and output factor measurements for field sizes down to 2 cm × 2 cm in the absence of appropriate small field dosimeters.

Entities:  

Keywords:  Dosimetry; electronic portal imaging device; portal dosimetry; small field dosimetry

Year:  2017        PMID: 28706350      PMCID: PMC5496271          DOI: 10.4103/jmp.JMP_131_16

Source DB:  PubMed          Journal:  J Med Phys        ISSN: 0971-6203


  39 in total

1.  Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device.

Authors:  K L Pasma; M L Dirkx; M Kroonwijk; A G Visser; B J Heijmen
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

2.  Dosimetric IMRT verification with a flat-panel EPID.

Authors:  B Warkentin; S Steciw; S Rathee; B G Fallone
Journal:  Med Phys       Date:  2003-12       Impact factor: 4.071

3.  Measurement and modeling of the effect of support arm backscatter on dosimetry with a varian EPID.

Authors:  Pejman Rowshanfarzad; Boyd M C McCurdy; Mahsheed Sabet; Christopher Lee; Daryl J O'Connor; Peter B Greer
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

4.  Measuring output factors of small fields formed by collimator jaws and multileaf collimator using plastic scintillation detectors.

Authors:  David M Klein; Ramesh C Tailor; Louis Archambault; Lilie Wang; Francois Therriault-Proulx; A Sam Beddar
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

5.  A new formalism for reference dosimetry of small and nonstandard fields.

Authors:  R Alfonso; P Andreo; R Capote; M Saiful Huq; W Kilby; P Kjäll; T R Mackie; H Palmans; K Rosser; J Seuntjens; W Ullrich; S Vatnitsky
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

6.  Experimental comparison of seven commercial dosimetry diodes for measurement of stereotactic radiosurgery cone factors.

Authors:  Sonja Dieterich; George W Sherouse
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

7.  A verification procedure to improve patient set-up accuracy using portal images.

Authors:  A Bel; M van Herk; H Bartelink; J V Lebesque
Journal:  Radiother Oncol       Date:  1993-11       Impact factor: 6.280

8.  Small SRS photon field profile dosimetry performed using a PinPoint air ion chamber, a diamond detector, a novel silicon-diode array (DOSI), and polymer gel dosimetry. Analysis and intercomparison.

Authors:  E Pappas; T G Maris; F Zacharopoulou; A Papadakis; S Manolopoulos; S Green; C Wojnecki
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

9.  Challenges of small photon field dosimetry are still challenging.

Authors:  S D Sharma
Journal:  J Med Phys       Date:  2014-07

10.  Dosimetric characterization of CyberKnife radiosurgical photon beams using polymer gels.

Authors:  E Pantelis; C Antypas; L Petrokokkinos; P Karaiskos; P Papagiannis; M Kozicki; E Georgiou; L Sakelliou; I Seimenis
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

View more
  1 in total

Review 1.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.