Anatomical background and generalized detectability in tomosynthesis and cone-beam CT.

G J Gang1, D J Tward, J Lee, J H Siewerdsen.   

Abstract

PURPOSE: Anatomical background presents a major impediment to detectability in 2D radiography as well as 3D tomosynthesis and cone-beam CT (CBCT). This article incorporates theoretical and experimental analysis of anatomical background "noise" in cascaded systems analysis of 2D and 3D imaging performance to yield "generalized" metrics of noise-equivalent quanta (NEQ) and detectability index as a function of the orbital extent of the (circular arc) source-detector orbit.
METHODS: A physical phantom was designed based on principles of fractal self-similarity to exhibit power-law spectral density (kappa/Fbeta) comparable to various anatomical sites (e.g., breast and lung). Background power spectra [S(B)(F)] were computed as a function of source-detector orbital extent, including tomosynthesis (approximately 10 degrees -180 degrees) and CBCT (180 degrees + fan to 360 degrees) under two acquisition schemes: (1) Constant angular separation between projections (variable dose) and (2) constant total number of projections (constant dose). The resulting S(B) was incorporated in the generalized NEQ, and detectability index was computed from 3D cascaded systems analysis for a variety of imaging tasks.
RESULTS: The phantom yielded power-law spectra within the expected spatial frequency range, quantifying the dependence of clutter magnitude (kappa) and correlation (beta) with increasing tomosynthesis angle. Incorporation of S(B) in the 3D NEQ provided a useful framework for analyzing the tradeoffs among anatomical, quantum, and electronic noise with dose and orbital extent. Distinct implications are posed for breast and chest tomosynthesis imaging system design-applications varying significantly in kappa and beta, and imaging task and, therefore, in optimal selection of orbital extent, number of projections, and dose. For example, low-frequency tasks (e.g., soft-tissue masses or nodules) tend to benefit from larger orbital extent and more fully 3D tomographic imaging, whereas high-frequency tasks (e.g., microcalcifications) require careful, application-specific selection of orbital extent and number of projections to minimize negative effects of quantum and electronic noise.
CONCLUSIONS: The complex tradeoffs among anatomical background, quantum noise, and electronic noise in projection imaging, tomosynthesis, and CBCT can be described by generalized cascaded systems analysis, providing a useful framework for system design and optimization.

Entities:  

Mesh:

Year:  2010        PMID: 20527529      PMCID: PMC2862054          DOI: 10.1118/1.3352586

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  38 in total

1.  On the statistical nature of mammograms.

Authors:  J J Heine; S R Deans; R P Velthuizen; L P Clarke
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

2.  A method for including redundant data in computed tomography.

Authors:  M D Silver
Journal:  Med Phys       Date:  2000-04       Impact factor: 4.071

3.  Parallel cascades: new ways to describe noise transfer in medical imaging systems.

Authors:  J Yao; I A Cunningham
Journal:  Med Phys       Date:  2001-10       Impact factor: 4.071

4.  Spectral analysis of full field digital mammography data.

Authors:  John J Heine; Robert P Velthuizen
Journal:  Med Phys       Date:  2002-05       Impact factor: 4.071

5.  A framework for noise-power spectrum analysis of multidimensional images.

Authors:  J H Siewerdsen; I A Cunningham; D A Jaffray
Journal:  Med Phys       Date:  2002-11       Impact factor: 4.071

6.  Temporal resolution and the evaluation of candidate algorithms for four-dimensional CT.

Authors:  Katsuyuki Taguchi
Journal:  Med Phys       Date:  2003-04       Impact factor: 4.071

7.  Performance evaluation of a modular gamma camera using a detectability index.

Authors:  John D Sain; Harrison H Barrett
Journal:  J Nucl Med       Date:  2003-01       Impact factor: 10.057

8.  penMesh--Monte Carlo radiation transport simulation in a triangle mesh geometry.

Authors:  Andreu Badal; Iacovos Kyprianou; Diem Phuc Banh; Aldo Badano; Josep Sempau
Journal:  IEEE Trans Med Imaging       Date:  2009-05-08       Impact factor: 10.048

9.  Tomographic mammography using a limited number of low-dose cone-beam projection images.

Authors:  Tao Wu; Alexander Stewart; Martin Stanton; Thomas McCauley; Walter Phillips; Daniel B Kopans; Richard H Moore; Jeffrey W Eberhard; Beale Opsahl-Ong; Loren Niklason; Mark B Williams
Journal:  Med Phys       Date:  2003-03       Impact factor: 4.071

10.  Human observer detection experiments with mammograms and power-law noise.

Authors:  A E Burgess; F L Jacobson; P F Judy
Journal:  Med Phys       Date:  2001-04       Impact factor: 4.071

View more
  38 in total

1.  On the orientation of mammographic structure.

Authors:  I Reiser; S Lee; R M Nishikawa
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

2.  Task-based modeling and optimization of a cone-beam CT scanner for musculoskeletal imaging.

Authors:  P Prakash; W Zbijewski; G J Gang; Y Ding; J W Stayman; J Yorkston; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

3.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

4.  Beyond noise power in 3D computed tomography: the local NPS and off-diagonal elements of the Fourier domain covariance matrix.

Authors:  Angel R Pineda; Daniel J Tward; Antonio Gonzalez; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

5.  A Statistical Model for Rigid Image Registration Performance: The Influence of Soft-Tissue Deformation as a Confounding Noise Source.

Authors:  Michael D Ketcha; Tharindu De Silva; Runze Han; Ali Uneri; Sebastian Vogt; Gerhard Kleinszig; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2019-03-27       Impact factor: 10.048

6.  Cascaded systems analysis of noise and detectability in dual-energy cone-beam CT.

Authors:  Grace J Gang; Wojciech Zbijewski; J Webster Stayman; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

7.  A statistical, task-based evaluation method for three-dimensional x-ray breast imaging systems using variable-background phantoms.

Authors:  Subok Park; Robert Jennings; Haimo Liu; Aldo Badano; Kyle Myers
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

8.  The effect of angular dose distribution on the detection of microcalcifications in digital breast tomosynthesis.

Authors:  Yue-Houng Hu; Wei Zhao
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

9.  Effects of Image Quality on the Fundamental Limits of Image Registration Accuracy.

Authors:  Michael D Ketcha; Tharindu De Silva; Runze Han; Ali Uneri; Joseph Goerres; Matthew W Jacobson; Sebastian Vogt; Gerhard Kleinszig; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2017-07-11       Impact factor: 10.048

Review 10.  Task-based measures of image quality and their relation to radiation dose and patient risk.

Authors:  Harrison H Barrett; Kyle J Myers; Christoph Hoeschen; Matthew A Kupinski; Mark P Little
Journal:  Phys Med Biol       Date:  2015-01-07       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.