Wear measurement of retrieved polyethylene ABG 1 cups by universal-type measuring microscope and X-ray methods.

Jiri Gallo1, Vitezslav Havranek, Ivana Cechova, Jana Zapletalova.   

Abstract

BACKGROUND: Polyethylene wear is considered a most important part of periprosthetic osteolysis development. Thus, its measurement is central to contemporary orthopaedics. AIMS: The aim of this paper was to compare the accuracy of three radiographic techniques for wear measurement. Secondly, the influence of the abduction angle of the cup on measurement accuracy was investigated.
METHODS: Wear was measured manually in 80 patients by a single observer according to the Livermore, Charnley, and Dorr description. A multi-component statistical analysis was used to test the hypothesis that the Livermore technique was superior. In vitro data obtained from a Universal-type measuring microscope served as a gold standard.
RESULTS: In vitro measurements showed an average linear wear of 0.363 mm per year (0.000-0.939, SD 0.241) with a corresponding volumetric wear rate of 161 mm3 per year (0-467, SD 118.2). The Livermore technique showed the least deviation from the optical reference standard and a superior position from the viewpoint of error analysis but the correlation coefficient was slightly less (r = 0.761) than for the Dorr and Charnley techniques (r = 0.795 and r = 0.778, respectively). In addition, the mean error of the Dorr method differed significantly from zero (p = 0.036). Overall, the Livermore technique was the most accurate method for polyethylene wear measurement regardless of the abduction angle of the cup.
CONCLUSIONS: The Livermore technique performed manually was more accurate than the Charnley and Dorr methods. Nevertheless, we consider the Dorr technique an adequate tool for day-to-day wear measurements, mainly due to its simplicity.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17426800     DOI: 10.5507/bp.2006.049

Source DB:  PubMed          Journal:  Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub        ISSN: 1213-8118            Impact factor:   1.245


  4 in total

1.  Male gender, Charnley class C, and severity of bone defects predict the risk for aseptic loosening in the cup of ABG I hip arthroplasty.

Authors:  Jiri Gallo; Vitezslav Havranek; Jana Zapletalova; Jiri Lostak
Journal:  BMC Musculoskelet Disord       Date:  2010-10-19       Impact factor: 2.362

2.  Wear pattern observations from TDR retrievals using autoregistration of voxel data.

Authors:  Yakov P Shkolnikov; Anton Bowden; Daniel MacDonald; Steven M Kurtz
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2010-08       Impact factor: 3.405

3.  Variation in cytokine genes can contribute to severity of acetabular osteolysis and risk for revision in patients with ABG 1 total hip arthroplasty: a genetic association study.

Authors:  Jiri Gallo; Frantisek Mrazek; Martin Petrek
Journal:  BMC Med Genet       Date:  2009-10-27       Impact factor: 2.103

4.  A simple method of measuring the wear of explanted acetabular component inserts.

Authors:  L Krakow; A Klockow; E Roehner; S Brodt; H Eijer; J Bossert; G Matziolis
Journal:  Bone Joint Res       Date:  2017-09       Impact factor: 5.853

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.