Jiri Gallo1, Vitezslav Havranek, Jana Zapletalova, Jiri Lostak.
Abstract
BACKGROUND: We studied which factor could predict aseptic loosening in ABG I hip prosthesis with hydroxyapatite coating. Aseptic loosening and periprosthetic osteolysis are believed to be caused, at least in part, by increased polyethylene (PE) wear rate via particle disease. Based on it, increased PE wear rate should be associated with aseptic loosening regardless of the type of implant.Entities:
Mesh:
Year: 2010 PMID: 20959001 PMCID: PMC2974673 DOI: 10.1186/1471-2474-11-243
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Categorical variables included in the study
| Stable implant | Unstable cup | Unstable stem | |||
|---|---|---|---|---|---|
| Variable | Categories | N. of hips (%) | N. of hips (%) | N. of hips (%) | |
| Gender | Men | 33 (27) | 11 (34) | 0.511 | 0 (0%) |
| Women | 88 (73) | 21 (66) | 2 (100%) | ||
| Osteoarthritis | 27 (22) | 2 (6) | 0 (0) | ||
| Hip dysplasia | 56 (46) | 14 (44) | 1 (50) | ||
| Primary diagnosis | Osteonecrosis | 20 (17) | 10 (31) | 0.078 | 0 (0) |
| Traumatic | 9 (7) | 5 (16) | 0 (0) | ||
| Inflammatory | 4 (3) | 1 (3) | 1 (50) | ||
| Others | 5 (4) | 0 (0) | 0 (0) | ||
| A | 42 (35) | 11 (34) | 0 (0) | ||
| Charnley type | B | 73 (60) | 14 (44) | 2 (100) | |
| C | 6 (5) | 7 (22) | 0 (0) | ||
| Liner geometry | Neutral | 96 (79) | 27 (84) | 2 (100) | |
| Hooded | 25 (21) | 5 (16) | 0.623 | 0 (0) | |
| Head material | CoCr | 111 (92) | 31 (97) | 1 (50) | |
| Zirconia | 10 (8) | 1 (3) | 0.460 | 1 (50) | |
| Cup relation to KL | Laterally | 10 (8) | 4 (12.5) | 0 (0) | |
| In contact | 23 (19) | 4 (12.5) | 0.565 | 0 (0) | |
| Medially | 88 (73) | 24 (75) | 2 (100) | ||
| True acetabular region | Yes | 97 (80) | 26 (81) | 0 (0) | |
| No | 24 (20) | 6 (19) | 1.000 | 2 (100) | |
| Type I, II | 41 (34) | 2 (6) | 1 (50) | ||
| Acetabular BD | Type III | 67 (55) | 19 (60) | 1 (50) | |
| Type IV, V | 13 (11) | 11 (34) | 0 (0) | ||
| Type I, II | 99 (82) | 27 (84) | 1 (50) | ||
| Femoral BD | Type III | 19 (16) | 5 (16) | 1.000 | 1 (50) |
| Type IV | 3 (2) | 0 (0) | 0 (0) | ||
| Implant removal | 1 (1) | 1 (3) | 0 (0) | ||
| Type of surgery | Cup revision | 36 (30) | 8 (25) | 0.489 | 0 (0) |
| THA revision | 84 (69) | 23 (72) | 2 (100) | ||
aFisher's exact test, KL-Kohler's line, BD-bone defects
Continuous variables included in the study
| Variable | Stable implant (N = 121) | Unstable cup (N = 32) | |
|---|---|---|---|
| Age at surgery (yrs.) | 45.6 ± 7.1 | 47.8 ± 5.8 | 0.102a |
| 45.4 (25.1-64.5) | 48.3 (37.6-56.7) | ||
| Height (cm) | 165.7 ± 8.7 | 165.4 ± 9.5 | 0.864a |
| 165 (150-195) | 165 (149-187) | ||
| Weight (kg) | 75.1 ± 14.3 | 75.2 ± 14.8 | 0.918b |
| 73 (42-114) | 73 (50-105) | ||
| Cup size (mm) | 50.2 ± 3.7 | 50.5 ± 4.1 | 0.830b |
| 50 (46-60) | 49 (46-60) | ||
| PE liner thickness (mm) | 7.0 ± 1.9 | 7.2 ± 2.0 | 0.805b |
| 6.9 (2.8-11.9) | 6.4 (4.9-11.9) | ||
| Abduction angle of the cup (°) | 46.0 ± 8.2 | 43.0 ± 6.3 | 0.073b |
| 46 (28-72) | 43 (28-54) | ||
| Time to revision surgery (yrs.) | 5.7 ± 1.8 | 6.9 ± 1.8 | |
| 5.7 (2.0-10.4) | 6.9 (3.7-10.5) | ||
| Linear wear rate (mm/yrs.) | 0.42 ± 0.38 | 0.39 ± 0.32 | 0.764b |
| 0.35 (0.0-2.28) | 0.31 (0.0-1.32) | ||
| Volumetric wear rate (mm3/yrs.) | 157 ± 139 | 144 ± 119 | 0.774b |
| 127 (0-815) | 107 (12-467) | ||
Mean ± SD (standard deviation), Median (Range), a Student's t-test, b Mann-Whitney U-test, PE-polyethylene
Variables associated with the probability of aseptic loosening of the cup in ABG I prosthesis
| Variable | Odds ratio | 95% confidence interval | p value | ||
|---|---|---|---|---|---|
| Men | 1.729 | 18.125 | 0.375 | ||
| Charnley C type of patient | 1.498 | 30.047 | |||
| Increasing acetabular BD by 1 step of severity | 1.647 | 9.980 | |||
| Increasing FU by 1 year | 1.096 | 2.086 | |||
| Increasing abduction angle of the cup by 1° | 0.846 | 0.987 | |||
BD-bone defects; FU-follow-up; R2-Nagelkerke R Square