Giuseppe Minniti1, Maurizio Valeriani, Enrico Clarke, Marco D'Arienzo, Michelangelo Ciotti, Roberto Montagnoli, Francesca Saporetti, Riccardo Maurizi Enrici.
Abstract
BACKGROUND: To assess the accuracy of fractionated stereotactic radiotherapy (FSRT) using a stereotactic mask fixation system. PATIENTS AND METHODS: Sixteen patients treated with FSRT were involved in the study. A commercial stereotactic mask fixation system (BrainLAB AG) was used for patient immobilization. Serial CT scans obtained before and during FSRT were used to assess the accuracy of patient immobilization by comparing the isocenter position. Daily portal imaging were acquired to establish day to day patient position variation. Displacement errors along the different directions were calculated as combination of systematic and random errors.Entities:
Mesh:
Year: 2010 PMID: 20070901 PMCID: PMC2823752 DOI: 10.1186/1748-717X-5-1
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Patient with mask fixation. The system consists of a semicircular metal frame, an upper and a lower mask conformed to the anterior and posterior surfaces of head, two lateral carbon bars for fixing the thermoplastic mask, and a mouth bite.
Figure 2Verification of isocenter position accuracy. During CT localization (A) and CT verification (B) the patient is positioned on the CT couch with the target positioner box aligned with anterior and lateral lasers using the radio-opaque markers (arrows). The amount of isocenter shift between CT localization (planning CT) (C) and CT verification (D) in relation to anatomical skull base cranial structures was then evaluated directly on the CT scans.
Positioning deviations of isocenter relocation at CT verification before and during radiation treatment.
| pre-treatment | 2 weeks | 4 weeks | end of treatment | |
|---|---|---|---|---|
| Direction (mm) | mean (SD) | mean (SD) | mean (SD) | mean (SD) |
| Craniocaudal | 0.3 (0.4) | 0.3 (0.5) | 0.3 (0.6) | 0.4 (0.5) |
| Mediolateral | 0.1 (0.3) | 0.1 (0.4) | 0.2 (0.4) | 0.2 (0.4) |
| Anteroposterior | 0.1 (0.4) | 0.2 (0.4) | 0.2 (0.4) | 0.2 (0.5) |
| 3D-displacement | 0.5 (0.4) | 0.5 (0.4) | 0.6 (0.4) | 0.6 (0.5) |
SD, standard deviation
Mean and standard deviation of overall, systematic, and random setup errors at portal images during the treatment (n = 456)
| Distribution of displacements (1 SD, mm) | ||||
|---|---|---|---|---|
| Overall displacement | Overall | Systematic | Random | |
| (mm) | (n = 456) | (n = 16) | (n = 456) | |
| Mediolateral | 0.3 | 0.9 | 0.7 | 0.5 |
| Anteroposterior | -0.2 | 1 | 0.8 | 0.6 |
| Craniocaudal | 0.2 | 1.1 | 0.9 | 0.7 |
SD, standard deviation