Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system.

C P Karger1, O Jäkel, J Debus, S Kuhn, G H Hartmann.   

Abstract

PURPOSE: Conformal radiotherapy in the head and neck region requires precise and reproducible patient setup. The definition of safety margins around the clinical target volume has to take into account uncertainties of fixation and positioning. Data are presented to quantify the involved uncertainties for the system used. METHODS AND MATERIALS: Interfractional reproducibility of fixation and positioning of a target point in the brain was evaluated by biplanar films. 118 film pairs obtained at 52 fractions in 4 patients were analyzed. The setup was verified at the actual treatment table position by diagnostic X-ray units aligned to the isocenter and by a stereotactic X-ray localization technique. The stereotactic coordinates of the treated isocenter, of fiducials on the mask, and of implanted internal markers within the patient were measured to determine systematic and random errors. The data are corrected for uncertainty of the localization method.
RESULTS: Displacements in target point positioning were 0.35 +/- 0.41 mm, 1.22 +/- 0.25 mm, and -0.74 +/- 0.32 mm in the x, y, and z direction, respectively. The reproducibility of the fixation of the patient's head within the mask was 0.48 mm (x), 0.67 mm (y), and 0.72 mm (z). Rotational uncertainties around an axis parallel to the x, y, and z axis were 0.72 degrees, 0.43 degrees, and 0.70 degrees, respectively. A simulation, based on the acquired data, yields a typical radial overall uncertainty for positioning and fixation of 1.80 +/- 0.60 mm.
CONCLUSIONS: The applied setup technique showed to be highly reproducible. The data suggest that for the applied technique, a safety margin between clinical and planning target volume of 1-2 mm along one axis is sufficient for a target at the base of skull.

Entities:  

Mesh:

Year:  2001        PMID: 11286858     DOI: 10.1016/s0360-3016(00)01562-5

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  25 in total

Review 1.  Current progress in adaptive radiation therapy for head and neck cancer.

Authors:  David L Schwartz
Journal:  Curr Oncol Rep       Date:  2012-04       Impact factor: 5.075

2.  Clinical results of a pilot study on stereovision-guided stereotactic radiotherapy and intensity modulated radiotherapy.

Authors:  Shidong Li; Lawrence R Kleinberg; Daniele Rigamonti; Moody D Wharam; Abdul Rashid; Juan Jackson; David Djajaputra; Shenjen He; Tunisia Creasey; Theodore L DeWeese
Journal:  Technol Cancer Res Treat       Date:  2010-12

3.  Rotational and translational reproducibility of newly developed Leksell frame-based relocatable fixation system.

Authors:  Etsuo Kunieda; Hossain M Deloar; Masayuki Kitamura; Osamu Kawaguchi; Hideyuki Shiba; Atsuya Takeda; Takatsugu Kawase; Satoshi Seki; Naoyuki Shigematsu; Atsushi Kubo
Journal:  Radiat Med       Date:  2006-08

4.  Variable planning margin approach to account for locoregional variations in setup uncertainties.

Authors:  Jinzhong Yang; Adam S Garden; Yongbin Zhang; Lifei Zhang; Lei Dong
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

5.  Analysis of intrafraction motion in CyberKnife-based stereotaxy using mask based immobilization and 6D-skull tracking.

Authors:  Tejinder Kataria; Kushal Narang; Deepak Gupta; Shyam S Bisht; Ashu Abhishek; Shikha Goyal; Trinanjan Basu; K P Karrthick
Journal:  J Radiosurg SBRT       Date:  2016

6.  Relocatable fixation systems in intracranial stereotactic radiotherapy. Accuracy of serial CT scans and patient acceptance in a randomized design.

Authors:  A Theelen; J Martens; G Bosmans; R Houben; J J Jager; I Rutten; P Lambin; A W Minken; B G Baumert
Journal:  Strahlenther Onkol       Date:  2011-12-24       Impact factor: 3.621

7.  Adaptive radiation therapy for head and neck cancer-can an old goal evolve into a new standard?

Authors:  David L Schwartz; Lei Dong
Journal:  J Oncol       Date:  2010-08-18       Impact factor: 4.375

Review 8.  Modern techniques for pituitary radiotherapy.

Authors:  G Minniti; D C Gilbert; M Brada
Journal:  Rev Endocr Metab Disord       Date:  2009-06       Impact factor: 6.514

9.  Fractionated stereotactic radiotherapy for skull base tumors: analysis of treatment accuracy using a stereotactic mask fixation system.

Authors:  Giuseppe Minniti; Maurizio Valeriani; Enrico Clarke; Marco D'Arienzo; Michelangelo Ciotti; Roberto Montagnoli; Francesca Saporetti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2010-01-13       Impact factor: 3.481

10.  Quality assurance and commissioning of an infrared marker-based patient positioning system for frameless extracranial stereotactic radiotherapy.

Authors:  Tejpal Gupta; Reena Phurailatpam; Mishra Ajay; Pai Rajeshri; Mohindra Pranshu; Chopra Supriya
Journal:  Int J Biomed Sci       Date:  2007-12
View more