Catastrophic failure of an uncemented acetabular component due to high wear and osteolysis: an analysis of 154 omnifit prostheses with mean 6-year follow-up.

Thord von Schewelov1, Lennart Sanzén, Ingemar Onsten, Ake Carlsson.   

Abstract

BACKGROUND: The purposes of this study were (1) to evaluate the wear pattern of the hydroxyapatite-coated "Dual Radius" Omnifit cup, (2) to investigate whether wear is correlated to any demographic or prosthesis-related factors, and (3) to describe micromotion of both the cup and the stem. PATIENTS AND METHODS: 154 hips were implanted between 1990 and 1996 and followed for an average of 6 years. Wear was measured according to the "Charnley-duo" method and, in 79 hips, with radiostereometry IRSA). RSA was also used to evaluate micromotion. We analyzed the femoral heads using scanning electron microscopy, energy dispersive X-ray spectroscopy and an atomic force microscope. RESULT: 66 cups were revised and had a mean annual wear of 0.32 mm compared to 0.12 mm in hips not revised. Osteolytic processes were observed in 35 hips but at revision osteolysis was present in 51 cases. 43/66 sockets were loose. Micromotion evaluated by RSA, weight, age, side, size of cup, screws, polyethylene thickness or shelf-life of the polyethylene did not correlate to wear, whereas male gender did.
INTERPRETATION: It is still unclear why about half of our cases had an abnormal wear rate. Annual wear exceeding 0.2 mm is prognostic of late failure and should be considered a warning sign.

Entities:  

Mesh:

Year:  2004        PMID: 15260420     DOI: 10.1080/00016470410001213

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  6 in total

1.  Early polyethylene wear and osteolysis with ABG acetabular cups (7- to 12-year follow-up).

Authors:  Sachin Badhe; Peter Livesley
Journal:  Int Orthop       Date:  2005-11-08       Impact factor: 3.075

Review 2.  Uncemented Cups and Impaction Bone Grafting for Acetabular Bone Loss in Revision Hip Arthroplasty: A Review of Rationale, Indications, and Outcomes.

Authors:  Rocco D'Apolito; Luigi Zagra
Journal:  Materials (Basel)       Date:  2022-05-23       Impact factor: 3.748

3.  Male gender, Charnley class C, and severity of bone defects predict the risk for aseptic loosening in the cup of ABG I hip arthroplasty.

Authors:  Jiri Gallo; Vitezslav Havranek; Jana Zapletalova; Jiri Lostak
Journal:  BMC Musculoskelet Disord       Date:  2010-10-19       Impact factor: 2.362

Review 4.  Early subsidence of shape-closed hip arthroplasty stems is associated with late revision. A systematic review and meta-analysis of 24 RSA studies and 56 survival studies.

Authors:  Paul van der Voort; Bart G Pijls; Marc J Nieuwenhuijse; Jorrit Jasper; Marta Fiocco; Josepha W M Plevier; Saskia Middeldorp; Edward R Valstar; Rob G H H Nelissen
Journal:  Acta Orthop       Date:  2015       Impact factor: 3.717

Review 5.  Adherence of hip and knee arthroplasty studies to RSA standardization guidelines. A systematic review.

Authors:  Rami Madanat; Tatu J Mäkinen; Hannu T Aro; Charles Bragdon; Henrik Malchau
Journal:  Acta Orthop       Date:  2014-06-23       Impact factor: 3.717

Review 6.  Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survivalstudies.

Authors:  Bart G Pijls; Marc J Nieuwenhuijse; Marta Fiocco; Josepha Wm Plevier; Saskia Middeldorp; Rob Ghh Nelissen; Edward R Valstar
Journal:  Acta Orthop       Date:  2012-11-05       Impact factor: 3.717

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.