Determination of linear x-ray attenuation coefficients of pathological brain tissues and use of filters in tissue contrast enhancement in computed tomography.

M Erdem Sagsoz1, Fazlı Erdogan2, Salih Z Erzeneoglu3, İhsan Yuce1.   

Abstract

OBJECTIVE: X-ray attenuation coefficients are used in common radiological, pathological and spectroscopic examinations and in the determination of the radiation dose distribution in biological tissues. In radiology, these coefficients enable diagnosis by differentiating the abnormal tissues from the normal ones using their morphological structure and contrast differences. In this study, our aim is to precisely determine the linear x-ray attenuation coefficients of pathological brain tissues and to use x-ray beam filters to enhance the tissue contrast in computed tomography.
MATERIALS AND METHODS: To directly measure the relative linear attenuation coefficients, an energy dispersive x-ray spectroscopy system (EDXRS-Canberra, Si(Li) with DSA-1000 spectrum analyzer 1998; CT, USA) was used with collimators and a medical-purpose x-ray tube (Siemens, Siremobil, 1985; Erlangen, Germany) in a linear geometry.
RESULTS: Using a Mo filter with Computed Tomography CT and photon energies from 15 to 25 keV, EDXRS acquisitions were found to significantly distinguish grades of brain tumors (p<0.05). For the data acquired from CT systems with the decreasing filtered photon mean energy, the x-ray attenuation coefficients (i.e., the Hounsfield units) show that the ratio of EDXRS to CT for water's attenuation coefficient are increased. With our suggested x-ray filters, the tissue contrast has been found to be increased in ex vivo brain tumor slices compared with slices scanned in conventional CT scanners.
CONCLUSION: X-ray attenuations measured with the EDXRS are found to be statistically more reliable because of the length of acquisition times in this study.

Entities:  

Keywords:  Computed tomography; X-ray attenuation coefficient; X-ray spectral filters; X-ray tube

Year:  2010        PMID: 25610123      PMCID: PMC4261332          DOI: 10.5152/eajm.2010.17

Source DB:  PubMed          Journal:  Eurasian J Med        ISSN: 1308-8734


  19 in total

1.  Preliminary experiment of fluorescent X-ray computed tomography to detect dual agents for biological study.

Authors:  Q Yu; T Takeda; T Yuasa; Y Hasegawa; J Wu; K Hyodo; F A Dilmanian; Y Itai; T Akatsuka
Journal:  J Synchrotron Radiat       Date:  2001-05-01       Impact factor: 2.616

2.  Computed tomography with monochromatic bremsstrahlung radiation

Authors: 
Journal:  Appl Radiat Isot       Date:  2000-10       Impact factor: 1.513

3.  Systematic bias in basis material decomposition applied to quantitative dual-energy x-ray imaging.

Authors:  E L Gingold; B H Hasegawa
Journal:  Med Phys       Date:  1992 Jan-Feb       Impact factor: 4.071

4.  Dual energy x-ray absorptiometry: the effects of beam hardening on bone density measurements.

Authors:  G M Blake; D B McKeeney; S C Chhaya; P J Ryan; I Fogelman
Journal:  Med Phys       Date:  1992 Mar-Apr       Impact factor: 4.071

5.  A solid iodinated phantom material for use in tomographic x-ray imaging.

Authors:  Melissa L Hill; James G Mainprize; Gordon E Mawdsley; Martin J Yaffe
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

6.  Near optimal energy selective x-ray imaging system performance with simple detectors.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

7.  Single-and dual-energy CT with monochromatic synchrotron x-rays.

Authors:  F A Dilmanian; X Y Wu; E C Parsons; B Ren; J Kress; T M Button; L D Chapman; J A Coderre; F Giron; D Greenberg; D J Krus; Z Liang; S Marcovici; M J Petersen; C T Roque; M Shleifer; D N Slatkin; W C Thomlinson; K Yamamoto; Z Zhong
Journal:  Phys Med Biol       Date:  1997-02       Impact factor: 3.609

8.  Contrast and dose with Mo-Mo, Mo-Rh, and Rh-Rh target-filter combinations in mammography.

Authors:  E L Gingold; X Wu; G T Barnes
Journal:  Radiology       Date:  1995-06       Impact factor: 11.105

9.  Selective iodine imaging using K-edge energies in computerized x-ray tomography.

Authors:  S J Riederer; C A Mistretta
Journal:  Med Phys       Date:  1977 Nov-Dec       Impact factor: 4.071

10.  Phase-contrast X-ray computed tomography for observing biological soft tissues.

Authors:  A Momose; T Takeda; Y Itai; K Hirano
Journal:  Nat Med       Date:  1996-04       Impact factor: 53.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.