Beyond noise power in 3D computed tomography: the local NPS and off-diagonal elements of the Fourier domain covariance matrix.

Angel R Pineda1, Daniel J Tward, Antonio Gonzalez, Jeffrey H Siewerdsen.   

Abstract

PURPOSE: To investigate the correlation and stationarity of noise in volumetric computed tomography (CT) using the local discrete noise-power spectrum (NPS) and off-diagonal elements of the covariance matrix of the discrete Fourier transform of noise-only images (denoted Σ(DFT)). Experimental conditions were varied to affect noise correlation and stationarity, the effects were quantified in terms of the NPS and Σ(DFT), and practical considerations in CT performance characterization were identified.
METHODS: Cone-beam CT (CBCT) images were acquired using a benchtop system comprising an x-ray tube and flat-panel detector for a range of acquisition techniques (e.g., dose and x-ray scatter) and three phantom configurations hypothesized to impart distinct effects on the NPS and Σ(DFT): (A) air, (B) a 20-cm-diameter water cylinder with a bowtie filter, and (C) the cylinder without a bowtie filter. The NPS and off-diagonal elements of the Σ(DFT) were analyzed as a function of position within the reconstructions.
RESULTS: The local NPS varied systematically throughout the axial plane in a manner consistent with changes in fluence transmitted to the detector and view sampling effects. Variability in fluence was manifest in the NPS magnitude-e.g., a factor of ~2 variation in NPS magnitude within the axial plane for case C (cylinder without bowtie), compared to nearly constant NPS magnitude for case B (bowtie filter matched to the cylinder). View sampling effects were most prominent in case A (air) where the variance increased at greater distance from the center of reconstruction and in case C (cylinder) where the NPS exhibited correlations in the radial direction. The effects of detector lag were observed as azimuthal correlation. The cylinder (without bowtie) had the strongest nonstationarity because of the larger variability in fluence transmitted to the detector. The diagonal elements of the Σ(DFT) were equivalent to the NPS estimated from the periodogram, and the average off-diagonal elements of the Σ(DFT) exhibited amplitude of ~1% of the NPS for the experimental conditions investigated. Furthermore, the off-diagonal elements demonstrated fairly long tails of nearly constant amplitude, with magnitude somewhat reduced for experimental conditions associated with greater stationarity (viz., lower Σ(DFT) tails for cases A and B in comparison to case C).
CONCLUSIONS: Volumetric CT exhibits nonstationarity in the NPS as hypothesized in relation to fluence uniformity and view sampling. Measurement of the NPS should seek to minimize such changes in noise correlations and include careful reporting of experimental conditions (e.g., phantom design and use of a bowtie filter) and spatial dependence (e.g., analysis at fixed radius within a phantom). Off-diagonal elements of the Σ(DFT) similarly depend on experimental conditions and can be readily computed from the same data as the NPS. This work begins to check assumptions in NPS analysis examine the extent to which NPS is an appropriate descriptor of noise correlations, and investigate the magnitude of off-diagonal elements of the Σ(DFT). While the magnitude of such off-diagonal elements appears to be low, their cumulative effect on space-variant detectability remains to be investigated-e.g., using task-specific figures of merit.
© 2012 American Association of Physicists in Medicine.

Entities:  

Mesh:

Year:  2012        PMID: 22755707      PMCID: PMC3365918          DOI: 10.1118/1.4705354

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  26 in total

1.  Cone-beam computed tomography with a flat-panel imager: effects of image lag.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

2.  Optimization of x-ray imaging geometry (with specific application to flat-panel cone-beam computed tomography).

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

3.  Anatomical background and generalized detectability in tomosynthesis and cone-beam CT.

Authors:  G J Gang; D J Tward; J Lee; J H Siewerdsen
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

4.  Noise aliasing and the 3D NEQ of flat-panel cone-beam CT: effect of 2D/3D apertures and sampling.

Authors:  Daniel J Tward; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

5.  The noise power spectrum of CT images.

Authors:  M F Kijewski; P F Judy
Journal:  Phys Med Biol       Date:  1987-05       Impact factor: 3.609

6.  The noise power spectrum in computed X-ray tomography.

Authors:  S J Riederer; N J Pelc; D A Chesler
Journal:  Phys Med Biol       Date:  1978-05       Impact factor: 3.609

7.  Detectability in computed tomographic images.

Authors:  K M Hanson
Journal:  Med Phys       Date:  1979 Sep-Oct       Impact factor: 4.071

8.  Noise due to photon counting statistics in computed X-ray tomography.

Authors:  D A Chesler; S J Riederer; N J Pelc
Journal:  J Comput Assist Tomogr       Date:  1977-01       Impact factor: 1.826

9.  Human observer detection experiments with mammograms and power-law noise.

Authors:  A E Burgess; F L Jacobson; P F Judy
Journal:  Med Phys       Date:  2001-04       Impact factor: 4.071

10.  Characterizing anatomical variability in breast CT images.

Authors:  Kathrine G Metheany; Craig K Abbey; Nathan Packard; John M Boone
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

View more
  17 in total

1.  Investigation of Noise and Contrast Sensitivity of an Electron Multiplying Charge-Coupled Device (EMCCD) based Cone Beam Micro-CT System.

Authors:  Sumukh Bysani Krishnakumar; Alexander R Podgorsak; Sv Setlur Nagesh; Amit Jain; Stephen Rudin; Daniel R Bednarek; Ciprian N Ionita
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-25

2.  Effects of Image Quality on the Fundamental Limits of Image Registration Accuracy.

Authors:  Michael D Ketcha; Tharindu De Silva; Runze Han; Ali Uneri; Joseph Goerres; Matthew W Jacobson; Sebastian Vogt; Gerhard Kleinszig; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2017-07-11       Impact factor: 10.048

Review 3.  Task-based measures of image quality and their relation to radiation dose and patient risk.

Authors:  Harrison H Barrett; Kyle J Myers; Christoph Hoeschen; Matthew A Kupinski; Mark P Little
Journal:  Phys Med Biol       Date:  2015-01-07       Impact factor: 3.609

4.  Fundamental relationship between the noise properties of grating-based differential phase contrast CT and absorption CT: theoretical framework using a cascaded system model and experimental validation.

Authors:  Ke Li; Nicholas Bevins; Joseph Zambelli; Guang-Hong Chen
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

5.  Task-based detectability in CT image reconstruction by filtered backprojection and penalized likelihood estimation.

Authors:  Grace J Gang; J Webster Stayman; Wojciech Zbijewski; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

6.  Task-based optimization of dedicated breast CT via Hotelling observer metrics.

Authors:  Adrian A Sanchez; Emil Y Sidky; Xiaochuan Pan
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

7.  Task-driven image acquisition and reconstruction in cone-beam CT.

Authors:  Grace J Gang; J Webster Stayman; Tina Ehtiati; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-03-24       Impact factor: 3.609

8.  Generation and analysis of clinically relevant breast imaging x-ray spectra.

Authors:  Andrew M Hernandez; J Anthony Seibert; Anita Nosratieh; John M Boone
Journal:  Med Phys       Date:  2017-05-04       Impact factor: 4.071

9.  Methods for CT automatic exposure control protocol translation between scanner platforms.

Authors:  Sarah E McKenney; J Anthony Seibert; Ramit Lamba; John M Boone
Journal:  J Am Coll Radiol       Date:  2014-03       Impact factor: 5.532

10.  Predicting image properties in penalized-likelihood reconstructions of flat-panel CBCT.

Authors:  Wenying Wang; Grace J Gang; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Med Phys       Date:  2018-11-20       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.