Literature DB >> 23574889

Temperature dependence of BCF plastic scintillation detectors.

Landon Wootton1, Sam Beddar.   

Abstract

We examined temperature dependence in plastic scintillation detectors (PSDs) made of BCF-60 or BCF-12 scintillating fiber coupled to optical fiber with cyanoacrylate. PSDs were subjected to a range of temperatures using a temperature-controlled water bath and irradiated at each temperature while either the dose was measured using a CCD camera or the spectral output was measured using a spectrometer. The spectrometer was used to examine the intensity and spectral distribution of scintillation light emitted by the PSDs, Cerenkov light generated within the PSD, and light transmitted through an isolated optical coupling. BCF-60 PSDs exhibited a 0.50% decrease and BCF-12 PSDs a 0.09% decrease in measured dose per °C increase, relative to dose measured at 22 °C. Spectrometry revealed that the total intensity of the light generated by BCF-60 and BCF-12 PSDs decreased by 0.32% and 0.13%, respectively, per °C increase. The spectral distribution of the light changed slightly with temperature for both PSDs, accounting for the disparity between the change in measured dose and total light output. The generation of Cerenkov light was temperature independent. However, light transmitted through optical coupling between the scintillator and the optical fiber also exhibited temperature dependence.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23574889      PMCID: PMC3849698          DOI: 10.1088/0031-9155/58/9/2955

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  12 in total

1.  Measuring output factors of small fields formed by collimator jaws and multileaf collimator using plastic scintillation detectors.

Authors:  David M Klein; Ramesh C Tailor; Louis Archambault; Lilie Wang; Francois Therriault-Proulx; A Sam Beddar
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

2.  Plastic scintillation dosimetry: optimal selection of scintillating fibers and scintillators.

Authors:  Louis Archambault; Jean Arsenault; Luc Gingras; A Sam Beddar; René Roy; Luc Beaulieu
Journal:  Med Phys       Date:  2005-07       Impact factor: 4.071

3.  Measurement accuracy and cerenkov removal for high performance, high spatial resolution scintillation dosimetry.

Authors:  Louis Archambault; A Sam Beddar; Luc Gingras; René Roy; Luc Beaulieu
Journal:  Med Phys       Date:  2006-01       Impact factor: 4.071

4.  Transient noise characterization and filtration in CCD cameras exposed to stray radiation from a medical linear accelerator.

Authors:  Louis Archambault; Tina Marie Briere; Sam Beddar
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

5.  On possible temperature dependence of plastic scintillator response.

Authors:  Sam Beddar
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

6.  Toward a real-time in vivo dosimetry system using plastic scintillation detectors.

Authors:  Louis Archambault; Tina M Briere; Falk Pönisch; Luc Beaulieu; Deborah A Kuban; Andrew Lee; Sam Beddar
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-16       Impact factor: 7.038

7.  Spectral discrimination of Cerenkov radiation in scintillating dosimeters.

Authors:  A M Frelin; J M Fontbonne; G Ban; J Colin; M Labalme; A Batalla; A Isambert; A Vela; T Leroux
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

8.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

9.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

10.  In-phantom dose verification of prostate IMRT and VMAT deliveries using plastic scintillation detectors.

Authors:  David Klein; Tina Marie Briere; Rajat Kudchadker; Louis Archambault; Luc Beaulieu; Andrew Lee; Sam Beddar
Journal:  Radiat Meas       Date:  2012-08-24       Impact factor: 1.898

View more
  12 in total

1.  Inorganic scintillation detectors based on Eu-activated phosphors for 192Ir brachytherapy.

Authors:  Gustavo Kertzscher; Sam Beddar
Journal:  Phys Med Biol       Date:  2017-05-05       Impact factor: 3.609

Review 2.  In vivo dosimetry: trends and prospects for brachytherapy.

Authors:  G Kertzscher; A Rosenfeld; S Beddar; K Tanderup; J E Cygler
Journal:  Br J Radiol       Date:  2014-07-08       Impact factor: 3.039

3.  Characterization of a scintillating fibre detector for small animal imaging and irradiation dosimetry.

Authors:  Coralie Le Deroff; Anne-Marie Frelin-Labalme; Xavier Ledoux
Journal:  Br J Radiol       Date:  2016-09-16       Impact factor: 3.039

4.  Ruby-based inorganic scintillation detectors for 192Ir brachytherapy.

Authors:  Gustavo Kertzscher; Sam Beddar
Journal:  Phys Med Biol       Date:  2016-10-14       Impact factor: 3.609

5.  Real-time in vivo rectal wall dosimetry using plastic scintillation detectors for patients with prostate cancer.

Authors:  Landon Wootton; Rajat Kudchadker; Andrew Lee; Sam Beddar
Journal:  Phys Med Biol       Date:  2014-01-17       Impact factor: 3.609

6.  A method to correct for temperature dependence and measure simultaneously dose and temperature using a plastic scintillation detector.

Authors:  Francois Therriault-Proulx; Landon Wootton; Sam Beddar
Journal:  Phys Med Biol       Date:  2015-09-25       Impact factor: 3.609

7.  Exradin W1 plastic scintillation detector for in vivo skin dosimetry in passive scattering proton therapy.

Authors:  Fahed Alsanea; Landon Wootton; Narayan Sahoo; Rajat Kudchadker; Usama Mahmood; Sam Beddar
Journal:  Phys Med       Date:  2018-02-27       Impact factor: 2.685

8.  Preliminary evaluation of the dosimetric accuracy of the in vivo plastic scintillation detector OARtrac system for prostate cancer treatments.

Authors:  Slade J Klawikowski; Clint Zeringue; Landon S Wootton; Geoffrey S Ibbott; Sam Beddar
Journal:  Phys Med Biol       Date:  2014-04-15       Impact factor: 3.609

9.  Effects of Temperature and X-rays on Plastic Scintillating Fiber and Infrared Optical Fiber.

Authors:  Bongsoo Lee; Sang Hun Shin; Kyoung Won Jang; Wook Jae Yoo
Journal:  Sensors (Basel)       Date:  2015-05-11       Impact factor: 3.576

10.  Real-time in vivo dosimetry for SBRT prostate treatment using plastic scintillating dosimetry embedded in a rectal balloon: a case study.

Authors:  Justin L Cantley; Chee-Wai Cheng; Fredrick B Jesseph; Tarun K Podder; Valdir C Colussi; Bryan J Traughber; Lee E Ponsky; Rodney J Ellis
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.