Polyethylene wear and acetabular component orientation.

Shantanu Patil1, Arnie Bergula, Peter C Chen, Clifford W Colwell, Darryl D D'Lima.   

Abstract

BACKGROUND: Polyethylene wear contributes substantially to both periprosthetic osteolysis and aseptic loosening after total hip arthroplasty. Acetabular component orientation has been shown to affect the range of motion of the hip as well as contact stresses. A series of studies was designed to test the hypothesis that acetabular component orientation can affect the magnitude and direction of polyethylene wear.
METHODS: A finite-element model was used to compute contact stresses during a normal gait cycle. Wear at the end of each gait cycle was calculated with use of the sliding-distance-coupled finite-element formulation. The wear that was calculated with use of finite-element analysis was validated by comparison with the findings of hip wear simulator studies with the acetabular liner oriented to simulate 45 degrees and 55 degrees of abduction. In a clinical study, fifty-six patients who underwent sixty hip arthroplasties with use of a single prosthetic design were followed for as long as five years. Radiographs were analyzed to measure the abduction angle of the acetabular component and polyethylene wear.
RESULTS: The finite-element analysis predicted increased peak contact stresses with an increased abduction angle and reduced peak contact stresses with an increased anteversion angle. Linear wear rates ranging from 0.036 to 0.045 mm/million cycles were also predicted, and increased acetabular abduction angles were predicted to be associated with higher linear wear rates. In the hip wear simulator studies, significantly different wear rates were found between the cups with acetabular abduction angles of 45 degrees and 55 degrees (mean, 17.2 compared with 21.7 mg/million cycles; p < 0.01). In the clinical study, radiographic analysis revealed significant correlation between the acetabular abduction angle and the linear polyethylene wear rate. A 40% increase in mean linear polyethylene wear was seen in cups with an abduction angle of >or=45 degrees. The direction of wear was more medial (by 9.4 degrees ) in cups with an abduction angle of <45 degrees.
CONCLUSION: All three studies presented here underlined the importance of optimizing the position of the acetabular component. Careful attention to acetabular position may help to minimize wear.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14652394     DOI: 10.2106/00004623-200300004-00007

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  66 in total

1.  Defining the normal acetabular vault in adult males and females using a novel three-dimensional model.

Authors:  Wael K Barsoum; Travis Smith; Leonard Buller; Feno Monaco; Alison Klika; Constantine Mavroudis; Jason Bryan
Journal:  J Anat       Date:  2012-06-07       Impact factor: 2.610

2.  Improving cup positioning using a mechanical navigation instrument.

Authors:  Simon D Steppacher; Jens H Kowal; Stephen Barry Murphy
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

3.  Intraoperative radiographs for placing acetabular components in hip resurfacing arthroplasty.

Authors:  Thomas P Gross; Fei Liu; Lee Webb
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

4.  Predicting long-term wear performance of hard-on-hard bearing couples: effect of cup orientation.

Authors:  S Shankar; K Gowthaman; M S Uddin
Journal:  Med Biol Eng Comput       Date:  2015-12-30       Impact factor: 2.602

5.  ACETABULAR COMPONENT ORIENTATION IN TOTAL HIP ARTHROPLASTY: THE ROLE OF ACETABULAR TRANSVERSE LIGAMENT.

Authors:  Mohammed El Idrissi; Abdelhalim Elibrahimi; Mohammed Shimi; Abdelmajid Elmrini
Journal:  Acta Ortop Bras       Date:  2016 Sep-Oct       Impact factor: 0.513

6.  Risk factors for accelerated polyethylene wear and osteolysis in ABG I total hip arthroplasty.

Authors:  Jiri Gallo; Vitezslav Havranek; Jana Zapletalova
Journal:  Int Orthop       Date:  2009-02-13       Impact factor: 3.075

7.  Intra- and intersurgeon variability in image-free navigation system for THA.

Authors:  Hirotsugu Ohashi; Masanori Matsuura; Yusaku Okamoto; Yoshiaki Okajima
Journal:  Clin Orthop Relat Res       Date:  2009-04-21       Impact factor: 4.176

8.  Validation of a simple radiographic method to determine variations in pelvic and acetabular cup sagittal plane alignment after total hip arthroplasty.

Authors:  Won Yong Shon; Siddhartha Gupta; Sandeep Biswal; Chang Yong Hur; Nirmal Jajodia; Suk Joo Hong; Jae Sung Myung
Journal:  Skeletal Radiol       Date:  2008-08-07       Impact factor: 2.199

9.  Combined anteversion technique for total hip arthroplasty.

Authors:  Lawrence D Dorr; Aamer Malik; Manish Dastane; Zhinian Wan
Journal:  Clin Orthop Relat Res       Date:  2008-11-01       Impact factor: 4.176

10.  High stress conditions do not increase wear of thin highly crosslinked UHMWPE.

Authors:  Natalie H Kelly; Amar D Rajadhyaksha; Timothy M Wright; Suzanne A Maher; Geoffrey H Westrich
Journal:  Clin Orthop Relat Res       Date:  2010-02       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.