On the influence of noise correlations in measurement data on basis image noise in dual-energylike x-ray imaging.

Ewald Roessl1, Andy Ziegler, Roland Proksa.   

Abstract

In conventional dual-energy systems, two transmission measurements with distinct spectral characteristics are performed. These measurements are used to obtain the line integrals of two basis material densities. Usually, the measurement process is such that the two measured signals can be treated as independent and therefore uncorrelated. Recently, however, a readout system for x-ray detectors has been introduced for which this is no longer the case. The readout electronics is designed to obtain simultaneous measurements of the total number of photons N and the total energy E they deposit in the sensor material. Practically, this is realized by a signal replication and separate counting and integrating processing units. Since the quantities N and E are (electronically) derived from one and the same physical sensor signal, they are statistically correlated. Nevertheless, the pair N and E can be used to perform a dual-energy processing following the well-known approach by Alvarez and Macovski. Formally, this means that N is to be identified with the first dual-energy measurement M1 and E with the second measurement M2. In the presence of input correlations between M1 = N and M2 = E, however, the corresponding analytic expressions for the basis image noise have to be modified. The main observation made in this paper is that for positively correlated data, as is the case for the simultaneous counting and integrating device mentioned above, the basis image noise is suppressed through the influence of the covariance between the two signals. We extend the previously published relations for the basis image noise to the case where the original measurements are not independent and illustrate the importance of the input correlations by comparing dual-energy basis image noise resulting from the device mentioned above and a device measuring the photon numbers and the deposited energies consecutively.

Mesh:

Substances:

Year:  2007        PMID: 17441242     DOI: 10.1118/1.2514058

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Quantification of breast density with dual energy mammography: a simulation study.

Authors:  Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

2.  Near optimal energy selective x-ray imaging system performance with simple detectors.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

3.  Signal to noise ratio of energy selective x-ray photon counting systems with pileup.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

4.  The effects of extending the spectral information acquired by a photon-counting detector for spectral CT.

Authors:  Taly Gilat Schmidt; Kevin C Zimmerman; Emil Y Sidky
Journal:  Phys Med Biol       Date:  2015-01-23       Impact factor: 3.609

5.  Noise suppression for dual-energy CT via penalized weighted least-square optimization with similarity-based regularization.

Authors:  Joseph Harms; Tonghe Wang; Michael Petrongolo; Tianye Niu; Lei Zhu
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

Review 6.  Imaging atherosclerosis and vulnerable plaque.

Authors:  Mehran M Sadeghi; David K Glover; Gregory M Lanza; Zahi A Fayad; Lynne L Johnson
Journal:  J Nucl Med       Date:  2010-04-15       Impact factor: 10.057

7.  Spectral prior image constrained compressed sensing (spectral PICCS) for photon-counting computed tomography.

Authors:  Zhicong Yu; Shuai Leng; Zhoubo Li; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2016-08-23       Impact factor: 3.609

Review 8.  CT angiography: current technology and clinical use.

Authors:  Kanako K Kumamaru; Bernice E Hoppel; Richard T Mather; Frank J Rybicki
Journal:  Radiol Clin North Am       Date:  2010-03       Impact factor: 2.303

9.  A neural network-based method for spectral distortion correction in photon counting x-ray CT.

Authors:  Mengheng Touch; Darin P Clark; William Barber; Cristian T Badea
Journal:  Phys Med Biol       Date:  2016-07-29       Impact factor: 3.609

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.